Product Description: The PaxScan 2520D/CL is a third-generation X-ray imager designed for high-speed dental cone-beam CT imaging. It utilizes a Gigabit Ethernet interface or a customer-specified CameraLink standard for instantaneous image display on a user-supplied workstation. The device features a lightweight aluminum housing.
Technical Specifications:
  • Receptor Type: Amorphous Silicon
  • Conversion Screen: CsI, DRZ Plus, or Gd2O2S: Tb (Kodak Lanex Screen)
  • Pixel Area: Total: 19.5 x 24.4 cm; Active (Gd2O2S): 19.3 x 24.2 cm; (Csl, Direct Deposit): 19.0 x 23.8 cm
  • Pixel Matrix: Total: 1,536 x 1,920; Active (Gd2O2S): 1,516 x 1,900; (Csl, Direct Deposit): 1,496 x 1,874
  • Pixel Pitch: 127 μm²
  • Limiting Resolution: 3.94 lp/mm
  • MTF, X-Ray: >48% at 1 lp/mm, Gd2O2S:Tb screen (80 kVp)
  • DQE (0), RQA5: Direct Deposit CSI: 70% ±5%; Other Scintillator Options: ≈60%
  • Energy Range: 40 - 150 kVp
  • Fill Factor: 57%
  • Image Capture: Intel® Pro/1000GT Desktop Adaptor (PCI) or Epix® CameraLink Adaptor (PCI) (Optional)
  • Scan Method: Progressive
  • Data Output/System Interface: Gigabit Ethernet or CameraLink
  • A/D Conversion: 14-bits
  • Frame Rate: 1 - 10 fps (1 x 1); 15/30 fps (2x2) (Workstation dependent)
Software: The software includes ViVA™, an application for image acquisition and viewing on Windows™. It features a "Virtual Command Processor" for detector calibration, receptor setup, image acquisition, and corrections. ViVA™ supports .viv, .raw, .jpg, and .bmp formats and is compatible with Windows® XP.
Environmental Specifications:
  • Shock: High-shock tolerance
  • Temperature Range: Operating: 10°C to 35°C; Storage: -20°C to +70°C
  • Humidity: Operating and Storage: 10 to 90% (non-condensing)
  • Atmospheric Pressure: 70 kPa to 106 kPa
Regulatory Compliance: UL 60601-1 (U.S.), CSA 22.2 No. 601.1-M90 (Canada)
Mechanical Specifications:
  • Weight: 9.65 lbs. (4.38 kg)
  • Housing Material: Aluminum
  • Sensor Protection Material: Carbon fiber plate (2.5 mm thick) and aluminum
Power Specifications:
  • Power Input: 90 - 264 Vac, 47 - 63 Hz
  • Power Output: +15 Vdc
  • Rating: 1.35 A max. @ 115 Vac, 60 Hz
  • Size: 1.69 (h) x 2.99 (w) x 5.75 (d) Inch [4.29 x 7.59 x 14.61 cm]
Additional Information: PaxScan is a registered trademark of Varian Medical Systems. The document includes reference drawings and contact information for Varian Medical Systems in Salt Lake City, UT.
See more

