Overview: The BIOPURE 7/15 is a cost-effective and space-saving water purification system designed to deliver compliant purified water for washer disinfectors and autoclaves, minimizing contamination risks.
Key Features:
  • Cost-effective: Provides on-demand pure water at a price comparable to bottled water.
  • Space-saving: Compact size allows for under-bench, wall-mounted, or cabinet installation.
  • Guaranteed bacterial specification: Achieved through integrated purification technologies and easy system disinfection.
  • Consistent water quality: Maintained by regular recirculation through purification technologies.
System Components:
  • Feedwater Inlet
  • Optional Boost Pump
  • Sanitisation Port
  • Pre-Treatment
  • Reverse Osmosis
  • Water Quality Sensor
  • Recirculation Pump
  • Dispense Tap
  • UV Lamp
  • Ultrafiltration Membrane
  • Biopure Purification Cartridge
Treated Water Specifications:
  • Make-up rate: 7 l/hr for BIOPURE 7, 15 l/hr for BIOPURE 15 at 15°C.
  • Delivery flow rate: Maximum 2 l/min at 15psi (1bar).
  • Inorganics: Conductivity <1 µS/cm at 25°C.
  • Bacteria: <1 CFU/10ml.
  • Endotoxin: <0.25 EU/ml.
Dimensions and Weights:
  • Height: 460mm (18.1in), Width: 550mm (21.7in), Depth: 270mm (10.6in).
  • Operational weight: 20Kg for BIOPURE 7, 21Kg for BIOPURE 15.
  • Installation: Bench/Wall in a properly ventilated area.
Feedwater Requirements:
  • Source: Tap water with conductivity <2000 µS/cm.
  • Contaminants: Hardness <250 ppm as CaCO3, Free chlorine 0.5 ppm, Chloramine <0.2 ppm, Silica <30 SiO2 ppm, Fouling index <10 FI, Iron/manganese <0.2 ppm.
  • Temperature: 4 – 40°C (Recommended 15 - 25°C).
  • Flowrate: 1.3 l/min for BIOPURE 7, 1.4 l/min for BIOPURE 15.
  • Drain requirements: 1.2 l/min with gravity fall and air gap.
Feedwater Pressure:
  • Maximum without internal boost pump: 90psi (6.0bar).
  • Minimum without internal boost pump: 60psi (4.0bar).
  • Maximum with internal boost pump: 30psi (2.0bar).
Electrical Requirements:
  • Mains Input: 100-240V ac, 50-60Hz.
  • System control voltage: 24V dc.
  • Power consumption: 80VA with boost pump, 50VA without boost pump.
  • Electrical protection rating: 13 amp supply with suitable ELB.
  • Noise level: <45 dBA.
Contact Information: ELGA LabWater, tel: +44 (0) 203 567 7300, email: [email protected], website: www.elgalabwater.com.
See more

Catalog excerpts

BIOPURE 7/15-1

BIOPURE HEALTHCARE BIOPURE 7/15 The BIOPURE 7/15 has been designed to deliver compliant purified water at the point of use for a wide range of washer disinfectors and autoclaves to minimize contamination risk to patients. • Cost effective – provides the convenience of on-demand pure water at a price comparable with purified bottled water • Space saving – the BIOPURE’S compact size means that it can be positioned under a bench, wall mounted or fitted into a cabinet • Guaranteed bacterial specification, provided by integrated purification technologies combined with easy system disinfection • Water specification is maintained in the reservoir by regular recirculation through the purification technologies BIOPURE 7/15 Water Quality Sensor Feed from Docking Vessel or Reservoir Recirculation Pump Water Quality Sensor PreTreatment Reverse Osmosis Feedwater Boost Pump Inlet (optional) Reverse Osmosis Outlet to application Sanitisation Port UV Lamp Drain Biopure Purification Cartridge Ultrafiltration Membrane Drain Outlet to Docking Vessel o

 Open the catalog to page 1
BIOPURE 7/15-2

BIOPUREHEALTHCARE Treated Water Specifications MODEL Make-up rate @15°C Delivery flow rate - maximum Inorganics (conductivity @25°C) Bacteria - typical Endotoxin 1 System to be installed and regularly disinfected following ELGA LabWater installation design guidelines. Sample to be taken from hygienic sample valve and sampled in accordance with ELGA LabWater guidelines. 2Install in a properly ventilated area. Source Conductivity Contaminant Hardness Free chlorine Chloramine Silica Fouling index Iron/manganese Organics Temperature Flowrate (requirement at 15°C) Drain requirements (gravity fall...

 Open the catalog to page 2

All VEOLIA catalogs and technical brochures

  1. CENTRA Brochure

    16  Pages

  2. Medica Brochure

    16  Pages

  3. Flex 5 & 6

    2  Pages

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About MedicalExpo
MedicalExpo is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. MedicalExpo is an entity of VirtualExpo Group.