TECHNICAL REPORT PYTB3
12Pages

{{requestButtons}}

Catalog excerpts

TECHNICAL REPORT PYTB3 - 1

SATRA Technology Centre Ltd Wyndham Way, Telford Way, Kettering, Northamptonshire, NN16 8SD United Kingdom Tel: +44 (0) 1536 410000 Fax +44 (0) 1536 410626 email: info@satra.co.uk www.satra.co.uk Customer details: Versapak (International) Ltd The Versapak Centre 4 Veridion Way Erith Kent DA18 4AL Julie Goddard Samples received: 29 August 2014 Your reference: TECHNICAL REPORT Subject: Decribed as Insulated Pathology Transport Bag for testing as per HML0217296/1333/2 Conditions of Issue: This report may be forwarded to other parties provided that it is not changed in any way. It must not be published, for example by including it in advertisements, without the prior, written permission of SATRA. Results given in this report refer only to the samples submitted for analysis and tested by SATRA. Comments are for guidance only. A satisfactory test report in no way implies that the product tested is approved by SATRA and no warranty is given as to the performance of the product tested. SATRA shall not be liable for any subsequent loss or damage incurred by the client as a result of information supplied in the report. Please note uncertainty of measurement has not been applied to the results in this report. SATRA uncertainty of measurement values are available on request. Report signed by: Position: Department: Kaileigh Heels Technologist Homeware & Lifestyle (Page 1 of 12) SATRA Technology Centre Ltd (a subsidiary of SATRA). Registered in England No. 3856296 at the above address.

Open the catalog to page 1
TECHNICAL REPORT PYTB3 - 2

TECHNICAL REPORT Testing requested We have carried out testing on two insulated pathology transport bags reference PYTB3 using a modified version of BS EN 12546-2:2000 Clause 4.2 Materials and articles in contact with foodstuffs. Insulated containers for domestic use. Specification for insulated bags and boxes -Insulation performance The modifications which we have applied are as follows: A thermocouple was placed in approximately the centre of the sample measuring the internal air temperature of the sample. Six 140mm x 250mm freezer blocks (supplied) were placed into the base of the...

Open the catalog to page 2
TECHNICAL REPORT PYTB3 - 3

TECHNICAL REPORT Test Results Mean external temperature: 37°C Sample 1 Elapsed Time (hours) Elapsed Time (hours) Versapak (International) Ltd SATRA Reference: HML0227792/1435/3 Date: 25 September 2014

Open the catalog to page 3
TECHNICAL REPORT PYTB3 - 4

TECHNICAL REPORT Sample 1 Air Temperature at 37°C Atmosphere 40.0 Time (hours) Versapak (International) Ltd SATRA Reference: HML0227792/1435/3 Date: 25 September 2014

Open the catalog to page 4
TECHNICAL REPORT PYTB3 - 5

TECHNICAL REPORT Sample 2 Air Temperature at 37°C Atmosphere 40.0 Time (hours) Versapak (International) Ltd SATRA Reference: HML0227792/1435/3 Date: 25 September 2014

Open the catalog to page 5
TECHNICAL REPORT PYTB3 - 6

TECHNICAL REPORT Mean external temperature: 22°C Sample 1 Elapsed Time (hours) Elapsed Time (hours) Versapak (International) Ltd SATRA Reference: HML0227792/1435/3 Date: 25 September 2014

Open the catalog to page 6
TECHNICAL REPORT PYTB3 - 7

TECHNICAL REPORT Sample 1 Air Temperature at 22°C Atmosphere 30.0 Time (minutes) Versapak (International) Ltd SATRA Reference: HML0227792/1435/3 Date: 25 September 2014

Open the catalog to page 7
TECHNICAL REPORT PYTB3 - 8

TECHNICAL REPORT Sample 2 Air Temperature at 22°C Atmosphere 25.0 Time (hours) Versapak (International) Ltd SATRA Reference: HML0227792/1435/3 Date: 25 September 2014

Open the catalog to page 8
TECHNICAL REPORT PYTB3 - 9

TECHNICAL REPORT Mean external temperature: 5°C Sample 1 Elapsed Time (hours) Elapsed Time (hours) Versapak (International) Ltd SATRA Reference: HML0227792/1435/3 Date: 25 September 2014

Open the catalog to page 9
TECHNICAL REPORT PYTB3 - 10

TECHNICAL REPORT Sample 1 Air Temperature at 5°C Atmosphere 20.0 Time (hours) Versapak (International) Ltd SATRA Reference: HML0227792/1435/3 Date: 25 September 2014

Open the catalog to page 10
TECHNICAL REPORT PYTB3 - 11

TECHNICAL REPORT Sample 2 Air Temperature at 5°C Atmosphere 20.0 Time (hours) Versapak (International) Ltd SATRA Reference: HML0227792/1435/3 Date: 25 September 2014

Open the catalog to page 11
TECHNICAL REPORT PYTB3 - 12

TECHNICAL REPORT TERMS AND CONDITIONS OF BUSINESS 1. GENERAL Work done or services undertaken are subject to the terms and conditions detailed below and all other conditions, warranties and representations, expressed or implied are hereby excluded. PRICES Prices are based on current material and production costs, exchange rates, duty and freight and are subject to change without notice. DELIVERY ESTIMATES Delivery estimates are made in good faith and date from receipt of a written order and full information to enable us to proceed. While SATRA or its subsidiaries (hereafter referred to as...

Open the catalog to page 12