Pre-gelled electrosurgical plates BLAYCO - 1

TECHNICAL DATA SHEET TELIC GROUP Pre-gelled electrosurgical plates BLAYCO > Product description Electrosurgical plates are used as closing element in the circuit constituted together with the active electrode and the electrosurgical unit in electrosurgical interventions. The electrode provides a large contact surface with the patient, compared with the active electrode, that allows reducing the current flow density and minimize the risk of Product charasteristics « This product does not contain natural latex, phthalates or compounds of animal or biologic origin. « This product is in compliance with RoHS standard. « Dual plates are compatible with electrosurgical generators provided with REM monitoring. Weight limitations for the patient established according to plate model are the following: - Adult « See annex of the present specification for further details about cables and their compatibilities. Compatible models for high energies up to 1100mA: 2125 Models compatible with conventional procedure: 2225, 2425, 2500, 2510, 2600, 2700, 2900 Materials « Backing material: Waterproof flexible white PE foam, closed cell, with biocompatible acrylic adhesive. + Conductive area: Aluminium / PET film with acrylic based hydrogel, biocompatible. This document is the property of Telic, S.A.U. Transmission to third-parties or non-authorized copies are not allowed 1-5

Open the catalog to page 1
Pre-gelled electrosurgical plates BLAYCO - 2

TECHNICAL DATA SHEET « Release liner: PET. » Cable (references —C/XX y —C/XX/5): - Staples: Plated brass. - Wire: Copper (10 x 0.14mm) coated with non-toxic PVC. - Inner terminal: Nickel-plated brass. - Plug: Non-toxic PVC. - Pin (cable 4212): Polycarbonate. « Pouch: PET / aluminium / PE film. « Service Box: Carton. Main dimensional properties Contact area: Conductive surface unipolar plate (em?) Conductive surface dual plate (em?) Adult Adult oblong « Hydrogel thickness: 0,69mm gel + aluminium support « Cable length: - REF-C/XX: eo Plug: See drawing in annex. Electrical properties...

Open the catalog to page 2
Pre-gelled electrosurgical plates BLAYCO - 3

TECHNICAL DATA SHEET . Electrical features of the cable: - High frequency leakage current (IEC 60601-2-2, cl. 201.8.8.3.102): OK. - High frequency dielectric strength (IEC 60601-2-2, cl. 201.8.8.3.103): OK. - Mains frequency dielectric strength (IEC 60601-2-2, cl. 201.8.8.3.104): OK. - Conclusion: Requirements established by electrical safety standards EN IEC 60601-2-2:2018 and ANSI/ AAMI Electromagnetic compatibility: The product is classified as group 1, class A product according to EN 55011/CISPR 11. This means that this product does not generate and/or use intentionally radio-frequency...

Open the catalog to page 3
Pre-gelled electrosurgical plates BLAYCO - 4

TECHNICAL DATA SHEET TELIC GROUP Instructions for use « Do not open the envelope until the plate is ready to be used, to avoid the gel drying out. o Select a well vascularised area near the incision. « Do not place on bony protrusions, metallic prosthesis, tattoos or scars. « Avoid if possible the areas with a lot of fat under the skin. « Shave if necessary. ¢ Clean and dry the application area. « In the case of the elderly, smooth out the skin beforehand. « Apply the adhesive face of the plate smoothing and pressing lightly. « Once the intervention is finished, remove the plate gently....

Open the catalog to page 4
Pre-gelled electrosurgical plates BLAYCO - 5

TECHNICAL DATA SHEET Plate with cable ERBE type connector VALLEYLAB type connector VALLEYLAB type connector AMICA type connector This document is the property of Telic, S.A.U. Transmission to third-parties or non-authorized copies are not allowed 5-5

Open the catalog to page 5