Overview: The document provides detailed specifications and requirements for the Standard Vet X-ray system, highlighting its features, optional components, and installation requirements.
Specifications: The system includes a 32kW High Frequency Compact Generator with Anatomical Programming (APR), a variable focal distance, and a rotor brake. It features a 4-way float top table, heavy gauge steel construction, and a moveable grid cabinet. The X-ray generator operates at 400 mA, 125 kVp, and 500mAs, with a serial interface control and push-button console.
Optional Features: Options include a fractional focal spot X-ray tube, non-certified medical collimator, 3-phase and battery-powered generators, and a variable tube stand and table. Additional features include a rotating tube head and collimator, hands-free collimator light control, and a stainless steel cassette tray.
Warranty: The system is covered by a 5-year parts warranty, except for the X-ray tube, which has a 12-month warranty from the installation date.
Room Specifications: Recommended room size is 8’ x 10’, with a ceiling height of 7 feet 4 inches. Shielding requirements should be verified with local building codes or a State Physicist.
Power Requirements: The system requires a single-phase 208/230/240 VAC power line, with specific wire size requirements based on distance from the main power panel. For distances over 100 feet, a three-phase 208 VAC generator is recommended. The disconnect panel should have a minimum of 80AMP service.
Contact Information: For further details, contact via phone at 800.920.9525, email at [email protected], or visit www.vetray.com.
See more

Catalog excerpts

Standard Vet-1

Standard Vet Revolutionizing a new gold standard 32kW High Frequency Compact Generator • Anatomical Programming (APR) • Variable Focal Distance Rotor Brake • 4 Way Float Top Table • Heavy Gauge Steel Construction • Moveable Grid Cabinet High Frequency X-ray Generator Variable Tube Stand and Table X-ray Tube – Rotating Anode 400 mA 125 kVp 500mAs Serial interface control Push button console with APR window Anatomical Programmed Radiography (APR) 208/220/240 VAC Single Phase Power Exposure Foot pedal Variable SID for table top and cassette tray Tube stand longitudinal motion stationary Rotating tube head and collimator 5 foot long 4 way float top table Foot treadle to operate table locks Hands free collimator light control Manual Collimator with light switch Grid Cabinet with 8:1 103LP Grid Stainless steel cassette tray 1.0/2.0 Focal Spots 16° Target Angle 140,000 HU 6 meter HV cables Optional Features Fractional Focal Spot X-ray tube Non-certified medical collimator 3 phase and battery powered generators

 Open the catalog to page 1
Standard Vet-2

High Frequency Generator Vet Ray Standard Vet X-ray System Dimensions Hands Free Collimator Light Control Recommended X-ray Room Specifications Room size: 8’ x 10’ (it will work in a smaller room) Shielding requirements for X-ray room: Check with the local building code or State Physicist for specific requirements. Ceiling height: 7 feet 4 inches X-ray System Power Requirements The Standard Vet system is powered by a High Frequency X-ray Generator. There are certain requirements that must be met in order for the system to operate correctly. These requirements are dependant on these three factors:...

 Open the catalog to page 2

All Vet Ray Technology catalogs and technical brochures

  1. Premium Vet

    2  Pages

  2. Mobile X-ray

    2  Pages

  3. DVR CCD

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.