1. Catalogs
  2. Visocor
  3. visocor HM60

visocor HM60

visocor HM60
1 / 17 PagesView full catalog

visocor HM60

Product catalog summary
Introduction
This document provides comprehensive guidelines for the use of the visocor HM60 blood pressure monitoring device by UEBE Medical GmbH. It includes instructions on measurement procedures, error handling, technical specifications, maintenance, and warranty details.
Safety Instructions
The device is designed for adult use to measure blood pressure and pulse at the wrist. It is not suitable for newborns, infants, or pregnant individuals. Users undergoing dialysis or on certain medications should consult a doctor before use. The device should not be used near RF electrosurgical instruments or MRI equipment and must be kept out of reach of children and pets.
Operating the Unit
Unit Description: The device features a display, memory recall button, start/stop button, battery compartment, set button, and cuff. It displays systolic and diastolic values, pulse, irregular pulse waves, memory code, date/time, and WHO classification.
Important Usage Information: Avoid alcohol, caffeine, and smoking before measurements. Ensure the wrist is free from clothing or jewelry and maintain consistent conditions for accurate readings.
Inserting/Replacing Batteries: Use long-life alkaline batteries, replacing all simultaneously and ensuring correct polarity.
Setting Date/Time: Adjust using the SET button.
Attaching the Cuff: Attach correctly and maintain proper posture during measurement.
Measuring Blood Pressure
The device automatically inflates the cuff and displays systolic, diastolic, and pulse values, classifying blood pressure according to WHO guidelines. It detects irregular pulses, advising consultation with a doctor if irregularities persist.
Using the Memory
The device stores up to 120 results and calculates an average from the last three measurements. Data can be retrieved and deleted as needed.
Technical Information
Ensure a consistent power supply and avoid electromagnetic interference. Only use the provided cuff and accessories. Follow the troubleshooting guide for malfunctions.
Maintenance and Warranty
Handle the device carefully to avoid damage. The device comes with a two-year warranty covering defects, excluding normal wear and tear or improper use.
Electromagnetic Compatibility
The device complies with international standards for electromagnetic compatibility, ensuring safe use in various environments without interference.
Disposal and Environmental Considerations
Dispose of technical equipment and batteries at appropriate collection points, not with household waste.
Conclusion
This document serves as a guide for the proper use and maintenance of the visocor HM60 blood pressure monitor, ensuring accurate measurements and device longevity.
See more

Catalog excerpts

visocor HM60-1

Wrist blood pressure monitor

 Open the catalog to page 1
visocor HM60-2

4. Inserting/replacing batteries, battery icon_11 7. Body posture when taking measurements_13 9. Classification of the measured values (WHO)_14 Thank you for choosing the visocor HM60 wrist blood pressure monitor (hereinafter also referred to as the device). These instructions are intended to help the user to use the unit safely and efficiently, and must be kept with the product and forwarded, if applicable. The unit must be used in accordance with the procedures contained in these instructions for use and must not be used for any other purposes. It is important that you read all the instructions...

 Open the catalog to page 2
visocor HM60-3

english Safety instructions The intended purpose of this device is to enable adults to automatically measure their own systolic blood pressure, diastolic blood pressure and pulse at their wrist. The device can also be used by health care professionals. This unit uses the oscillometric method for measuring blood pressure and pulse rate. Warning: Not suitable for use on newborn babies or infants. Warning: Not suitable for use during pregnancy. The device is not suitable for use in close proximity to RF elec-trosurgical instruments or magnetic resonance imaging equipment. Do not use the unit without...

 Open the catalog to page 3
visocor HM60-4

The readings measured using this device do not provide a diagnosis. They are no substitute for visiting a doctor. Do not rely solely on the result obtained; always take the patient’s other symptoms into account too. In case of doubt, please call a doctor or paramedics to help. Discuss the measured values with your doctor. Do not assess the measurement results yourself. Never change the dose of medicines prescribed by your doctor. Before carrying out your self-measurement, please pay attention to the chapter “Important usage information” on page 10. Safety instructions batteries of different types....

