User Manual TO-A06

User Manual TO-A06
1 / 18 PagesView full catalog

User Manual TO-A06

Product catalog summary
Overview
The Adult Female Torso Phantom is a medical imaging simulator designed for X-Ray, CT, and MRI. It is developed using CT scans of a healthy 35-year-old patient and features advanced tissue-mimicking materials for realistic imaging results. The phantom includes a heart pump for cardiac motion and a lung pump for breathing motion, enhancing its utility for clinical and research purposes.
Anatomy
The phantom replicates a comprehensive range of anatomical structures including the spine, ribcage, shoulders, clavicles, pelvis, partial femoral bones, trachea, heart, lungs, diaphragm, breasts, liver, gallbladder, stomach, kidneys, spleen, pancreas, intestines, bladder, cervix, vaginal canal, ovaries, and uterus.
Functional Features
  • Beating Heart: The heart model includes cyclic motion, water-fillable chambers, and supports pulsatility from 0-240 BPM with a user-friendly touchscreen interface.
  • Breathable Lungs: The lungs support up to 60 breathing cycles per minute with volume adjustments in 1 mL increments, providing approximately 1 liter of total air volume.
Design and Components
The phantom is designed for comprehensive imaging with multiple views. The package includes the adult torso, secondary tubes, lung pump, and heart pump with accessories.
Usage Instructions
  • Heart and Lung Pumps: Instructions for connecting and disconnecting the pumps are provided, emphasizing proper setup and storage to prevent damage and discoloration.
  • Scanning: The phantom can be scanned in various positions and with different imaging techniques, including ultrasound, MRI, and CT. It supports the use of contrast mediums and provides guidelines for catheter insertions.
  • Contrast Medium/Water: Compatible contrast mediums include Barium-sulfate, Iodine-based, Gadolinium-based, Microbubble, and Iron Oxide-based. Proper flushing and drying procedures are necessary post-scanning.
Technical Properties
Details on sound velocity, density, attenuation, hardness, and T2 values for different tissues are provided, along with thermal properties and HU values.
Materials and Dimensions
The phantom is composed of urethane-based soft resin for tissues and a ceramic-reinforced epoxy-based composite for bones. Specific dimensions and weights are provided, along with shipment details.
Handling and Storage
  • The phantom can be positioned horizontally or vertically for scanning and should be moved using a stretcher or wheeled table.
  • Store in a dry, dark place or cover with plastic to prevent discoloration. Avoid prolonged exposure to direct sunlight.
  • Use and store at temperatures between 10°C and 63°C (50F-145.4F).
Cleaning
The phantom is water-resistant and can be cleaned with water or soapy water. It is resistant to chemicals, but stronger solvents should be tested on a small area first.
FAQs
For additional FAQs, visit: https://www.truephantom.com/faq/
See more

Catalog excerpts

User Manual TO-A06-1

TRUE PHANTOM SOLUTIONS User ManualAdult Female Torso (Dynamic) for X-Ray CT, MRI [TO-A06] VERSION 2.2 ■ 07 August 2026 [email protected] ■ http://www.truephantom.com

 Open the catalog to page 1
User Manual TO-A06-2

TRUE PHANTOM SOLUTIONS INC. TO-A06 | User Manual

 Open the catalog to page 2
User Manual TO-A06-3

TRUE PHANTOM SOLUTIONS INC. 1. Description The Adult Female Torso Phantom is a medical imaging simulator, uniquely developed using CT scans of a healthy 35-year-old patient (168 cm in height), offering unparalleled anatomical accuracy. Designed for X-Ray, CT, and MRI, it features advanced tissue-mimicking materials that closely replicate human soft tissue and bone. Developed in collaboration with experts from the U.S. and Switzerland, this phantom is a versatile tool for a wide range of clinical and research purposes. Its advanced tissue-mimicking materials accurately replicate soft tissue and...

 Open the catalog to page 3
User Manual TO-A06-4

TRUE PHANTOM SOLUTIONS INC. TO-A06 | User Manual Complete Spine Complete Ribcage Shoulders and Clavicles Pelvis Partial Femoral Bones Trachea Realistic Heart with: - 2 Atria, 2 Ventricles -Appendage Feature -Aorta - Flexible Valvular Structures -SVC - Esophagus Lungs Diaphragm Breasts: - Body Tissue Mimicking Material - Mammary Gland Tissues Liver - Bile Ducts Gallbladder Stomach Kidneys Spleen Pancreas Large and Small Intestines Vaginal Canal ✓ Beating heart model with the cyclic motion of the heart walls Water-fillable chambers with minimal air pockets Access to intravenous catheter insertion...

 Open the catalog to page 4
User Manual TO-A06-5

TRUE PHANTOM SOLUTIONS INC. Front View Side View View Back View Perspective View

 Open the catalog to page 5
User Manual TO-A06-6

TRUE PHANTOM SOLUTIONS INC. 6. Components Included Adult Torso Secondary Tubes for connecting the primary tubing to the Heart/Water Tank Lung Pump and accessories (Details in the Lung Pump Manual) Heart Pump and accessories (Details in the Heart Pump Manual)

 Open the catalog to page 6
User Manual TO-A06-7

TRUE PHANTOM SOLUTIONS INC. 7. Using a Heart Pump with the Phantom 7.1 Connecting the Heart Pump   Lay the secondary tubes on either side of the phantom. The tubing has been cut to specific length so that it dictates the arrangement of the tubing and connectors. If you have a custom order with the optional coronary arteries, the necessary branching/valving for the tubing has already been implemented. Secondary Tubes for phantom without Coronary Arteries Secondary Tubes for phantom with Coronary Arteries Attach the secondary tubes to the Phantom using the quick connect tabs provided. Attach...

