visocor OM60

Product catalog summary
Safety Instructions
The device is designed for adults to measure blood pressure and pulse at the upper arm. It is not suitable for newborns, infants, or pregnant individuals. Users undergoing dialysis or on certain medications should consult a doctor before use. Keep the device away from children, pets, and strong electromagnetic fields.
Operating the Unit
The unit includes a display, memory recall button, start/stop button, cuff, and battery compartment. Ensure the cuff is correctly sized and attached. Maintain consistent posture and environment for accurate readings. Avoid alcohol, caffeine, and smoking before measurements.
Initial Operation
Insert batteries or use a mains adapter. Set the date and time before use. Attach the cuff correctly and ensure the air hose is not kinked. Follow correct body posture guidelines during measurement.
Measuring Blood Pressure
Press the start/stop button to begin. The device inflates the cuff and measures during deflation. Results are displayed and classified according to WHO guidelines. Irregular pulse detection is included; consult a doctor if irregularities are frequent.
Memory and Data Management
The device stores up to 120 measurements and calculates an average of the last three. Older data is overwritten when the memory is full. Data can be accessed via the Memory button and is retained even when batteries are replaced. To delete data, press and hold the Memory button until values are replaced by zeroes.
Technical Information
Use only specified batteries and accessories. Avoid exposure to extreme conditions and do not modify the device. Follow troubleshooting guidelines if issues arise. The device, model visocor OM60, features an LCD display and operates on 4 AAA batteries, using the oscillometric method.
Maintenance and Warranty
Regular maintenance and calibration checks are recommended every two years. Repairs should only be performed by authorized personnel. The device has a 2-year warranty covering defects, excluding normal wear and tear or misuse. Claims require proof of purchase.
Disposal and EMC
Dispose of the device and batteries according to local regulations. Ensure electromagnetic compatibility by avoiding use near strong electromagnetic fields. Guidelines for maintaining separation distances from RF communications equipment are provided.
Care and Maintenance Instructions
  • Wash hands after every measurement. If shared, wash hands before and after use.
  • Protect the device from temperature changes, humidity, dust, and sunlight.
  • Handle the cuff carefully to avoid damage. Keep it away from sharp objects.
  • Clean the device with a soft, dry cloth. Avoid strong solvents.
  • Inspect the cuff for marks or discolorations after use. Clean with a damp soapy cloth, but do not immerse in water.
  • Store the device properly, ensuring no heavy objects are placed on it and the air hose is not kinked.
  • Detach the cuff by gently pulling the air connector, not the hose.
Contact Information
  • UEBE Medical GmbH, Service-Center, Bgm.-Kuhn-Str. 20, 97900 Külsheim, Germany
  • Email: [email protected]
  • Website: www.uebe.com
Additional Information
  • Medical aid code number: 21.28.01.2159
  • Trademarks: visocor and UEBE
  • Subject to technical modifications. Reprinting is prohibited.
See more

Catalog excerpts

visocor OM60-1

Upper arm blood pressure monitor

 Open the catalog to page 1
visocor OM60-2

Table of contents A Intended use 4 5. Inserting/replacing batteries, battery icon_11 8. Body posture when taking measurements_13 10. Classification of the measured values (WHO)_14 Thank you for choosing the visocor OM60 upper arm blood pressure monitor (also referred to in the following as the unit). These instructions are intended to help the user to use the unit safely and efficiently, and must be kept with the product and forwarded, if applicable. The unit must be used in accordance with the procedures contained in these instructions for use and must not be used for any other purposes. It...

 Open the catalog to page 2
visocor OM60-3

Intended use The intended purpose of this device is to enable adults to automatically measure their own systolic blood pressure, diastolic blood pressure and pulse at their upper arm. The device can also be used by health care professionals. This unit uses the oscillometric method for measuring blood pressure and pulse rate. Warning: Not suitable for use on newborn babies or infants. Warning: Not suitable for use during pregnancy. The device is not suitable for use in close proximity to RF electrosurgical instruments or magnetic resonance imaging equipment. Do not use the unit without first consulting...

 Open the catalog to page 3
visocor OM60-4

Safety instructions • The readings measured using this device do not provide a diagnosis. They are no substitute for visiting a doctor. Do not rely solely on the result obtained; always take the patient's other symptoms into account too. In case of doubt, please call a doctor or paramedics to help. • Discuss the measured values with your doctor. Do not assess the measurement results yourself. Never change the dose of medicines prescribed by your doctor. • Before carrying out your self-measurement, please pay attention to the chapter “Important usage information” on page 10. 2. Important technical...

