Introduction
Thermo Scientific Revco Blood Bank Refrigerators are designed to comply with AABB, ANRC, and FDA standards for blood product safety. This manual provides installation and operation instructions for various models, including undercounter, space saver, and upright models. Key features include a key-operated power and alarm switch, digital temperature display, and alarm systems.
Operating Standards
The refrigerators are intended for indoor use at altitudes up to 2000 meters, with specific humidity and temperature conditions. They should not be connected to GFCI outlets to avoid nuisance tripping.
Safety Precautions
Warnings and cautions are provided to prevent injury and equipment damage. The unit must be properly grounded, and unauthorized repairs will void the warranty. Flammable materials should not be stored in the refrigerator.
Unpacking and Inspection
Upon delivery, inspect the unit for damage. If damage is found, report it immediately and retain packing materials for inspection.
Transport and Storage
Use appropriate equipment like forklifts for moving the refrigerator to avoid injury. For storage, ensure the unit is dry and the door is open.
Intended Use
The refrigerators are designed for storing blood and blood products in sealed bags at temperatures between 2°C and 6°C. They are classified as Class 2 medical devices by the FDA and should only be used by trained personnel.
Installation
Install the unit in a level area with adequate space around it. Avoid direct sunlight and heat sources. Ensure proper wiring and grounding to prevent damage and ensure safety. Specific installation steps include connecting to a dedicated circuit, leveling the unit, and checking door seals.
Control Panel and Operation
The control panel includes features for temperature control and alarm systems. The unit supports automatic defrost and has guidelines for product loading. Temperature and alarm setpoints can be adjusted using the control panel buttons.
Maintenance
Regular cleaning of the drawers, cabinet interior, and condenser is necessary. Condensers should be cleaned every six months. Follow troubleshooting guidelines for any operational issues.
End of Life Care
Instructions are provided for the proper disposal and recycling of the unit at the end of its life cycle, in compliance with local regulations and the WEEE Directive.
Temperature Control
The unit is factory set to maintain a +4°C temperature. Adjustments can be made by accessing service parameters. Temperature is measured using probes in sensor bottles filled with glycerine and water.
Alarm Systems
The alarm system provides visual and audible warnings for power failure and temperature rise, with a battery backup. Units can have local or optional remote alarms, and instructions for installing a remote alarm system are provided.
Chart Recorders
Instructions for preparing and operating the chart recorder, including battery installation and chart paper setup, are provided. The chart recorder uses AC power with a battery backup for power failures.
See more

Catalog excerpts

Revco® Blood Bank Refrigerators-1

Thermo Scientific Revco® Blood Bank Refrigerators Installation and Operation Visit us online to register your warranty www.thermoscientific.com/labwarranty

 Open the catalog to page 1
Revco® Blood Bank Refrigerators-2

IMPORTANT Read this instruction manual. Failure to follow the instructions in this manual can result in damage to the unit, injury to operating personnel, and poor equipment performance. CAUTION All internal adjustments and maintenance must be performed by qualified service personnel. Material in this manual is for informational purposes only. The contents and the product it describes are subject to change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising...

 Open the catalog to page 2
Revco® Blood Bank Refrigerators-5

Blood bank refrigerators are designed to meet strict AABB, ANRC and FDA requirements for stored product safety. This manual provides installation and operation instructions for: Undercounter models (4.7 ft³), Space Saver models (11.5 ft³) and Upright models (23.3, 29.2 and 51.1 ft³). Refer to Table 1 to determine the specifications that apply to your unit. The control system, standard on all models, includes: • Key-operated power and alarm switch Preset temperature setpoint Digital temperature display with 0.1°C resolution Graphic temperature display Audible and visual power failure indicators...

 Open the catalog to page 5
Revco® Blood Bank Refrigerators-6

Table 1: Blood Bank Refrigerator Specifications Model Voltage Amps / Cu.Ft./ Cabinet Codea Breaker Liters Single solid Space Saver Single hinged glass Single Single hinged glass hinged glass W = 230 V / 50 Hz BS1363a Plug Cap A = 115 V / 60 Hz, NEMA 5-15P D = 208-230 V / 60 Hz, NEMA 6-15P Thermo Scientific Recvo Blood Bank Refrigerators

 Open the catalog to page 6
Revco® Blood Bank Refrigerators-7

Safety Precautions In this manual and on labels attached to this product, the words WARNING and CAUTION mean the following: CAUTION: A potentially hazardous situation which, if not avoided, may result in minor or moderate injury or damage to the equipment. WARNING: This symbol indicates potentially hazardous situation which, if not avoided, could result in serious injury or death. WARNING: This symbol indicates situations where dangerous voltages exist and potential for electric shock is present. CAUTION: This symbol, in the context of a CAUTION, indicates a potentially hazardous situation which...

