1. Catalogs
  2. VAREX Imaging
  3. PaxScan® 3030DX
video corpo

PaxScan® 3030DX

PaxScan® 3030DX

PaxScan® 3030DX

Product catalog summary
Product Description
The PaxScan® 3030DX is a digital X-ray imaging device known as a flat panel detector (FPD). It features a 30 x 30 cm amorphous silicon FPD with 194µm pixels and a universal power supply. It offers excellent low-dose performance through Varex Imaging’s proprietary electronics and a custom Cesium Iodide scintillator. A Windows® application and DLL library are available for OEM system integration. This device is intended for integration into a complete X-ray system by qualified manufacturers.
Technical Specifications
  • Receptor Type: Amorphous Silicon
  • Conversion Screen: Integral columnar CsI:TI
  • Pixel Area: 298mm x 298mm
  • Pixel Matrix: 1,536 x 1,536
  • Pixel Pitch: 194 µm
  • Limiting Resolution: 2.58 lp/mm @ 15 fps
  • Image Quality (MTF): 55% at 1.0 lp/mm, 25% at 2.0 lp/mm
  • DQE: 77% at 0 lp/mm
  • Quantum Limited Dose: 3.5 nGy/frame
  • Energy Range: 40 - 150 kVp
  • Fill Factor: 68%
  • Lag: <5% (first frame)
  • Scan Method: Parallel
  • Data Output: Gigabit Ethernet
  • A/D Conversion: 16-bit
  • Radiation Tolerance: 2000 Gy
  • Dynamic Range: 94 dB
Power Requirements
  • Input Voltage Range: 21 V - 33 V
  • Nominal Power Consumption: 20 W
  • Peak Power Consumption: 25 W
Mechanical
  • Weight: Approx. 5.6 kg (panel), 9.05 kg (with Tungsten)
  • Housing Material: Aluminum
  • Sensor Protection: Carbon fiber and aluminum
  • Mounting Provisions: Blind, threaded mounting holes
Image Acquisition Modes
  • Normal Fluoro: 768 x 768 (2x2 binned)
  • Full Resolution: 1,536 x 1,536
Environmental
  • Temperature Limit: 15 - 53°C
  • Relative Humidity: 10% - 90%
  • Atmospheric Pressure: 70 kPa - 106 kPa
  • Shock Tolerance: 20G
Regulatory
  • U.S.: ANSI/AAMI ES60601-1:2005
  • Canada: CAN/CSA C22.2 No. 60601-1:08
  • EU: IEC/EN 60601-1:2005
Additional Notes
All Varex Imaging Amorphous Silicon Receptors are designed for integration into complete X-ray systems by qualified integrators, who are responsible for obtaining FDA clearance for medical use.
See more

Catalog excerpts

PaxScan® 3030DX-1

PaxScan® 3030DX Flat Panel Detector Product Description The PaxScan 3030DX is a real-time digital X-ray imaging device commonly referred to as a flat panel detector (FPD). The main system components are the 30 x 30cm 194µm-pixel amorphous silicon FPD and universal power supply. Excellent low-dose performance is achieved by combining Varex Imaging’s proprietary readout electronics with a custom Cesium Iodide scintillator. A Windows® based application program and a communications command (DLL) library has also been developed to assist OEM customers tasked with developing their own system interface. This imager is intended for incorporation into a complete X-ray system by a qualified equipment manufacturer. ® Technical Specifications Power Requirements Receptor Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Amorphous Silicon Input voltage range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 V - 33 V (measured at the input of the imager) Conversion Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Integral columnar CsI:TI Pixel Area - Total . . . . . . . . . . . . . . . . 298mm (h) x 298mm (v) (11.7 x 11.7 in) Peak Power Consumption1 (initialization) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 W Pixel Matrix -Total . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1,536 (h) x 1,536 (v) Effective . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1,516 (h) x 1,516 (v) Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . approx. 5.6 kg panel with Tungsten - 9.05 kg Limiting Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.58 lp/mm @ 15 fps (1 x 1) 1.29 lp/mm @ 30 fps (2 x 2) Housing Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Aluminum Image Quality (RQA5) (typical) Sensor Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Carbon fiber and aluminum Mounting Provisions . . . . . . . . . . . . Blind, threaded mounting holes on the back. Image Acquisition Modes Normal Fluoro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 768 (h) x 768 (v) (2x2 binned) 2.58 lp/mm (Nyquist Frequency) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16% Temperature Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 - 53°C (as reported by imager internal sensor) Relative Humidity (non-condensing) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10% - 90% Quantum Limited Dose (2x2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5 nGy/frame Shock Tolerance . . . . . . . . . . . . . . . . . . . . . . 20G (any direction no power applied) Scan Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Parallel Data Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Gigabit Ethernet Note1 - Power drop across supply cables is not included Radiation Tolerance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2000 Gy (active area) PaxScan is a Registered Trademark of Varex Imaging Corporation Dynamic Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 dB std modes Copyright © 2018, Varex Imaging Corporation. All Rights Reserv

 Open the catalog to page 1
PaxScan® 3030DX-2

PaxScan® 3030DX Flat Panel Detector Drawing for reference only Inches [mm] Manufactured by Varex Imaging Corporation Specifications subject to change without notice. Salt Lake City, UT 1-801-972-5000 Charleston, SC 1-843-767-3005 www.vareximaging.com Note: All Varex Imaging Amorphous Silicon Receptors are designed to be integrated into a complete X-ray system by a qualified system integrator. The system Integrator is responsible for obtaining FDA clearance for medical use. Copyright © 2018, Varex Imaging Corporation. All Rights Reserved.

 Open the catalog to page 2

All VAREX Imaging catalogs and technical brochures

  1. 4336W

    2  Pages

  2. 4336X

    4  Pages

  3. 2530W

    4  Pages

Archived catalogs

  1. 4343CB

    2  Pages

  2. GS-3074

    8  Pages

  3. MCS-6074

    8  Pages

  4. 2530C

    2  Pages

  5. Linatron K15

    2  Pages

  6. B-130HM

    6  Pages

  7. B-135H

    6  Pages

  8. B-130H

    6  Pages

  9. B-121

    6  Pages

  10. B-115

    6  Pages

  11. B-110

    6  Pages

  12. B-112

    6  Pages

  13. B-100

    6  Pages

  14. B-147

    6  Pages

  15. B-148H

    6  Pages

  16. B-146H

    6  Pages

  17. B-146HM

    6  Pages

  18. B-145W

    6  Pages

  19. B-145A

    6  Pages

  20. B-185H

    6  Pages

  21. B-180H

    6  Pages

  22. B-165H

    6  Pages

  23. 4343R ver3

    2  Pages

  24. 4336R

    2  Pages

  25. AMX4

    4  Pages

  26. OPTICA™ 40

    2  Pages

  27. Linatron® M3

    4  Pages

  28. GS-2079

    8  Pages

  29. GS-2077

    8  Pages

  30. GS-1584

    8  Pages

  31. GS-1587

    8  Pages

  32. Nexus DRF

    2  Pages

  33. Nexus DR

    2  Pages

  34. Attila

    2  Pages

  35. 3024M

    2  Pages

  36. XRpad 4343 F

    2  Pages

  37. XRD 4343RF

    2  Pages

  38. XRpad2 4336

    2  Pages

  39. PaxScan 2020X

    2  Pages

  40. Nexus DR

    2  Pages

  41. Optica™ 30

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.