Operating Specifications
The Waters 2432 Conductivity Detector operates within a measuring range of 0 to 10,000 µS/cm without the need for range switching. It maintains a drift of less than 0.5% per hour and noise levels at or below 0.005 µS/cm. Temperature control is available from 20 °C to 50 °C in 5 °C increments, with stability at ±0.1 °C. The sampling rate is 10 points per second.
Cell Specifications
The detector features ring-shaped electrodes made from stainless steel, with a cell volume of 0.8 µL and a pressure limit of 725 psi. Wetted materials include chemically inert PCTFE and stainless steel.
Electrical Specifications
Power requirements range from 100 to 240 VAC, with a line frequency of 50 to 60 Hz and a nominal power consumption of 160 VA. It includes an input for inject start.
Physical/Environmental Specifications
The detector's dimensions are 34.3 cm in width, 20.8 cm in height, and 61.0 cm in depth, with a weight of 12.3 kg. It operates within a temperature range of 5 to 35 °C and a humidity range of 20% to 80% (non-condensing), producing audible noise below 58 dBA.
Ordering Information
Various part numbers are available for different configurations, including modular and ACQUITY UPLC H-Class, Alliance, ACQUITY Arc, and ACQUITY QDa systems.
Additional Information
The document highlights the detector's ability to provide highly sensitive and stable performance in non-suppressed ion chromatography, with precise temperature control to eliminate baseline drift. Waters Corporation trademarks and contact information are also provided.
See more

Catalog excerpts

2432-1

2432 Conductivity Detector The Waters® 2432 Conductivity detector offers highly sensitive and stable performance when performing non-suppressed ion chromatography. Precise temperature control of the flow cell, which eliminates baseline drift, coupled with the ability to collect absolute and auto-zeroed data frees the chromatographer for other tasks. OPERATING SPECIFICATIONS1 Measuring range 0 to 10,000 yS/cm without range switching <0.005 yS/cm Temperature control Temperature stability Sampling rate CELL SPECIFICATIONS Electrodes Ring-shaped electrodes made from stainless steel Wetted materials Chemically inert PCTFE and stainless steel

 Open the catalog to page 1
2432-2

ELECTRICAL SPECIFICATIONS Power requirements Line frequency Power consumption Inject start Height: 20.8 cm (8.2 inches) Depth: 61.0 cm (24.0 inches) Weight 12.3 kg (27.2 lbs.) Operating humidity range 20% to 80% (non-condensing) 2432 Conductivity Detector for Modular 176003340 and ACQUITY UPLC® H-Class/ACQUITY UPLC H-Class Bio 2432 Conductivity Detector for Alliance® 176003339 2432 Conductivity Detector for ACQUITY® Arc™ 176003341 2432 Conductivity Detector for ACQUITY QDa® 176003342 1. The percent drift is relative to the base conductivity of the mobile phase in a system running...

 Open the catalog to page 2
Related Searches
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.