Automatic Aseptic Filling Isolator ZHEJIANG TAILIN BIOENGINEERING CO., LTD
Open the catalog to page 1Automatic Aseptic Filling lsolator Automatic Aseptic Filling lsolator AFL Series Aseptic Filling Isolator adopts a three chamber design, consisting of transfer chamber, filling chamber and capping chamber, and is equipped with integrated Vaporized Hydrogen Peroxide sterilization system to sterilise the chambers together or separately.Customisation is possible. It adopts modular assembly and automated production to improve the efficiency of aseptic preparation production. At the same time, it can solve the risk of contamination during the filling process of aseptic preparations and ATMPs. Applications...
Open the catalog to page 2Automatic Aseptic Filling lsolator - Automated Leak Testing: Equipped with a automated leak test system, each chamber is individually leak tested according to ISO10648-2, leakage rate is less than 0.5% volume/hour at 2 times working pressure - Glove Integrity Testing: Integrated automated glove integrity tester(GIT series) that performs an independent leakage test on each glove mounted on the isolator according to ISO 14644-7; Capable of detecting holes down to 100μm diameter - Data Management: Integrated control of the isolator from an IPC With multi-level login access control, audit trail and...
Open the catalog to page 3Zhejiang Tailin Bioengineering Co., Ltd No.2930 Nanhuan Road, Binjiang, Hangzhou, Zhejiang, China. TEL: 0086-571-86589087 E-MAIL: [email protected] @Tailin All Copyright Reserved By Tailin