A502-480

A502-480

Product catalog summary
Intended Use: The albumin reagent is designed for the quantitative determination of albumin concentration in human serum.
Principle: The method involves serum albumin binding selectively to bromcresol green dye at pH 4.2. The absorbance increase of the albumin-dye complex at 630 nm is proportional to the albumin concentration.
Contact Information: TECO DIAGNOSTICS, 1268 N. Lakeview Avenue, Anaheim, CA 92807. Tel: 714-463-1111, Fax: 714-463-1169.
Test Details:
  • Test: Albumin Reagent Set (A502-480)
  • Number of tests: 480
  • Format: Liquid
  • Method: Endpoint BCG
  • Testing Procedure: Manual & Automated
  • Storage Temperature: Reagent at room temperature (15-30°C), Albumin standard refrigerated (2-8°C)
  • Wavelength: 630 nm
  • Linearity: 0.5 – 8.0 g/dL
  • Expected Values: 3.5 – 5.3 g/dL
Reagent Deterioration: The reagent should be a clear, yellow-green solution. Discard if turbidity or precipitation occurs.
Limitations of Procedure:
  • Ampicillin and other medications can interfere with dye-binding properties.
  • Dye-binding properties vary among species.
  • Samples above 8.0 g/dL should be diluted with 0.9% saline 1:1, re-run, and results multiplied by 2.
  • Samples below 0.5 g/dL should be analyzed electrophoretically.
See more

Catalog excerpts

A502-480-1

Test: Albumin Reagent Set (A502-480) Intended Use Albumin reagent is used for the quantitative determination of albumin concentration in human serum. Number of tests: 480 tests Format: Liquid Method: Endpoint BCG Principle Serum albumin binds selectively to the dye bromcresol green at pH 4.2. The increase in absorbance of the resulting albumin-dye complex, read at 630 nm, is proportional to the albumin concentration. Testing Procedure: Manual & Automated Storage Temperature: Reagent is stored at room temperature (15-30°C) Albumin standard is stored refrigerated (2-8°C) Wavelength: 630 nm Linearity: 0.5 - 8.0 g/dL Expected Values: 3.5 - 5.3 g/dL Reagent Deterioration: The reagent should be a clear, yellow-green solution. If turbidity or precipitation has occurred, the reagent should be discarded. Limitations of Procedure: Ampicillin and other medications seriously interfere with the dye-binding properties of albumin. The dye-binding properties of albumin, other than human, differ among species. Sample with values above 8.0 g/dL should be diluted with 0.9% saline 1:1, re-run, and results multiplied by 2. Samples with results below 0.5 g/dL should be done electrophoretically. CONTACT US: TECO DIAGNOSTICS 1268 N. Lakeview Avenue Anaheim, CA 92807 Tel: 714-463-1111 Fax: 714-463-1169

 Open the catalog to page 1

All Teco Diagnostics catalogs and technical brochures

  1. RPR – 100

    1  Page

  2. C509-400

    1  Page

  3. RF‐100

    1  Page

  4. URS‐11

    2  Pages

  5. P605-510TM

    1  Page

  6. ASO-100

    1  Page

  7. B577-480

    1  Page

  8. URS‐1A

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.