 Open the catalog to page 4
visocor OM60-5

english Unit description Operating the unit Operating the unit 2. Displays ^ Cuff socket ^ Displays ^ Memory recall button ^ SYS = Systole (upper blood pressure value) ll DIA = Diastole (lower blood pressure value) ^ Irregular pulse waves ^ Memory code 3 Date/time 3 Connecting socket for [8 Battery compartment mains adapter calculated pulse frequency per minute or memory space ^ Battery check display ^ Average value memory ^ Pulse signal display 10 WHO classification (page 14)

 Open the catalog to page 5
visocor OM60-6

Operating the unit3. Important usage information The measurement results of automated blood pressure monitors can be influenced by the measuring location, posture, previous exertion and general physical condition. Please pay attention to the usage information in order to obtain accurate readings. • Do not consume alcohol, caffeine or smoke for at least one hour before measuring. • Rest for at least 5 minutes before taking the measurement. Depending on the previous degree of exertion, you may even need to rest for up to an hour. • Expose the upper arm. Under no circumstances may clothes hinder...

 Open the catalog to page 6
visocor OM60-7

6. Setting the date/time Operating the unit Operating the unit • With the device switched off, press and hold the start/stop button until the four-digit year appears on the display (Figure 1). • Press the Memory button to set the year. Press the Start/Stop button to confirm the selected year and proceed to setting the Figure 1 Figure 2 20 20 05/0S • Set the month with the Memory button and confirm this with the Start/Stop button. Set the day in the same way (Figure 2). • Use the memory button to set the hour on the clock and confirm by pressing the start/stop button. Follow the same procedure...

 Open the catalog to page 7
visocor OM60-8

Operating the unit Operating the unit 9. Measuring blood pressure Switch on the unit by pressing the Start/Stop button. The full display appears (Figure 1). • After the device has completed its calibration against the ambient air pressure, the automatically controlled inflation process starts. The cuff is inflated to the pressure required for measurement (Figure 2). • The actual measurement takes place with the start of deflation (Figure 3). The V symbol in the display flashes until the pulse frequency is displayed. Once the measurement has been taken, the cuff automatically deflates. The determined...

 Open the catalog to page 8
visocor OM60-9

Measured results are automatically stored in the memory. The memory can store up to 120 results and the average value. The average value is calculated from the last three measurements taken. Press the MEM key in the shut-off state to get to the memory. Press and hold the MEMORY button again until the measured values are replaces by zeroes (Figure 4). The memory has now been deleted. When more than 120 measured values have been stored, the oldest value (NO. 120) is deleted to allow the latest value (NO. 1) to be recorded. Data retrieval The memory is not automatically deleted when the battery...

 Open the catalog to page 9
visocor OM60-10

p| What you should know about blood pressure1. Systolic and diastolic blood pressure values The cardiovascular system has the important function of supplying all organs and tissues in the body with sufficient amounts of blood and of transporting metabolites. To do so, the heart contracts and expands at a regular rate of about 60 to 80 times per minute. The pressure of the flowing blood on the artery walls caused by the heart contracting is termed systolic. The pressure in the ensuing relaxation phase, when the heart refills with blood, is termed diastolic. During daily measurement you determine...

 Open the catalog to page 10
visocor OM60-11

Technical information Technical information english Failure encountered Measurement range: 2. Customer service Device repairs may only be performed by the manufacturer or an expressly authorised body. Please contact: UEBE Medical GmbH [email protected] 3. Technical data 4. Original spare parts and accessories The following original spare parts and accessories are available from specialist dealers: Ring-cuff 22-42 cm type VWR2 Part. no. 2506001, PZN-16353137 visocor U2MC mains adapter Part no. 2506020, PZN-16353166 Subject to technical modifications.

 Open the catalog to page 11
visocor OM60-12

Technical information 5. Technical inspection / calibration check Protect from sunlight Generally, it is recommended that a metrological inspection is performed every two years. However, professional users in Germany are obligated to do so in accordance with the “Medical Products Operator Ordinance” (Medizinprodukte-Betreiberverordnung). Medical device This can be performed either by UEBE Medical GmbH, an authority responsible for metrology or an authorised maintenance service. Please refer to your national regulations. Reference number = item number Serial number Protected against solid foreign...