Catalog excerpts

X-RAY PRODUCTS Rotating Anode X-Ray Tube Tubes Radiogénes à Anode Tournante Röntgenröhre mit rotierender Anode Tubos de Rayos-X con Ánodo Giratorio Note: Document originally drafted in the English language. Product Description The M-171 is a 2.8” (71 mm) 39 kV, 222 kJ (300 kHU) maximum anode heat content, rotating anode insert. This insert is specifically designed for use in Mammography systems. The insert features a 0° and 10° biangular molybdenum alloy target and is available with the following nominal focal spots: Description du Produit Le tube M-171, est un tube à anode tournante de 71 mm, (2,8 pouces) de diamètre; sa dissipation thermique calorifique maxima est de 222 kJ, soit 300 kUC et a une puissance de 39 kV. L’anode est particulicèrement adaptée à la mammographie; la pente de l’anode est double: 0° et 10° et composée de molybdène; le tube a des foyers de: 0.1 x 0.05 (0°) - 0.3 (10°) 0,1 x 0,05 (0°) - 0,3 (10°) Nominal Anode Input Power: Small - 800 W IEC 60613 Large - 3.6 kW IEC 60613 For the equivalent anode input power of 60 Watts Loading Factor for slit focal: Small - 26 kV, 25 mA 5° Tilt from Central Ray Large - 28 kV, 100 mA on Central Ray This insert is intended for use in Varian B-110 housings. Puissance anodique nominale de l’anode: Petit foyer - 800 W CEI 60613 Grand foyer - 3.6 kW CEI 60613 Pour la puissance anodique d’equilibre thermique de 60 Watts Facteur de charge pour foyer à fente: Petit - 26 kV, 25 mA 5° Incliner de Rayon Central Grand - 28 kV, 100 mA sur Rayon Central Ce tube est destiné à être inséré dans les gaines Varian B-110. Manufactured by Varian Medical Systems Fabrique par Varian Medical Systems Hergestellt von Varian Medical Systems Fabricado por Varian Medical Systems Produktbeschreibung Die M-171 ist eine 71 mm (2.8”) Doppelfokus DrehanodenRöntgenröhre, mit einer Anoden Wärmespeicherkapazität von 222 kJ (300 kHU) und einer max. Spannungsfestigkeit von 39 kV. Diese Röhre findet ihren speziellen Einsatz in Mammographie Röntgensystemen. Der Molybdän Anodenteller besitzt einen Doppelwinkel von 0°, bzw. 10°. Folgende Brennfleckkombination sind lieferbar: 0.1 x 0.05 (0°) - 0.3 (10°) Nominale Anodenbezugsleistung: Klein - 800 W IEC 60613 Gross - 3.6 kW IEC 60613 Gilt bei einer Aquivalent Anodenleistung von 60 Watt Ladefaktor: Klein - 26 kV, 25 mA 5° Neigung aus Zentralstrahl Gross - 28 kV, 100 mA auf Zentralstrahl Die Röntgenröhre ist für den Einbau in die Varian Strahlerhaube B-110 vorgesehen. Descripcion del Producto El M-171 es un tubo de ánodo giratorio de 71 mm (2.8”), 39 kV, 222 kJ (300 kUC) diseñado especificamente para uso en el sistema de mamografia. Consta de un objectivo aleado de molibdeno se dos grados (0° y 10°) biangulares. Disponible con las siguientes combinaciones de marcas focales: 0.1 x 0.05 (0°) - 0.3 (10°) Potencia nominal de entrada del anodo: Foco fine - 800 W IEC 60613 Foco grueso - 3.6 kW IEC 60613 Para una potencia equivalente del anodo de 60 Watts Carga Electrica Para la Abertura Focal: Pequeño - 26 kV, 25 mA 5° Ladeo de Rayo Central Grande - 28 kV, 100 mA en Rayo Central Este tubo es diseñado, para uso en los encajes Varian de la B-110. Specifications subject to change without notice. Spécifications susceptibles d’être modifiées sans préavis. Technische Daten ohne Gewähr. Especificaciones sujetas a cambio sin previo aviso.
Open the catalog to page 1
medical systems Tube Outline Drawing Dessin d'Encombrement de la Tube Esquema Detallado del Tubos REFERENCE AXIS RAYON CENTRAL RAYO CENTRAL Large - Black Grand - Noir Gross - Schwarz Largo - Negro Small -White Petit - Blanc Klein - Weiss Pequeno - Blanco X-Ray Tube Tube Radiogene Radiation Filter or Filtration Copyright © 2007, Varian Medical Systems. All Rights Reserved.
Open the catalog to page 2
medical systems Single Load Ratings IEC 60613 Abaques de Chargepour Pose Unique CEI 60613 Diagramas de Exposition Radiografica IEC 60613 Copyright © 2007, Varian Medical Systems. All Rights Reserved.
