General Overview
The MiniTemp Veterinary Temperature Monitor is a digital thermometer designed for veterinary use, featuring detachable probes for accurate temperature monitoring. It offers fast response times due to its small sensor size and can display temperatures in Celsius or Fahrenheit. The device indicates if a patient's temperature is within a 'Comfort Zone'.

Preparation
To prepare the MiniTemp for use, install a 9v PP3-type battery and attach a temperature probe using the provided extension lead.

Sensors/Probes
The device includes a rigid and a flexible probe, both using a 1.2mm thermistor. Probes should be used with a thermometer sleeve and cleaned between uses. They are suitable for rectal or oesophageal use.

Operation
Turn the unit on by pressing the ON/OFF button. The LED indicator will flash green if the temperature is within the Comfort Zone (36.6°C to 39.6°C) and red if outside this range. Temperature units can be switched by holding the ON/OFF button.

Maintenance
The unit can be cleaned with a damp cloth, while probes should be cleaned with alcohol or disinfectants. Probes should not be heat sterilized.

Specifications
Mechanical dimensions are 117 x 68 x 26mm, weighing 125g without the rubber boot. It operates on a 9v battery with an accuracy of +/- 0.1°C within the specified range of 5°C to 45°C.

Manufacturer Information
Manufactured by Vetronic Services LTD, located in Devon, UK.
See more

Catalog excerpts

MiniTemp Operators Guide-1

The MiniTemp Veterinary Temperature Monitor from Vetronic Services LTD Users Manual Ver 1.01

 Open the catalog to page 1
MiniTemp Operators Guide-3

The MiniTemp Veterinary Temperature Monitor General The MiniTemp Veterinary Temperature Monitor is a digital single channel thermometer that has been designed specifically for the veterinary market. The unit uses detachable probes to accurately monitor and report the environmental or body temperature. Each probe uses a tiny 1.2mm sensing thermistor resulting in a very small sensor body size. The small physical size and low thermal inertia of the sensor produces very fast response times to changes in temperature. Two probes are supplied; a rigid probe and a flexible probe. Temperature measurements...

 Open the catalog to page 3
MiniTemp Operators Guide-4

Sensors/Probes The unit is supplied with two sensors; a rigid probe and a flexible probe. Both types use a 1.2mm fast-response thermistor and are identical in their electrical performance. Wherever possible use a thermometer sleeve over the temperature probe before inserting into an orifice. The probes are suitable for rectal or oesophageal use and should be cleaned between uses. Care should be exercised when inserting the probes, which should glide in smoothly without any obstruction. If obstruction is felt, withdraw the probe, lubricate and re-position. Never force a probe if resistance is...

 Open the catalog to page 4
MiniTemp Operators Guide-5

LED Indicator The LED indicator at the top of the unit will flash either Red or Green. Flashing Green: The temperature measured by the sensor lies within the Comfort Zone Flashing Red: The temperature measured by the sensor lies outside the Comfort Zone Comfort Zone The Comfort Zone is a region of temperature values that indicate for veterinary mammalian species that the patient is neither to hot nor too cold. This region is taken to be from 36.6◦C to 39.6◦C. Within the Comfort Zone: If the temperature measured lies within the comfort zone then the LED will flash green and there are no warning...

 Open the catalog to page 5
MiniTemp Operators Guide-6

Maintenance and Cleaning The main unit can be wiped over with a damp cloth to remove any stains and marks. The front panel is resistant to normal detergents and ammonia-based cleaning compounds. Temperature probes should be carefully cleaned with a soft cloth using surgical spirit, Isopropyl alcohol or general disinfectants such as Chlorhexidine. DO NOT HEAT STERILIZE THE TEMPERATURE PROBES OR EXTENSION LEADS AS THIS WILL DESTROY THEM

 Open the catalog to page 6
MiniTemp Operators Guide-7

MiniTemp Specifications Mechanical: Unit Size: With fitted rubber boot: Weight: With fitted rubber boot: Electrical Battery Type: Current: Operating Temperature range Accuracy PP3 9v Alkaline or re-chargeable Average 5mA Standby 90uA 5◦C to 45◦C +/- 0.1C within specified range Accuracy outside of this range is unspecified

 Open the catalog to page 7
MiniTemp Operators Guide-8

MiniTemp is manufactured in the United Kingdom by Vetronic Services LTD 12 Henleys Business Park Manor Road Abbotskerswell Newton Abbot Devon, TQ12 5NF Telephone +44 (0) 1626 365505 Fax +44 (0)870 129 4705 www.vetronic.co.uk [email protected]

 Open the catalog to page 8
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.