Overview
The VHP Decontamination Lock is a material airlock designed for rapid decontamination using hydrogen peroxide (H₂O₂) vapor. It is primarily used to transfer materials between rooms with different cleanliness levels, ensuring surface sterilization before entry.

Specifications
  • Dimensions for Lock 1: Width 1900 mm, Height 2640 mm, Depth 1600 mm, Closing door height 2100 mm, Swing door width 800 mm.
  • Dimensions for Lock 2: Width 1150 mm, Height 2795 mm, Depth 840 mm, Closing door height 800 mm, Swing door width 800 mm.
  • Material: Jacket made of stainless steel AISI 304; Pass-through chamber made of AISI 316L with a polished surface finish and rounded corners.
  • Filtration: Inlet and outlet H14 HEPA filtration.

Features
  • Integrated VHP generator for a 6-log reduction in contaminants.
  • Complies with ISO 10648-2 leak tightness classes and EU GMP Standard for 'B' class cleanliness.
  • Electronic control system with Siemens PLC and color touchscreen controls.
  • Mobile perforated shelves for comprehensive sterilization.
  • Operates at normal temperatures (25-30°C) and pressures.
  • Low operating costs with minimal maintenance demands.

Control and Safety
  • HMI Control Panel with display for measured values, operation modes, and user login features.
  • H₂O₂ concentration sensors for safety inside and outside the chamber.
  • Chamber door made of PMMA with a silicone pneumatic seal and monitored by sensors.

Additional Options
  • Paperless recorder for data recording.
  • Distribution pipe for optimal gassing inside the chamber.

Standards and Compliance
The device adheres to relevant standards for safety and efficiency, ensuring reliable operation in controlled environments.
See more

Catalog excerpts

BLOCK Technology_Decontamination Lock_var 062021-1

VHP Decontamination Lock Material airlock providing rapid H₂O₂ decontamination

 Open the catalog to page 1
BLOCK Technology_Decontamination Lock_var 062021-2

Dimensions VHP Decontamination Lock VHP Decontamination Lock Dimensions in mm - VHP Decontamination Lock 1 Width Closing door height Swing door width Dimensions in mm - VHP Decontamination Lock 2 Width Closing door height Swing door width

 Open the catalog to page 2
BLOCK Technology_Decontamination Lock_var 062021-3

Decontamination locks are most frequently used as a pass-through to transfer material between rooms with different classes of cleanliness (typically B and C), whereas material must be decontaminated (by means of surface sterilization) before entry into the room. Loaded materials are decontaminated using hydrogen peroxide vapour with a 6-log reduction. A hydrogen peroxide vapour generator is a part of the equipment. Integrated VHP generator Decontamination pass-through cabin may be fitted onto the wall separating individual rooms Lock complies with the leak tightness classes pursuant to ISO 10648-2...

 Open the catalog to page 3
BLOCK Technology_Decontamination Lock_var 062021-4

Main switch H₂O₂ concentration sensor – high Process sensor inside the chamber. H₂O₂ concentration sensor – low Safety sensor inside the chamber. Chamber closing door 03 It is made of PMMA with a thickness of 25 mm. The door is sealed with a silicone pneumatic seal. A sensor monitors the closing of the door; another sensor controls the mechanism that prevents the door from opening. H₂O₂ concentration sensor – low Safety sensor outside the chamber. Perforated shelves 04 4 pieces, standard perforated shelves for loading material. Compressed air inlet Air exhaust Air inlet Paperless recorder VHP...