TA41 series - 1

Product Segments TiMOTION's TA41 is the ideal linear actuator for medical applications, particularly for dentist chairs and electric wheelchairs. Its physical design is similar to the TA7, yet without the IP rating. The TA41 provides multiple options of cable exits and it supports a maximum of 600Kg force in push. General Features Voltage of motor Maximum load Maximum load Maximum speed at full load Stroke Minimum installation dimension Color Operational temperature range Options 12, 24, or 36V DC 6,000N in push 4,000N in pull 23.4mm/s (with 1,000N in a push condition) > 25~1000mm > Stroke + 163mm Black or grey +5°C~+45°C Hall sensors, Reed sensor

Open the catalog to page 1
TA41 series - 2

Retracted Length Note 1  The above dimension is with motor cover.

Open the catalog to page 2
TA41 series - 3

Load and Speed CODE Load (N) Self Locking Force (N) Typical Current (A) No Load With Load Typical Speed (mm/s) No Load With Load Motor Speed (2600RPM, Duty Cycle 10%) C 5000 4000 Motor Speed (3400RPM, Duty Cycle 10%) L 6000 4000 Motor Speed (3800RPM, Duty Cycle 10%) U 5000 4000 Note 1 Please refer to the approved drawing for the final authentic value. 2 This self-locking force level is reached only when a short circuit is applied on the terminals of the motor. All the TiMOTION control boxes have this feature built-in. 3 The current & speed in table are tested with 24V DC motor. With a 12V...

Open the catalog to page 3
TA41 series - 4

Performance Data (24V DC Motor) Motor Speed (2600RPM, Duty Cycle 10%) Speed vs. Thrust

Open the catalog to page 4
TA41 series - 5

Performance Data (24V DC Motor) Motor Speed (3400RPM, Duty Cycle 10%) Speed vs. Thrust

Open the catalog to page 5
TA41 series - 6

Motor Speed (3800RPM, Duty Cycle 10%) Speed vs. Thrust 40.0 35.0 30.0 Performance Data (24V DC Motor)

Open the catalog to page 6
TA41 series - 7

Load and Speed See page 3 Retracted Length (mm) See page 8 Rear Attachment (mm) See page 9 Front Attachment (mm) See page 9 2=Aluminum casting, U clevis, slot 6.2, depth 17.0, hole 10.2 3=Aluminum casting, U clevis, slot 6.2, depth 17.0, hole 12.2 4=Aluminum casting, U clevis, slot 8.2, depth 17.0, hole 10.2 5=Aluminum casting, U clevis, slot 8.2, depth 17.0, hole 12.2 0 = Without punched hole on inner tube, without slot, M22*2P inner threaded 1 = Punched hole on inner tube + plastic cap, without slot, hole 10.2, with plastic bushing 2 = Punched hole on inner tube + plastic cap, without...

Open the catalog to page 7
TA41 series - 8

Retracted Length (mm) 1. Calculate A+B + C = Y 2. Retracted length needs to > Stroke+Y A. Front Rear Attach. C. Spindle Function N < 6000 (N) Front Attach. 0, 1 2, 3

Open the catalog to page 8
TA41 series - 9

TA41 Ordering Key Appendix Rear Attachment (mm) C = Aluminum casting, U clevis, slot 8.2, depth 17.0, hole 10.2, T-bush 5 = luminum casting, U clevis, slot 8.2, A depth 17.0, hole 12.2 4 = luminum casting, U clevis, slot 8.2, A depth 17.0, hole 10.2 3 = luminum casting, U clevis, slot 6.2, A depth 17.0, hole 12.2 2 = luminum casting, U clevis, slot 6.2, A depth 17.0, hole 10.2 F = luminum CNC, no slot, hole 10.2, A T-bush 3 = lastic, U clevis, slot 8.2, depth 20.2, P hole 10.2, for load push < 4000N & pull < 2500N ø26 2 = unched hole on inner tube + plastic P cap, without slot, hole 12.2...

Open the catalog to page 9
TA41 series - 10

TA41 Ordering Key AppendixFront Attachment (mm) 8 = Aluminum casting, U clevis, slot 6.2, depth 17.0, hole 12.2 9 = Aluminum casting, U clevis, slot 6.2, depth 17.0, hole 10.2, with plastic T-bushing J = Aluminum casting, without slot, hole 10.2, for dental chair Direction of Rear Attachment (Counterclockwise) 1 = 0° 3 = 90° Functions for Limit Switches Wire Definitions CODE Pin 1 (Green) 2 (Red) O 3 (White) 4 (Black) 5 (Yellow) A (Blue) 1 extend (VDC+) N/A N/A N/A retract (VDC+) N/A 2 extend (VDC+) N/A middle switch pin B middle switch pin A retract (VDC+) N/A 3 extend (VDC+) common upper...

Open the catalog to page 10
TA41 series - 11

C = Y cable (For direct cut system, no water proof, anti pull) Cable Length for Direct Cut System (mm) CODE LI L2 L3 F = DIN 6P, 180° plug, for TEC G = Audio plug extension cable standard option M = DIN 4P, dental chair plug (40510-143, standard) N = DIN 4P, dental chair plug (40510-040) Terms of Use The user is responsible for determining the suitability of TiMOTION products for a specific application. TiMOTION products are subject to change without prior notice.

Open the catalog to page 11

All TiMOTION Technology catalogs and technical brochures

  1. ERGO MOTION

    48 Pages

  2. CARE MOTION

    32 Pages

  3. series TL3

    12 Pages

  4. JP4 series

    7 Pages

  5. JP3 series

    7 Pages

  6. TA6 Series

    10 Pages

  7. TA1 Series

    12 Pages

  8. TGM3 Series

    6 Pages

  9. TGM2 Series

    7 Pages

  10. TGM1 series

    7 Pages