TR-55i

TR-55i

TR-55i

Product catalog summary
Overview
This document provides detailed specifications and features of the TR-5i Series data loggers, focusing on temperature, voltage, 4-20mA, and pulse count measurements. It outlines the data collection methods, device compatibility, and warning notification systems.

Data Collection and Access
The TR-5i Series supports data collection through T&D devices, with data accessible via a local PC. Data can be downloaded directly to a PC using the TR-50U2 USB communication port or on-site using the TR-57DCi data collector via infrared communication.

Warning Notification
The devices feature a warning LED that flashes when readings exceed set limits, accompanied by an on-screen error message. This warning remains until data is downloaded.

TR-5i Series Features
  • Compact, waterproof, and dustproof design suitable for harsh environments.
  • Logging capacity of up to 16,000 readings with 15 selectable recording intervals ranging from 1 second to 1 hour.
  • Temperature adjustment function for viewing and recording adjusted values.

TR-51i and TR-52i Specifications
  • TR-51i: Internal sensor, temperature range -40 to 80°C, IP67 waterproof rating.
  • TR-52i: External sensor, temperature range -60 to 155°C, splashproof with a wide selection of optional sensors.

TR-55i Specifications
The TR-55i model supports various sensors and measurement types:
  • TR-55i-TC: Thermocouple sensors (K, J, T, S), temperature range -199 to 1760°C.
  • TR-55i-Pt: Pt100/Pt1000 sensors, temperature range -199 to 600°C.
  • TR-55i-V: Voltage measurement, range 0 to 22 V.
  • TR-55i-mA: 4-20mA current measurement, operational up to 40mA.
  • TR-55i-P: Pulse count measurement, maximum count 61,439 per interval.

Additional Features
  • Scale conversion for TR-55i models to automatically convert recorded data to desired units.
  • Two recording methods: instantaneous and average.
  • Preheat function for TR-55i-V to save battery life by powering the sensor only before measurement.

Power and Battery Life
All models use a lithium battery (LS14250) with varying battery life depending on usage and communication settings. Battery life ranges from approximately 10 months to 4 years.

Environmental and Physical Specifications
All models are designed to operate in environments ranging from -40 to 80°C, with varying degrees of waterproof and dustproof capabilities.
See more

Catalog excerpts

TR-55i-1

Compact Waterproof Data Logger Temperature / Voltage / 4-20mA / Pulse Count T&D Data Collection Devices The TR-51i with internal sensor provides optimum waterproof and dustproof capabilities. Common Features Warning Monitoring Function Collect Data using our Data Collection Devices For Direct Data Downloading to PC via USB Communication Port TR-50U2 (sold separately) For On-Site Downloading (up to 16 units) Data Collector TR-57DCi (sold separately) Download data via Infrared Communication Graph Display 15 Recording Intervals with Large Logging Capacity Up to 16,000 readings per unit. Select from 15 recording intervals (from 1 second to 1 hour). Example: At a 60-minute recording interval, that gives you 666 days (almost 2 years) of non-stop recording. Temperature Adjustment Function If a reading exceeds the set limit the warning LED will flash and an on-screen error message will appear. Warning display will stay on until data is downloaded. Measurement and error message shown alternately Upper Limit Exceeded (Hi) A data logger built for tough environments Scale Conversion (TR-55i-V / mA / P) Upon the downloading of recorded data, the scale will be automatically converted according to pre-made settings. Graph will be displayed using the desired unit. Two Recording Methods (TR-55i-V / mA) Instantaneous: Records the measured value instantaneously at the set recording interval. Average: Records the average of the measurements taken during the set recording interval. Entering adjustment values beforehand makes it possible to view and record the adjusted measurement values. Preheat Function (TR-55i-V) This supplies power to the sensor immediately before measurement and turns off power until the next measurement. This saves battery life allowing for longer continuous recording.

 Open the catalog to page 1
TR-55i-2

*1: Battery life depends upon multiple factors including measuring environment, recording interval, and quality of the battery being used. When infrared communication function is enabled, battery life may be further shortened if the unit is used under the inverter type fluorescent lighting. *2: This is the waterproof capacity of the data logger with the sensor connected. The specifications listed above are subject to change without notice.

 Open the catalog to page 2
TR-55i-3

*1: In the case of a 4-wire sensor, one wire will be left unused. *2: For TR-55i-TC and TR-55i-Pt, sensor inaccuracies are not included. *3: Battery life varies depending upon multiple factors including ambient temperature, recording interval, frequency of communication, and battery performance. All estimates are based on operations carried out with a new battery and are in no way a guarantee of actual battery life. Battery life may be shortened further if the unit is used under inverter type fluorescent lighting. *4: This is the waterproof capacity of the data logger with the Input Module connected....

 Open the catalog to page 3

All T&D Corporation catalogs and technical brochures

  1. TR4A

    1  Page

  2. TR7A

    1  Page

  3. RTR500B