1. Catalogs
  2. Videomed
  3. otoacustin emission

otoacustin emission

otoacustin emission

otoacustin emission

Product catalog summary
System Overview: The document describes the HOMOTH ABR/BERA module, a real-time brainstem audiometer designed for audiological diagnostics. It utilizes modern processors and is adaptable to future diagnostic demands through software updates.
System Requirements: The system requires a Pentium PC with a minimum of 500 MHz, USB 2.0 port, and Windows 98, ME, 2000, or XP.
Standards and Compliance: The device complies with EN 60601-1/1-1/1-2 standards and AGERA rules, ensuring safety and performance.
Technical Specifications: The system features galvanically separated electrodes, a stimulus range of 50-500 us clicks, and sinus stimuli in preparation. It supports positive, negative, and alternating polarity with an intensity range of 0-110 dB SPL. The rate is adjustable from 1-50 per second in 0.1 steps, with masking capabilities of 0-80 dB white noise.
Measurement and Analysis: The system includes a 1-channel EEG amplifier with 80 dB gain and input impedance greater than 48 MOhm. It features a 12-bit A/D converter at 100 kHz, an averager for up to 10,000 sweeps, and online artefact elimination. Filters include highpass (100-300 Hz), lowpass (1-8 kHz), software filters, and a 50 Hz notch filter.
Results Presentation: Results are displayed as curve diagrams, latency diagrams, and direct comparisons between right and left measurements.
Physical Characteristics: The device dimensions are 32 cm (W) x 27 cm (D) x 7.5 cm (H) with a weight of 1.8 kg. It includes a measure cable of 275 cm and an electrode cable of 60 cm.
Power and Accessories: The system consumes 15 W at 13.2 V and comes with accessories such as headphones, measure cables, electrodes, a power pack, and a program CD.
Additional Features: The system is expandable with electronystagmography (ENG/VNG) and otoacoustic emissions (OAE) modules. It supports both transient evoked and distortion-product otoacoustic emissions, suitable for all age groups and various diagnostic needs.
Innovative Analysis: The HOMOTH system uses an expert system with artificial neural networks and fuzzy logic for objective hearing test results, providing high safety and reliability even for untrained users.
See more

Catalog excerpts

otoacustin emission-2

BERA 4000 The HOMOTH ABR / BERA- module is a real time brainstem audiometer. It is developed considering the latest knowledge of the ENT research. By use of most modern processors, a wide range of possibilities is opened. It is a system lasting into the future for a lot of years, because all changes in diagnostic demands can be loaded as an update or upgrade via software into the system. The software contains pre-selected standard programs, to reduce the operation of the system onto a few keymoves. Beside this, it is possible to create storable measuring programs with individual parameter selections....

 Open the catalog to page 2
otoacustin emission-3

otoacoustic emissions leading companies

 Open the catalog to page 3
otoacustin emission-4

DP-OAE 4000 Lately the Otoacoustic Emissions (OAE) have become more and more important in the field of objective diagnosis at sensorineural deafness. Using the T-OAE measurement (Transient Evoked Otoacoustic Emissions) clicks will cause a wide-band-irritation, using the DP-OAE´s specific frequencies for irritation, (Distorsion-Products Otoacoustic Emissions) it is possible to provoke, measure and evaluate frequency related parts of the cochlear. A sound proof cabin is not necessary, only a noise-reduced area. By continuous stimulation with two presetable primary sinus tones(f1, f2), the selected...

 Open the catalog to page 4
otoacustin emission-5

otoacoustic emissions

 Open the catalog to page 5
otoacustin emission-6

T-OAE 4000 As a world novelty, HOMOTH treads new paths in evaluating results of measured Otoacoustic Emissions. All measured data is classified according to signalstatistic and signaldynamic features. The result is transmitted to an expert system as parameter vectors. The expert system, containing more than 10.000 evaluated OAE measurements, compares all parameter vectors with the stored samples. The analysis module consists of an artificial neuronal network, supplemented by modern fuzzy logic. The result is transmitted to the personal computer, via USB port, and is presented on the monitor....

 Open the catalog to page 6
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.