TECHNICAL DATA SHEET Product description The vein strippers are a single use medical device sterilized by EtO, intended to use on surgical interventions were stripping a vein is required. The vein stripper could be used on the two most common removal techniques: conventional stripping and invagination strip- Product charasteristics « Ref. VE-022, conventional stripping: No hole. « Ref. VE-025, invagination stripping: One side hole. Technical specifications Materials « Cable, olives, end cap: » Packaging directly in contact with product: » Protective pouch: LDPE « Sterile pouch: o PA/PP film o Cellulose paper, medical grade Main dimensional properties « Cable length: 100 +10cm o Cable diameter: 1.54 — 1.66mm o End cap diameter: 3.25 +0.10mm » Big olive diameter: 15.40 +0,10mm Medium « Little olive diameter: 9.5 #0,10mm This document is the property of Telic, S.A.U. Transmission to third-parties or non-authorized copies are not allowed 1-3
Open the catalog to page 1TECHNICAL DATA SHEET Funcionality tests » Cable — Stopper tensile strength: >25kg » Cable breaking force: >45kg Biocompatibility « No cytotoxic (ISO 10993-5) eNegative intracutaneous irritation test (ISO 10993-10) oNo sensitizing (ISO 10993-10) Sterilitzation and expiry date « Sterile by Ethylene Oxide «Shelf life: 3 years after manufacturing date » Keep protected from direct sunlight / the outside. o Keep ina dry place. » Single use product. No maintenance required « Do not re-sterilize. Conventional stripping : Pouch content: 2 cables, 3 olives (big, medium and small size), 1 handle. 12...
Open the catalog to page 2TECHNICAL DATA SHEET TELIC GROUP Instructions for use « The olive and the handle must be attached in the opposite ends of the cable, once the cable has been introduced in the vein to be removed. « Olive assembly: Insert the cable into the slot. A “click” must be heard to ensure it has been well inserted. Press the stopper, located in the cable end, into the olive so that it is locked. « Handle assembly: Insert the cable into the slot. Press the stopper, located at the end of the cable, into the handle so that it is. Regulatory information TELIC, S.A.U. guarantees that this product is in...
Open the catalog to page 3All Telic catalogs and technical brochures
-
CUp Electrodes
3 Pages
-
Surface electrodes
2 Pages
-
Concentric needles
2 Pages
-
Corkscrew subdermal needles
2 Pages
-
Conductive gel
3 Pages
-
Telic Groupe
27 Pages
-
Blayco dermal curette
2 Pages
-
Blayco biopsy punches
3 Pages
-
Blayco-Pad
2 Pages
-
Skin marker
2 Pages
-
Desfi-Dormo ED-1010
3 Pages