P605-510TM

P605-510TM

Product catalog summary
Intended Use: This document describes the colorimetric determination of potassium in human serum and plasma using TC Matrix analyzers.
Principle: Potassium levels are measured by creating a colloidal suspension with sodium tetraphenylboron, where the turbidity is proportional to the potassium concentration.
Contact Information: TECO DIAGNOSTICS, 1268 N. Lakeview Avenue, Anaheim, CA 92807. Tel: 714-463-1111, Fax: 714-463-1169.
Test Details: The test is a Potassium Reagent (Colorimetric Method) for TC Matrix (P605 – 510TM) with 510 tests available in liquid format. The method is colorimetric and endpoint, conducted automatically.
Storage and Testing Conditions: Store at 15 – 30°C. The wavelength for testing is 510 nm.
Expected Values: The expected potassium concentration range is 3.4 – 5.3 mEq/L, but laboratories should establish their own expected values.
Linearity: The test is linear between 2 – 7 mEq/L.
Reagent Deterioration: Do not use the reagent if it is very cloudy or fails to achieve the assigned value on fresh control serum.
Limitations of Procedure: Turbid or icteric samples, bilirubin above 40 mg/dL, urea nitrogen above 80 mg/dL, and hemolyzed sera can produce falsely elevated results. Sera with high ammonia levels should be avoided.
See more

Catalog excerpts

P605-510TM-1

Intended Use For the colorimetric determination of potassium in human serum and plasma on TC Matrix analyzers. Principle The amount of potassium is determined by using sodium tetraphenylboron in a specifically prepared mixture to produce a colloidal suspension, the turbidity of which is proportional to the potassium concentration in the sample. CONTACT US: TECO DIAGNOSTICS 1268 N. Lakeview Avenue Anaheim, CA 92807 Tel: 714‐463‐1111 Fax: 714‐463‐1169 Test: Potassium Reagent (Colorimetric Method) for TC Matrix(P605 – 510TM) Number of Tests: 510 tests Format: Liquid Method: Colorimetric, Endpoint Testing Procedure: Automated Storage Temperature: 15 – 30°C Wavelength: 510 nm Expected Values: 3.4 – 5.3 mEq/L It is recommended that each laboratory establish its own range of expected values. Linearity: 2 – 7 mEq/L Reagent Deterioration : Do not use if: (1) The reagent is very cloudy; (2)T he reagent fails to achieve assigned value on fresh control serum. Limitations of Procedure: Turbid or icteric samples produce falsely elevated results

 Open the catalog to page 1

All Teco Diagnostics catalogs and technical brochures

  1. RPR – 100

    1  Page

  2. C509-400

    1  Page

  3. RF‐100

    1  Page

  4. URS‐11

    2  Pages

  5. ASO-100

    1  Page

  6. B577-480

    1  Page

  7. A502-480

    1  Page

  8. URS‐1A

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.