 Open the catalog to page 7
User Manual TO-A06-8

TRUE PHANTOM SOLUTIONS INC. TO-A06 | User Manual The overall setup of the Phantom’s connection with the Heart Pump and Reservoir should be as shown below. 7.2 Disconnecting the Heart Pump and Storing the Phantom ■ To disconnect the tubes from the phantom, gently press the quick connect adapters and pull the extension tube out. ■ Tilt the phantom to allow the water to flow out. ■ Make sure the phantom is dry before storing it. Storing water in the phantom for long periods of time (over 2 weeks) can result in discoloration.

 Open the catalog to page 8
User Manual TO-A06-9

TRUE PHANTOM SOLUTIONS INC. TO-A06 | User Manual ■ Lay the extension tube near the top of the phantom. ■ Attach the tube to the phantom using the quick connect tab provided (The trachea opening). Lung pump _ connection point ■ Please refer to the Lung Pump's Manual for Details about how to connect the extension tubing to the pump and how to operate the pump. ■ This single outlet will be used to simulate the "inhaling" and "exhaling" motions of the lungs inside.

 Open the catalog to page 9
User Manual TO-A06-10

TRUE PHANTOM SOLUTIONS INC. The overall setup of the Phantom’s connection with the Lung Pump should be as shown below 8.2 Disconnecting the Pump and Storing the Phantom  To disconnect the tubes from the phantom, gently press the quick connect adapters and pull the extension tube out. Do not pull on the blue tubes as that could damage the phantom.

 Open the catalog to page 10
User Manual TO-A06-11

TRUE PHANTOM SOLUTIONS INC. TO-A06 | User Manual The complete assembled phantom should have the following configuration HEART PUMP AND RESERVOIR 11 | Page

 Open the catalog to page 11
User Manual TO-A06-12

TRUE PHANTOM SOLUTIONS INC. TO-A06 | User Manual ■ The phantom is fabricated based on the anatomy of a real human, and it can be scanned in supine, prone, and lateral positions; more importantly, scanning can be performed from multiple directions. ■ For ultrasound applications, this phantom can be scanned either with or without the ultrasonic gel. To get clearer images, we recommend applying a little gel to the area before scanning. You can simply wipe off the excess with a soft cloth or tissue once done. ■ You can scan the phantom while the pumps are running to get imaging results of the breathing...

 Open the catalog to page 12
User Manual TO-A06-13

TRUE PHANTOM SOLUTIONS INC. TO-A06 | User Manual Attenuation Measured at 2.25 MHz [dB/cm] Organs with Organs without Speckles

 Open the catalog to page 13
User Manual TO-A06-14

TRUE PHANTOM SOLUTIONS INC. TO-A06 | User Manual 12.1 Thermal Properties of Bone-Mimicking Tissue Thermal Volumetric Specific Thermal Conductivity Heat Capacity Diffusivity Specific Heat 12.2 HU Values of the Tissue Mimicking Materials S.No. Tissue Type HU Value (average) Vena Cava 80 (Tissue Part), -1000 (Air Filled Part)

 Open the catalog to page 14
User Manual TO-A06-15

TRUE PHANTOM SOLUTIONS INC. Mammary Gland Tissues Vaginal Canal Rectum Wall Sigmoid Colon Wall Large and Small Intestines 80 (Tissue Part), -1000 (Air Filled Part)

 Open the catalog to page 15
User Manual TO-A06-16

TRUE PHANTOM SOLUTIONS INC. TO-A06 | User Manual Soft tissue and organs: Composition of urethane-based soft resin Synthetic bones: Patented ceramic-reinforced epoxy-based composite material 14. Size and Weight of the Phantom Phantom Size: 74 x 35 x 28 cm (29 x 14 x 11 Inches), Weight: 36 kg (80 lbs) (approx.) Lung Pump Size: Size: 36 x 12 x 8 Inches, Weight: 48 Lbs (approx.) Heart Pump Size: Size: 27.5 x 8 x 10 Inches, Weight: 39 Lbs (approx.) Shipment Size with Hard Carry Case: 46 x 36 x 17 Inches, Weight: 136 Lbs (approx.) Size of Lung Pump Package: 49 x 17 x 13 Inches, Weight: 62.5 Lbs (approx.)...

 Open the catalog to page 16
User Manual TO-A06-17

TRUE PHANTOM SOLUTIONS INC. TO-A06 | User Manual 17.Storage • The phantom is fabricated from the CT scans of a real human body. It can be scanned per the standard protocols used while positioning a real person. • The phantom can be put on horizontal and vertical positions for scanning purposes. • Anterior, Lateral, and Posterior techniques can also be performed. • When moving the phantom to another location, use a stretcher or wheeled table instead of carrying it. • The experiments conducted on the phantom can be performed in a sunny lab, but leaving the phantom under direct sunlight for weeks...