1. Catalogs
  2. Techlab Works
  3. Xe Plug-in Environment
video corpo

Xe Plug-in Environment

Xe Plug-in Environment

Xe Plug-in Environment

Product catalog summary
Overview: The document provides technical specifications for the Xe TMOR's Control Touch Panel and the SensOR Box, both designed for use in medical environments such as operating theatres.
Control Touch Panel Specifications:
  • Display: Integrated touch display with a 24-inch model (Model n. TLW_MS_XEPL_004_01).
  • Inputs: Includes analog/digital inputs and interrupt isolated inputs with galvanic isolation.
  • Outputs: Features analog, digital isolated relay, and PWM isolated outputs.
  • Connections: Ethernet and HDMI video outputs with a resolution up to 1920 x 1080.
  • Touch Technology: PCAP (TouchPro® Projected Capacitive) supporting 10 touch points.
  • Power Supply: AC 100 - 240 V, 50/60 Hz, max 450 W.
  • Certifications: UL60601-1, EN60601-1, CCC, CE, FCC Parts 18.
  • Operating Conditions: Temperature range from 0°C to 40°C, humidity from 20% to 80%.
SensOR Box Specifications:
  • Model: TLW_MS_XE_BSENS_001.
  • Parameters Monitored: Temperature, Humidity, eCO2, Differential pressure, VOC, Particulate matter.
  • Sensor Operating Range: Temperature from -40°C to +85°C, Humidity from 0% to 100% r.H., eCO2 from 400-60000 ppm.
  • Accuracy: Temperature ± 0.5°C at 25°C, Humidity ± 3% r.H., TVOC and particulate matter with specified accuracy ranges.
  • Precision Measure: High resolution for temperature, humidity, eCO2, differential pressure, and particulate matter.
  • Installation: Can be installed on walls or ceilings.
Additional Information: The document notes that specifications, designs, and accessories are subject to change without notice. The company, Techlab Works s.r.l., is based in Italy and provides contact information for further inquiries.
See more

Catalog excerpts

Xe Plug-in Environment-1

Copyright© 2022 TechLab Works s.r.l.- P.I. 04603060874 – All rights reserved. OR’s Control Touch Panel Control Environment The OR’s control touch panel for the control of your Operating Theatre Integrated touch display Web data storage and analytics Light control Gas alarms Ventilation and air flow Synchronization with Hospital Time System Xe Clock Environment Sensors Monito

 Open the catalog to page 1
Xe Plug-in Environment-2

An/Dig Input 10bit - (x4) [0 to 10Vac Input Impedance: 39K Separated PCB ground Rated Voltage: 10Vac 7 to 24Vdc I min: 2 to 12 mA Galvanic Isolation Rated Voltage: 24 Vdc] • Interrupt isolated Input HS - (x2) [7 to 24Vdc I min: 2 to 12 mA Galvanic Isolation Rated Voltage: 24 Vdc] • Analog Output 8bit - (x3) [0 to 10Vac I max: 20 mA Separated PCB ground Rated Voltage: 10Vac] • Digital Isolated Output Relay - (x8) [220V Vdc I max: 5A Galvanic Isolation Diode protected for Relay] PWM Isolated Output 8bit - (x3) [5 to 24Vdc I max: 70 mA Galvanic Isolation Diode Protected for Relay Rated Voltage:...

 Open the catalog to page 2

All Techlab Works catalogs and technical brochures

  1. Endostation

    3  Pages

  2. Xe™

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.