BLAYCO active electrodes are designed for electrosurgical pencils with non-stick coating. The product range includes ball, needle, and blade electrodes with varying dimensions and references: TCET1 (Ball, 80mm), TCET2 (Needle, 70mm), TCET3 (Needle, 150mm), TCET4 (Blade, 70mm), and TCET5 (Blade, 160mm).
Technical Specifications
  • Materials: Stainless steel AISI 304 for the electrode rod, polytetrafluoroethylene for the anti-adherent tip coating, ABS for the mould, PVC for the protector, and PE for the inner pouch.
  • Dimensional Properties: The stainless steel rod has a diameter of 2.36 ± 0.02mm. Electrode insertion is either circular or hexagonal, with specific dimensions for each type.
Biocompatibility
The product is biocompatible, meeting ISO 10993-5 and ISO 10993-10 standards, indicating no cytotoxicity, sensitization, or intracutaneous irritation.
Sterilization and Expiry Date
The product is sterilized using ethylene oxide and has a shelf life of 3 years from the manufacturing date.
Storage and Maintenance
Store in a dry place, away from direct sunlight, within a temperature range of 5-40ºC. The product is single-use and requires no maintenance.
Packaging
Each unit is individually packaged, with a service box containing 50 units.
Product Characteristics
The electrodes are compatible with all Blayco Electrosurgical pencils and are latex-free.
Regulatory Information
The product complies with Directive 93/42/EEC and is manufactured under ISO 13485 certified Quality Assurance System. It is classified as a Class IIb sterile product with GMDN code 11499.
See more

Catalog excerpts

Active electrodes for electrosurgical pencil-1

TELIC GROUP Product BLAYCO Active electrodes for electrosurgical pencil, non-stick coated TCETl - Ball electrode 80mm, REF. ABT-501 TCET2 - Needle electrode 70mm, REF. ABT-60 TCET3 - Needle electrode 150mm, REF. ABT-70 UlSJ / TCET4 - Blade electrode 70mm, REF. ABT-80 TCET5 - Blade electrode 160mm, REF. ABT-90 Materials ■ Electrode rod: Stainless steel AISI 304 ■ Anti-adherent tip coating: Polytetrafluoroethylene ■ Mould: ABS ■ Protector: PVC ■ Inner pouch (REF ABT-70, ABT-90): PE ■ Package: - Cellulose film, medical grade - Film PET/PE Main dimensional properties ■ Stainless steel rod: 0 2,36 ± 0,02mm ■ Electrode insertion: - Circular: insertion diameter 2,88 ± 0,02mm - Hexagonal, distance between hexagon opposite faces: 3,2 ± 0,02 mm Code Reference Dimensions of active tip (mm) Total length(mm) Electrode insertion Biocompatibility ■ ISO 10993-5: No cytotoxic ■ ISO 10993-10: no sensitizing ■ ISO 10993-10: negative intracutaneous irritation test Conclusion: the product is biocompatible. Sterilization and expiry date ■ Sterile product by ethylene oxide Page 1 of 2 This document is the property of Telic, S.A.U. Transmission to third-parties or non-authorized copies are not allowed.

 Open the catalog to page 1
Active electrodes for electrosurgical pencil-2

TELIC TECHNICAL DATA SHEET GROUP ■ Shelf life: 3 years upon the manufacturing date Storage and maintenance ■ Keep protected from direct sunlight / the outside ■ Keep in a dry place ■ Temperature limitation: 5-40°C ■ Single use product. No maintenance required Packaging ■ Unitary package ■ Service box of 50 units ■ Suitable for all models of Blayco Electrosurgical pencils ■ Latex Free TELIC, S.A.U. guarantees that this product is in conformity with Directive 93/42/EEC and that it has been manufactured following the directives of the Quality Assurance System certified as ISO 13485. This product...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.