 Open the catalog to page 7
Revco® Blood Bank Refrigerators-8

Safety Precautions Before installing, using or maintaining this product, please be sure to read this manual and product warning labels carefully. Failure to follow these instructions may cause this product to malfunction, which could result in injury or damage. WARNING: Unauthorized repair of your refrigerator will invalidate your warranty. Contact Technical Service at 1-800-438-4851 for additional information. WARNING: No equipment that uses an open flame should be placed inside the refrigerator. This will harm the unit, hamper functionality and compromise your safety. CAUTION: Do not use any...

 Open the catalog to page 8
Revco® Blood Bank Refrigerators-9

At delivery, examine the exterior for physical damage while the carrier’s representative is present. If exterior damage is present, carefully unpack and inspect the unit and all accessories for damage. If there is no exterior damage, unpack and inspect the equipment within five days of delivery. If you find any damage, keep the packing materials and immediately report the damage to the carrier. Do not return goods without written authorization. When submitting a claim for shipping damage, request that the carrier inspect the shipping container and equipment. Thermo Scientific Recvo Blood Bank...

 Open the catalog to page 9
Revco® Blood Bank Refrigerators-10

For transport, do not lift the device using the doors or components attached to the device as lift points which may result in personal injury and damage to the product itself. CAUTION: Heavy loads! Lift only with a forklift! Or at minimum a pallet jack! To avoid injury through physical strain, such as strain trauma and slipped discs, do not attempt to lift the refrigerator without the use of a forklift! Or at minimum a pallet jack! To avoid injury through dropped loads, be sure to wear personal protection equipment, such as safety shoes, when lifting or moving the refrigerator. If the unit is...

 Open the catalog to page 10
Revco® Blood Bank Refrigerators-11

Operating Standards The refrigerators described in this manual are classified for use as stationary equipment in a Pollution Degree 2 and Overvoltage Category II environment. These units are designed to operate under the following environmental conditions: • Maximum relative humidity 60% for temperatures from 15 to 32°C (59 to 90°F). For V models: Maximum relative humidity 75% for temperatures from 10 to 35°C (50 to 95°F). Main supply voltage fluctuations not to exceed 10% of the nominal voltage. The refrigerator must not be connected to a GFCI (Ground Fault Circuit Interrupter) protected outlet...

 Open the catalog to page 11
Revco® Blood Bank Refrigerators-12

The refrigerators described in this manual are intended for the storage of blood and blood products. All materials stored in these products must be in sealed blood bags. Product must not be stored in open containers. These products are intended for storing samples or inventory at operating temperatures between 2°C (35.6°F) and 6°C (42.8°F). Expected users of this equipment include but are not limited to personnel from the following areas: Professional and clinical (blood bank) laboratories, Pharma and Biotech facilities, Industrial and Government facilities or those trained in laboratory protocols...

 Open the catalog to page 12
Revco® Blood Bank Refrigerators-13

7 Installation WARNING: Do not exceed the electrical rating printed on the dataplate located on the upper left side of the unit. CAUTION: Do not move the unit using the drain pan on the back. This could cause damage to the equipment. 7.1 Location WARNING: Do not move the unit while loaded. Unit shall always be moved when empty and by pushing slowly at handle level or lower on the unit. Be especially careful on any uneven surfaces. Install the unit in a level area free from vibration with a minimum of 6 inches of space on the sides and rear and 12 inches at the top. Do not position the equipment...

 Open the catalog to page 13

All Thermo Scientific catalogs and technical brochures

  1. CryoStar NX50

    2  Pages

  2. Multifuge

    58  Pages

  3. MaxQ 6000

    40  Pages

  4. 1300 Series A2

    76  Pages

  5. LTQ XL™

    4  Pages

  6. Furnaces

    2  Pages

  7. native MS

    16  Pages

  8. RM 200 EG

    4  Pages

  9. Darwin LIMS

    2  Pages

  10. Konelab 20

    2  Pages

  11. Indiko

    6  Pages

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.