Perivascular Flowprobe Specifications
2Pages

{{requestButtons}}

Catalog excerpts

Perivascular Flowprobe Specifications - 1

Volume Flow PS-Series, V-Series, PR-Series, PAU-Series & PMP-Series Probes measure bidirectional flow up to 5 times the transonic THE MEASURE OF BETTER RESULTS. 1. Resolution: represents the smallest detectable change in flow, a factor in accuracy. 2. Probes operate in either low flow or normal flow scales, determined by the flow range under study. selected scale setting. The scale settings calibrate the 1 volt reference signal for data collection; the linear range of the Flowmeter is equal to ± 5V. By using the low flow button, measurement sensitivity is increased by a factor of four. This linear overrange is important for the proper recording of highly pulsatile peak flows. 3. Maximum Range for each Probe reflects the highest flow rate that can be processed. 4. Zero offset on individual Probes is often lower than this value. 5. The Absolute Accuracy percentage can be raised to relative accuracy levels by in situ calibration.

Open the catalog to page 1
Perivascular Flowprobe Specifications - 2

Reflector Depth Probe Length Probe Width 1. Lumen is the ultrasonic window within the reflector. Dimensions are approximately equal. 2. Reflector depth is the external dimension for Nanoprobes. 3. Handle comes standard with indicated MA- acute Probes. MC- chronic or custom orders are available without a handle. Probe Length Reflector Depth Reflector Length Shell Length Probe Width Reflector Width PROBE SIZE & SERIES transonic THE MEASURE OF ^ BETTER RESULTS.

Open the catalog to page 2