CM250

CM250

CM250

Product catalog summary
Overview: The document describes a fully automated, random access analyzer designed for clinical chemistry and turbidimetry, capable of processing up to 230 tests per hour.
Sample Tray: The analyzer features 48 sample positions, accommodating primary tubes or sample cups with a volume range of 2 - 100 μL. It supports continuous sample loading, pre-dilution, and includes a built-in barcode reader.
Reagent Tray: It has 48 reagent positions with a Peltier cooling system maintaining temperatures up to 12ºC below room temperature. Vials can be 20, 30, or 50 mL, with programmable volumes for R1 (0-700 μL) and R2 (0-450 μL), and a standard reagent volume of 200 μL.
Dispensing System: The system includes liquid level detection via a capacitive sensor, a collision sensor, and an internal and external washing system for the dispenser, with a preheated probe.
Reaction Tray: Contains 80 washable cuvettes with a 0.6 cm light path and a heating system with air circulation at 30ºC and 37ºC.
Washing System: Features a fully automatic 5-step washing and drying system with water consumption of less than 1 L/hour.
Optical System: Utilizes a true double beam optical system with interference filters ranging from 340 to 767 nm, a band width of 10 nm, and a range of -0.1 to 3.5 Abs. It includes a 6V - 20W halogen lamp.
Analytical Mode: Supports various modes including end point, kinetics, immunoturbidimetry, and automatic dilution of pathological samples. It also includes quality control features like Levey-Jennings charts and Westgard rules.
Data Handling: Requires an external PC and printer, supports data import/export, and includes software for Windows 10 (64 bits) with automatic backup and two USB 2.0 ports.
Communication: Uses a serial port standard communication following the ASTM 1394 protocol.
Installation Requirements: Operates on a power supply of 85 - 240 VAC, 50/60 Hz, 400 VA, with room temperature requirements of 15 - 35ºC and humidity below 80%.
Dimensions and Weight: The analyzer measures 80 cm (W) x 48 cm (H) x 57 cm (D) and weighs 89 kg (total) and 57 kg (net).
Certifications: Certified under ISO 13485:2016, CE, ANMAT 3266/13, EMC 61326-2-6-2006, and ES 61010-2-101:2002.
Contact Information: Wiener Lab Switzerland S. A., Av. Cardinal - Mermillod 36, 1227 Geneva, Switzerland. Email: [email protected]
Analytes: The analyzer can process a wide range of enzymes, substrates, specific proteins, and inorganic ions, including ALT, ALP, AST, CK, glucose, cholesterol, and more.
See more

Catalog excerpts

CM250-1

Fully automated, random access analyzer for Clinical Chemistry and Turbidimetry

 Open the catalog to page 1
CM250-2

TECHNICAL FEATURES Throughput: up to 230 tests/hour Analytical Mode: • End Point, reagent or sample blank end point • One and two-points kinetics • Immunoturbidimetry • Factor, standard or calibration curve • Routine, STAT, profiles • Automatic dilution of pathological samples or substrate consumption • Samples pre-dilution if method requires • QC: Levey-Jennings charts, Westgard rules Sample Tray: • 48 sample positions • Primary tubes or sample cups • Sample volume: 2 - 100 μL • Continuous sample loading • Sample pre-dilution • Built-in barcode reader Reagent Tray: • 48 reagent positions • Peltier...

 Open the catalog to page 2