Open the catalog to page 3
medical systems Single Load Ratings I EC 60613 Abaques de Chargepour Pose Unique CEI 60613 Diagramas de Exposition Radiografica IEC 60613 NOMINAL FOCAL SPOT SIZE - 0.1 MAXIMUM EXPOSURE TIME (SECONDS) NOMINAL FOCAL SPOT SIZE - 0.3 MAXIMUM EXPOSURE TIME (SECONDS) Copyright © 2007, Varian Medical Systems. All Rights Reserved.
Open the catalog to page 4
X-RAY PRODUCTS Filament Emission Charts IEC 60613 Abaques d’ Émissions des Filaments CEI 60613 Heizfadenemissionsdiagramm IEC 60613 Curvas de Emisión de los Filamentos IEC 60613 THREE PHASE EMISSION (± .15 A) M-171 THREE PHASE EMISSION (± .15 A) M-171 When using these emission curves for trial exposures, refer to the power rating curves shown for maximum kV, tube emission, filament current, exposure time, and target speed. Lors de l’utilisation de ces abaques pour des expositions d’essai, référez-vous aux courbes maximales de kV, d’émission du filament, de temps d’exposition et de vitesse...
Open the catalog to page 5
medical systems Anode Heating & Cooling Chart Abaques d' Echauffement et de Refroidissement de L'Anode Anoden Aufheiz - und Abkiihl Kurven Curvas de Calentamiento y Enfriamiento del Anodo kJ % ANODE HEATING AND COOLING CURVES Time (Minutes) Duree (Minutes) Zeit (Minuten) Tiempo (Minutos) medical systems Salt Lake City, UT Copyright © 2007, Varian Medical Systems. All Rights Reserved.
Open the catalog to page 6All Varian Oncology catalogs and technical brochures
-
Halcyon
15 Pages
-
Eclipse
7 Pages
-
TrueBeam Safety Features
2 Pages
-
TrueBeam Brochure
11 Pages
-
ProBeam Brochure
16 Pages
Archived catalogs
-
Brachytherapy Applicator Catalog
150 Pages
-
GammaMedplus™ iX Afterloader
4 Pages
-
Trilogy Brochure
7 Pages
-
VitalBeam
8 Pages
-
HyperArc Overview
2 Pages
-
Eclipse™
2 Pages
-
INTRACRANIAL SRS PACKAGE
4 Pages
-
ProBeam 360
2 Pages
-
VELOCITY SOFTWARE
3 Pages
-
ProBeam Compact Brochure
9 Pages
-
oNCOLOGY CARE MANAGEMENT
3 Pages
-
360 ONCOLOGY
3 Pages
-
Proton Therapy reatment
4 Pages
-
HALCYON
12 Pages
-
PaxScan ® 3030X
2 Pages
-
PaxScan ® 4336X
4 Pages
-
PaxScan ® 4343R
2 Pages
-
Nexus-DRF™
2 Pages
-
2520DX
2 Pages
-
1515DXT
2 Pages
-
PaxScan 1508 DXT
2 Pages
-
PaxScan 1313DXT
2 Pages
-
1308DXT
2 Pages
-
Calypso Product Brief
2 Pages
-
Clinac Brochure
12 Pages
-
Capri? Applicator Set
2 Pages
-
Edge Brochure
22 Pages
-
NDI-452
4 Pages
-
M-101G
6 Pages
-
M-105SP
6 Pages
-
M-109
6 Pages
-
M-113
6 Pages
-
M-143
6 Pages
-
M-147SP
6 Pages
-
M-146
6 Pages
-
M-147
6 Pages
-
RAD-85S
6 Pages
-
M-149
6 Pages
-
M-151
10 Pages
-
M-152
12 Pages
-
M-153
8 Pages
-
PaxScan 3030+
2 Pages
-
PaxScan 2520E+
2 Pages
-
PaxScan 4030E
2 Pages
-
PaxScan 4336R
2 Pages
-
PaxScan® 2520XI/V
2 Pages
-
PaxScan® 2520D/CL
2 Pages
-
PaxScan® 2020X
2 Pages
-
PaxScan® 2020+
2 Pages
-
1313
2 Pages
-
PaxScan® 1308
2 Pages
-
Vitesse 3.0
3 Pages
-
BrachyVision
7 Pages
-
RapidArc
8 Pages
-
Varian Exact© IGRT couch
2 Pages
-
ARIA Oncology Information System
12 Pages
-
PaxScan ® 4030CB
2 Pages
-
PaxScan ® 1508DXT
2 Pages
-
PaxScan ® 1313DX
2 Pages
-
PaxScan ® 1308DX
2 Pages
-
Vitesse? 2.5
2 Pages
-
VariSeed? 8.0 Brochure
7 Pages
-
BrachyVision Acuros
3 Pages
-
BrachyVision? 8.5
7 Pages
-
Varian BrachyTherapy
147 Pages