User Manual CA-C01
1 /7Pages

User Manual CA-C01

User Manual CA-C01
1 /7Pages

Catalog excerpts

User Manual CA-C01-1

TRUE PHANTOMSOLUTIONS Standard Calvaria for X-Ray CT, US, MRI User ManualStandard Calvaria for X-Ray CT, US, MRI[CA-A01] VERSION 2.1 ■ 1 May 2026 [email protected] ■ www.truephantom.com

 Open the catalog to page 1
User Manual CA-C01-2

TRUE PHANTOM SOLUTIONS INC.

 Open the catalog to page 2
User Manual CA-C01-3

TRUE PHANTOM SOLUTIONS INC. Description The design of this phantom is based on an average human male skull, and It is developed using our patented ceramic-reinforced epoxy-based composite material suitable for Ultrasound, MRI, and CT applications (any other part of the skull can be fabricated based on the customer order). In terms of MRI applications, the phantom tissues have realistic T2 relaxation time values, which makes this product the best fit for any T2-weighted MRI imaging methods. Very good results can also be achieved with proton-density imaging methods. The phantom can still be imaged...

 Open the catalog to page 3
User Manual CA-C01-4

TRUE PHANTOM SOLUTIONS INC. Anatomy Calvaria - Cortical Bone - Trabecular Bone Front View Lateral View Perspective View

 Open the catalog to page 4
User Manual CA-C01-5

TRUE PHANTOM SOLUTIONS INC. CA-C01 | User Manual TRUE PHANTOM SOLUTIONS INC. CA-C01 | User Manual Sound Density Velocity [m/s] [g/cm3] Attenuation Speckles Measured at 2.25 MHz [dB/cm] Materials Used Soft tissue and organs: Composition of urethane-based soft resin Synthetic bones: Patented ceramic-reinforced epoxy-based composite material Phantom Size: 6.7 x 5.5 x 2.4 Inches, Weight: 2 Lbs (approx.). Shipment Size: 12 x 10 x 10 Inches, Weight: 6 Lbs (approx.) Standard Calvaria Phantom User Manual/Assembly Instructions 2-Year Warranty & Unlimited Customer Service

 Open the catalog to page 5
User Manual CA-C01-6

TRUE PHANTOM SOLUTIONS INC. CA-C01 | User Manual The phantom is fabricated based on the anatomy of the real human body. It can be scanned per the standard protocols used while positioning a real person. The phantom can be put on horizontal and vertical positions for scanning purposes. Anterior, Lateral, and Posterior techniques can also be performed. • The phantom is made of urethane-based material and should be protected from direct exposure to any intense UV light. • The experiments conducted on the phantom can be performed in a sunny lab, but leaving the phantom under direct sunlight for weeks...

 Open the catalog to page 6
User Manual CA-C01-7

TRUE PHANTOM SOLUTIONS INC. CA-C01 | User Manual • The phantom is water-resistant and can be cleaned after use. It is ideal for cleaning it either with water or with soapy water. It is also quite resistant to chemicals, but if stronger solvents are meant to be used with the phantom, we recommend doing a small test on the bottom part of the neck or legs. The phantom should be placed on its belly, and the chosen solvent can be applied on a small, selected surface of the phantom for a few hours. If there is no visible effect on the phantom's surface, it is safe to assume that the phantom is resistant...

 Open the catalog to page 7

All True Phantom Solutions catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.