Catalog excerpts
Datasheet Dual Channel Isolation Amplifier Dual Channel Isolation Amplifier 95-0450 0450-006-1 The TMSi isolation amplifier is intended to provide a patient safe galvanic he isolation between two external analog inputs and two AUX ports on TMSi signal acquisition devices. NOTE: Keep all patient unsafe connections outside the patient patient-unsafe environment, i.e. 1.5m from the patient. The package set contains a dual channel isolation amplifier as well as two AUX cables and this specifications sheet. specifications Physical Dimensions Size 105mm x 50mm x 25mm Weight 75 gram Material ABS (grey) Electrical Properties Isolation breakdown voltage Gain Input Voltage range Maximum Allowed Input Voltage Leakage current Bandwidth Input Impedance Input referred Noise (RMS) Cross-talk talk Offset Power consumption Input Connectors Output Connectors Pinout Front-outside view, female connector outside
Open the catalog to page 1Maintenance • Before cleaning make sure equipment is turned off and not in contact with a patient. • Use a slightly damp cloth for cleaning. • Never use aggressive chemicals for cleaning. • Only use water or isopropyl alcohol for cleaning. • Do not sterilize equipment. Notes • You need two AUX ports on your TMSi signal acquisition device to connect both external inputs. • When connecting only a single channel the power consumption will be the same as with both channels connected (300mW). • The power consumption is beyond the advised maximum rating of TMSi AUX ports, which is 50mW per...