 Open the catalog to page 4
visocor HM60-5

3 Memory recall button 2 Average value memory 7 Battery check display 3 PUL. /min = Pulse calculated pulse frequency per minute or memory space 8 Irregular pulse waves 4 Pulse signal display 10 SYS = Systole (upper blood pressure value) 9 DIA = Diastole (lower blood pressure value)

 Open the catalog to page 5
visocor HM60-6

Operating the unit3. Important usage information The measurement results of automated blood pressure monitors can be influenced by the measuring location, posture, previous exertion and general physical condition. Please pay attention to the usage information in order to obtain accurate readings. • Do not consume alcohol, caffeine or smoke for at least one hour before measuring. • Rest for at least 5 minutes before taking the measurement. Depending on the previous degree of exertion, you may even need to rest for up to an hour. • Clear the wrist from all clothing or watches/jewellery since this...

 Open the catalog to page 6
visocor HM60-7

5. Setting the date/time Operating the unit Operating the unit 7. Body posture when taking measurements With the device switched off, press and hold the SET button until the four-digit year appears on the display (Figure 1). Press the Memory button to set the year. Press the SET button to confirm the se- _ lected year and proceed to setting the date Figure 1 (Figure 2). Set the month with the Memory button and confirm this with the SET button. Set the day (Figure 2), hours and minutes (Figure 3) in the same way. Sit at a dining table, lean your back against the backrest of the chair and bend...

 Open the catalog to page 7
visocor HM60-8

Operating the unit Operating the unit starts. The cuff is inflated to the pressure required for measurement (Figure 2). • The actual measurement takes place with the start of deflation (Figure 3). The 1 symbol in the display flashes until the pulse frequency is displayed. • Once the measurement has been taken, the cuff automatically deflates. The determined systole, diastole and pulse values are shown on the display (Figure 4). • The device switches itself off after approximately 60 seconds. You can also switch the device off before this by pressing the Start/Stop button. The device will classify...

 Open the catalog to page 8
visocor HM60-9

Deleting data Measured results are automatically stored in the memory. The memory can store up to 120 results and the average value. The average value is calculated from the last three measurements taken. Only the entire memory can be deleted. It is not possible to delete individual measured values. When more than 120 measured values have been stored, the oldest value (NO. 120) is deleted to allow the latest value (NO. 1) to be recorded. Data retrieval To retrieve the data, press MEM with the device switched off. The average value of the stored results is displayed along with “AVG.” (Figure 1)....

 Open the catalog to page 9
visocor HM60-10

1. Systolic and diastolic blood pressure values The cardiovascular system has the important function of supplying all organs and tissues in the body with sufficient amounts of blood and of transporting metabolites. To do so, the heart contracts and expands at a regular rate of about 60 to 80 times per minute. The pressure of the flowing blood on the artery walls caused by the heart contracting is termed systolic. The pressure in the ensuing relaxation phase, when the heart refills with blood, is termed diastolic. During daily measurement you determine both values. What you should know about blood...

 Open the catalog to page 10
visocor HM60-11

Technical information Technical information 1. Error messages

 Open the catalog to page 11
visocor HM60-12

Technical information Technical information 4. Technical inspection / calibration check Generally, it is recommended that a metrological inspection is performed every two years. However, professional users in Germany are obligated to do so in accordance with the “Medical Products Operator Ordinance” (Medizinprodukte-Betrei-berverordnung). This can be performed either by UEBE Medical GmbH, an authority responsible for metrology or an authorised maintenance service. Please refer to your national regulations. Upon request, responsible authorities or authorised maintenance services receive a “Test...

 Open the catalog to page 12
visocor HM60-13

fSN~| Serial number PF| Reference number = item number IP22 Protected against solid foreign particles with a diameter of more than 12.5 mm, protection against water droplets. m Direct current Hil Manufacturer Technical equipment and batteries do not belong in household waste. They must be disposed of at appropriate collection and disposal points. The unit is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use. Portable and mobile HF communications equipment can affect the unit. Use of the unit in...

 Open the catalog to page 13

All Visocor catalogs and technical brochures

  1. visocor OM60

    17  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.