MD490M

MD490M

MD490M

Product catalog summary
Applications
The MD490M Helium Leak Detector is designed for use in air conditioning and refrigeration systems, automotive components, and appliances.
Benefits
The MD490M is a core component of VIC's turnkey vacuum mass spectrometer systems, known for their ease of service and adaptability. It utilizes a helium mass spectrometer, which is at least 1000 times more sensitive than other leak detection methods. The system is reliable, rugged, easy to maintain, and portable, with many units operating for over 20 years. It features fast response times and password protection to prevent unauthorized access.
Features
The system includes a display leak rate range, power specifications (115 VAC/60 Hz, 4 Amp), and dimensions (13-9/16”W, 24-5/16”D, 15-9/16”H). It has a high vacuum rated solenoid actuated fore-line isolation valve and pressure measurement capabilities. The high vacuum pump is a 53 c/second air-cooled turbo-molecular type, and the mechanical pump is an external forepump with a 7 CFM capacity. The system mode testing outputs 0-10 Volts to the main PLC, with interface signals for auto-zero, ready, leak rate, and fault signals. Sniffer mode testing ranges from 10-3 to 10-7 atm-cc/sec He.
Performance
The system mode has a minimum detectable helium leak rate of 2.5 x 10-9 atm-cc/sec, while the sniffer mode detects down to 2.5 x 10-7 atm-cc/sec. Response times are less than 1.0 second for system mode and less than 1.5 seconds for sniffer mode. Startup times are less than 3 minutes for cold start-up and less than 1 minute for standby start-up. The system features one-button automatic operation, a hard vacuum leak detection system, PLC control, modular assembly, and turn-key operation. It includes a front panel key lock-out, a five-year warranty on major components, adjustable turbo speed, and an internal fast clean-up sniffer manifold.
Available Options
Options include 24Vdc control, 110Vdc control, an atmospheric sniffer module, side or bottom mount configurations, and an electrical converter box.
Contact Information
Vacuum Instruments Corporation, 2101 9th Ave, Ronkonkoma, NY 11779, United States. Phone: +1 (631) -737-0900, 24 hr Service Hotline: +1 (800) 221-2440, Email: [email protected].
See more

Catalog excerpts

MD490M-1

— VACUUM INSTRUMENTS ■ CORPORATION t Leak Detection Solutions Leak Detection Solutions The MD-490M is at the heart of VIC's turnkey vacuum mass spectrometer systems which feature ease of service and its adaptability for system use. Our Leak Detection Systems use the most sensitive and accurate leak detection device science has developed -- the helium mass spectrometer. The helium mass spectrometer is at least 1000 times more sensitive than other leak detection methods. V.I.C. Leak Detection has taken the helium mass spectrometer out of the laboratory, where it has served as a research tool for over 30 years, and redesigned it to meet the stringent requirements of ruggedness, reliability and ease of maintenance that are essential in any industrial environment. -»Industry's most sensitive and accurate leak detection -* Reliable -»Rugged with many systems operating for 20 years Easy to maintain -»Fast response time -> Portable Password Protection prevents unauthorized access -»Helium mass spectrometer is a least 1000 times more sensitive than other leak detection methods APPLICATIONS VACUUM INSTRUMENTS CORPORATION Leak Detection Solutions Air Conditioning & Refrigeration Systems, Automotive Components, Appliances

 Open the catalog to page 1
MD490M-2

Performance • Sensitivity System Mode, minimum detectable helium leak rate 2.5 x 10-9 atm-cc/sec. Sniffer Mode, minimum detectable helium leak rate 2.5 x 10-7 atm- cc/sec. • Response Time < 1.0 second system mode testing, < 1.5 seconds sniffer mode testing. • Startup Time One button automatic operation. Cold Start-Up, less than 3 minutes. Standby Start-Up, less than 1 minute. Display Leak Rate Range • System Mode Testing: 0-10 Volts output to main PLC. Leak rate configured to systems calibrated gas leak. • Interface Signals: Auto-Zero, Ready, Leak Rate and Fault Signals. • Sniffer Mode Testing:...

 Open the catalog to page 2