PaxScan® 4030X-1

PaxScan® 4030X Flat Panel Detector Product Description The PaxScan® 4030X is a real-time digital X-ray imaging device commonly referred to as a flat panel detector (FPD). The main system components are the 40 x 30cm 194m-pixel amorphous silicon FPD and Command Processor CP2LC. Excellent lowdose performance is achieved by combining Varex Imaging’s proprietary readout electronics with a custom Cesium Iodide scintillator. A Windows® based application program and a communications command (DLL) library has also been developed to assist OEM customers tasked with developing their own system interface. This imager is intended for incorporation into a complete X-ray system by a qualified equipment manufacturer. Technical Specifications Power Requirements Receptor Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . Amorphous Silicon Input voltage range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-33 Volts (measured at the input of the imager) Conversion Screen . . . . . . . . . . . . . . . . . . . . Integral columnar CsI:Tl Nominal Power Consumption1 . . . . . . . . . . . . . . . . . . . . . . . 25 Watts Pixel Area - Total . . . . . . . . 397mm (h) x 298mm (v) (15.6 x 11.7 in) Peak Power Consumption1 (initialization) . . . . . . . . . . . . . . 30 Watts Pixel Matrix - Total . . . . . . . . . . . . . . . . . . . . . . . 2,048 (h) x 1,536 (v) Effective . . . . . . . . . . . . . . . . . . . . 2,028 (h) x 1,516 (v) Limiting Resolution . . . . . . . . . . . . . . . . . 2.58 lp/mm @ 15 fps (1 x 1) 1.29 lp/mm @ 30 fps (2 x 2) Housing Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Aluminum Sensor Protection . . . . . . . . . . . . . . . . . Carbon fiber and aluminum Image Quality (RQA5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (typical) Mounting Provisions . . . . . . . . . . . . Blind, threaded mounting holes on the back. Image Acquisition Modes Normal Fluoro . . . . . . . . . . . . . . . . . 1024 (h) x 768 (v) (2x2 binned) Temperature Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 - 49°C (as reported by imager internal sensor) Full Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . 2,048 (h) x 1,536 (v) Environmental Energy Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 - 150 kVp Relative Humidity . . . . . . . . . . . . . . . . . . 10 - 90% Non-Condensing Fill Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68% Atmospheric Pressure . . . . . . . . . . . . . . . . . . . . . . . 70 kPa - 106 kPa Lag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <5% (first frame) Shock Tolerance . . . . . . . . . . 20G (any direction no power applied) Scan Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Parallel Regulatory Data Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . LVDS, CameraLink U.S. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ANSI/AAMI ES60601-1:2005 A/D Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-bit Canada . . . . . . . . . . . . . . . . . . . . . . . . . . . CAN/CSA C22.2 No. 60601-1:08 Cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Passive EU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IEC/EN 60601-1:2005 Radiation Tolerance . . . . . . . . . . . . . . . . . . . . 2000 Gy (active area) Dynamic Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 dB std modes 108 dB DGS mode Note 1 Power drop across supply cables not included ®PaxScan is a Registered Trademark of Varex Imaging Corporation Copyright © 2017, Varex Imaging Corporation. All

 Open the catalog to page 1
PaxScan® 4030X-2

PaxScan® 4030XFlat Panel Detector 8X REAR MOUNTING HOLES POWER CONN DATA CONN. Drawings are for reference only Dimensions are in inches [mm] 5.98 ACTIVE AREA THREAD DEPTH IMAGING PLANE ZOOM MODE DIRECTION ACTIVE AREA Manufactured by Varex Imaging Corporation Specifications subject to change without notice. Note: All Varex Imaging Amorphous Silicon Receptors are designed to be integrated into a complete X-ray system by a qualified system integrator. The system Integrator is responsible for obtaining FDA clearance for medical use. Copyright © 2017, Varex Imaging Corporation. All Rights Reserved....

 Open the catalog to page 2

All VAREX Imaging catalogs and technical brochures

  1. 4336W

    2  Pages

  2. 4336X

    4  Pages

  3. 2530W

    4  Pages

Archived catalogs

  1. 4343CB

    2  Pages

  2. GS-3074

    8  Pages

  3. MCS-6074

    8  Pages

  4. 2530C

    2  Pages

  5. Linatron K15

    2  Pages

  6. B-130HM

    6  Pages

  7. B-135H

    6  Pages

  8. B-130H

    6  Pages

  9. B-121

    6  Pages

  10. B-115

    6  Pages

  11. B-110

    6  Pages

  12. B-112

    6  Pages

  13. B-100

    6  Pages

  14. B-147

    6  Pages

  15. B-148H

    6  Pages

  16. B-146H

    6  Pages

  17. B-146HM

    6  Pages

  18. B-145W

    6  Pages

  19. B-145A

    6  Pages

  20. B-185H

    6  Pages

  21. B-180H

    6  Pages

  22. B-165H

    6  Pages

  23. 4343R ver3

    2  Pages

  24. 4336R

    2  Pages

  25. AMX4

    4  Pages

  26. OPTICA™ 40

    2  Pages

  27. Linatron® M3

    4  Pages

  28. GS-2079

    8  Pages

  29. GS-2077

    8  Pages

  30. GS-1584

    8  Pages

  31. GS-1587

    8  Pages

  32. Nexus DRF

    2  Pages

  33. Nexus DR

    2  Pages

  34. Attila

    2  Pages

  35. 3024M

    2  Pages

  36. XRpad 4343 F

    2  Pages

  37. XRD 4343RF

    2  Pages

  38. XRpad2 4336

    2  Pages

  39. PaxScan 2020X

    2  Pages

  40. Nexus DR

    2  Pages

  41. Optica™ 30

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.