Catalog excerpts

PaxScan® 2520D/CL-1

X-RAY PRODUCTS PaxScan® 2520D/CL Amorphous Silicon Digital X-Ray Imager Product Description The PaxScan 2520D/CL is a third-generation X-ray imager designed for high-speed dental cone-beam CT imaging. Based upon the new Gigabit Ethernet interface standard, or customer-specified CameraLink standard, images are displayed instantaneously on a user-supplied workstation. The housing is lightweight aluminum. Technical Specifications Receptor Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Amorphous Silicon Conversion Screen . . . . . . . . . . . . . . . . . CsI, DRZ Plus, or Gd2O2S: Tb (Kodak Lanex Screen) The software release includes ViVA™, a basic application for image acquisition and viewing on an end-user workstation or laptop running Microsoft® Windows™. The developer’s software package includes a “Virtual Command Processor” software interface that performs detector calibration, receptor set-up, image acquisition, and image corrections. ViVA™ includes file translators for .viv, .raw, .jpg, and .bmp formats. Windows® XP compatible. Pixel Area Total . . . . . . . . . . . . . . . . . . . . 19.5 (h) x 24.4 (v) cm (7.68 x 9.6 inch) Active (Gd2O2S) . . . . . . . . . . 19.3 (h) x 24.2 (v) cm (7.60 x 9.49 inch) (Csl, Direct Deposit) . . . . . . 19.0 (h) x 23.8 (v) cm (7.48 x 9.37 inch) Pixel Matrix Total . . . . . . . . . . . . . . . . . . . . . . . 1,536 (h) x 1,920 (v) Active (Gd2O2S) . . . . . . . . . . . . . . . . . . . . . . . . . 1,516 (h) x 1,900 (v) (Csl, Direct Deposit) . . . . . . . . . . . . . . . . . . . . . . . . . 1,496 (h) x 1,874 (v) Environmental Shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . High-shock tolerance Temperature Range - Operating . . . . . . . . . . . . . . . 10°C to 35°C (max.) (Ambient) - Storage . . . . . . . . . . . . . . . . . . . . . . -20°C to +70°C Limiting Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.94 lp/mm MTF, X-Ray . . . . . . . . >48% 1 lp/mm, Gd2O2S:Tb screen (80 kVp) DQE (0), RQA5 Direct Deposit CSI. . . . . . . . . . . . . . . . . . 70% ±5% DQE (0), RQA5 Other Scintillator Options . . . . . . . . . . . . . . . ≈60% Humidity - Operating (non-condensing) . . . . . . . . . . . . . . . . . 10 to 90% Storage (non-condensing) . . . . . . . . . . . . . . . . . . . 10 to 90% Atmospheric Pressure . . . . . . . . . . . . . . . . . . . . . . . . 70 kPa to 106 kPa Regulatory Energy Range Standard . . . . . . . . . . . . . . . . . . . . . . . . . 40 - 150 kVp Fill Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57% Image Capture . . . . . . . . Intel® Pro/1000GT Desktop Adaptor (PCI) (Customer Supplied) Epix® CameraLink Adaptor (PCI) (Optional) Scan Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Progressive Data Output/System Interface . . . . . Gigabit Ethernet or CameraLink A/D Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-bits Mechanical Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.65 lbs. (4.38 kg) Housing Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Aluminum Sensor Protection Material Carbon fiber plate (2.5 mm thick) and aluminum Power Power Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 - 264 Vac, 47 - 63 Hz Power Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +15 Vdc Frame Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 10 fps (1 x 1) 15/30 fps (2x2) (Workstation dependent) Rating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.35 A max. @ 115 Vac, 60 Hz Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.69 (h) x 2.99 (w) x 5.75 (d) Inch [4.29 x 7.59 x 14.61 cm] ®PaxScan is a Registered Trademarks of Varian Medical Systems Manufactured by Varian Medical Systems Specifications subject to change without notice.

 Open the catalog to page 1
PaxScan® 2520D/CL-2

medical systems Product Specifications Drawings are for reference only (ACTIVE AREA)! -IMAGING PLANE THREAD HOLES (ACTIVE AREA) ^Threaded Hole (Chassis Ground} -CHASSIS GROUND THREADED HOLE; -POWER CONNECTOR medical systems Salt Lake City, UT Copyright © 2011, Varian Medical Systems, Inc. All Rights Reserved.

 Open the catalog to page 2

All Varian Oncology catalogs and technical brochures

  1. Halcyon

    15  Pages

  2. Eclipse

    7  Pages

Archived catalogs

  1. VitalBeam

    8  Pages

  2. Eclipse™

    2  Pages

  3. ProBeam 360

    2  Pages

  4. 360 ONCOLOGY

    3  Pages

  5. Nexus-DRF™

    2  Pages

  6. 2520DX

    2  Pages

  7. 1515DXT

    2  Pages

  8. 1308DXT

    2  Pages

  9. Clinac Brochure

    12  Pages

  10. Edge Brochure

    22  Pages

  11. NDI-452

    4  Pages

  12. M-101G

    6  Pages

  13. M-105SP

    6  Pages

  14. M-109

    6  Pages

  15. M-113

    6  Pages

  16. M-143

    6  Pages

  17. M-147SP

    6  Pages

  18. M-146

    6  Pages

  19. M-147

    6  Pages

  20. RAD-85S

    6  Pages

  21. M-149

    6  Pages

  22. M-151

    10  Pages

  23. M-152

    12  Pages

  24. M-153

    8  Pages

  25. PaxScan 3030+

    2  Pages

  26. M-171

    6  Pages

  27. PaxScan 4030E

    2  Pages

  28. PaxScan 4336R

    2  Pages

  29. 1313

    2  Pages

  30. Vitesse 3.0

    3  Pages

  31. BrachyVision

    7  Pages

  32. RapidArc

    8  Pages

  33. Vitesse? 2.5

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.