Product catalog summary
Introduction
The WishBone Medical Superior Clavicle Plating System is designed for both pediatric and adult patients, offering sterile, single-use kits for individualized care. The system includes narrow, low-profile clavicle plates available in 6-hole and 8-hole configurations, along with 3.5mm locking and non-locking screws.

System Overview
  • Kit Components: Includes a double-ended drill guide, threaded drill guide, drill bits, small depth gauge, solid T15 screwdriver, and a screw assortment.
  • Plate Features: Designed for superior mid-shaft clavicle fixation with compression slots for fracture compression, pre-contoured to fit anatomy, and rounded tips to reduce soft tissue irritation.
  • Screws: 3.5mm cortical and locking titanium screws are included, with additional sizes available separately.

Indications
The system is indicated for clavicle fractures requiring surgical intervention, with specific considerations for pediatric patients.

Surgical Technique
  • Patient Positioning: The patient is positioned in a beach chair or supine position with the head rotated away from the operative side.
  • Approach and Exposure: Incisions are made inferior to the clavicle or along Langer’s lines to minimize nerve damage and improve cosmetic outcomes.
  • Radiologic Evaluation: Preoperative planning and imaging are crucial for assessing fracture and determining plate length.
  • Fracture Reduction and Plate Selection: The lateral fragment is manipulated for reduction, and the appropriate plate is selected based on fracture location and anatomy.
  • Non-Locking Screw Insertion: Non-locking screws are used to lag the plate to the clavicle, with caution to protect the subclavian vascular bundle.
  • Locking Screw Insertion (Optional): Locking screws are inserted into threaded holes using a locking drill guide.
  • Implant Removal: All screws are unlocked and removed, followed by the plate.
  • Closure: An intraoperative radiograph is recommended to confirm fracture reduction and screw positioning before wound closure.

Conclusion
The WishBone Medical Clavicle Plating System provides a comprehensive solution for clavicle fracture fixation, emphasizing ease of use, sterility, and patient-specific care.
See more

Catalog excerpts

SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE-1

SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE

 Open the catalog to page 1
SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE-3

WISHBONE MEDICAL SUPERIOR CLAVICLE PLATING SYSTEM Introduction The WishBone Medical Clavicle Plating System is designed for pediatric and adult patients—delivered in sterile, single-use kit packaging for individualized procedural care. This complete system offers narrow, low-profile, superior mid-shaft clavicle plates in 6-Hole and 8-Hole configurations. Procedure kits include 3.5mm locking and non-locking screws. SYSTEM COMPONENTS 6-Hole or 8-Hole Superior Clavicle Plate Threaded Drill Guide Small Depth Gauge Drill Bits Screw Assortment* Double-Ended Drill Guide * Non-locking (cortical) and...

 Open the catalog to page 3
SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE-4

EVERYTHING YOU NEED, BRAND NEW EVERY TIME. Complete procedure kits in sterile packaging eliminate set processing and help avoid delays or cancellations attributed to missing components. Readily available for immediate use, individual kits deliver sharp and pristine instruments, new for each procedure. SYSTEM FEATURES: SUPERIOR MID-SHAFT CLAVICLE PLATES Smaller, low-profile design for the pediatric patient. Pre-contoured plate designed to fit patient’s anatomy. Compression slots designed to provide plate positioning and fracture compression. Rounded plate tips and smooth edges designed to lessen...

 Open the catalog to page 4
SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE-5

FOR PLATES & SCREWS KITS 2.7mm and 3.5mm cortical (8mm–50mm lengths) and locking screws (10–50mm lengths) are available as sterile packed ancillary items in the WishBox for plates & screws systems. For more information, please refer to the Plates & Screws Delivered Goods Form (LIT-DGF-PS). SYSTEM FEATURES: SCREWS WishBone Medical Superior Clavicle Plates utilize 3.5mm cortical and locking titanium screws. An assortment of each is included in every procedure kit. See quantities below. LOCKING SCREWS CORTICAL SCREWS Designed for fixation in cortical bone. Threaded cortical screw head. Rounded screw...

 Open the catalog to page 5
SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE-6

TABLE OF CONTENTS SURGICAL TECHNIQUE Radiologic Evaluation Straight AP, augmented with an oblique view allows for assessment of the body of the clavicle, acromioclavicular (AC) and sternoclavicular (SC) joints and coracoclavicular (CC) ligaments. Scapular Y views are helpful in assessing the relationship between acromium/coracoid and scapula. Depending on stabilization method, a horizontal incision is placed inferior to the clavicle (fig1). Another option is an incision perpendicular to the long axis along the Langer’s line. This incision may provide better cosmetic results and less damage to...

 Open the catalog to page 6
SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE-7

TABLE OF CONTENTS SURGICAL TECHNIQUE Fracture Reduction and Plate Selection Distract, elevate and rotate the lateral fragment to obtain reduction (fig 2). Select the appropriate Superior Mid-Shaft Clavicle Plate to best address the location of the fracture and patient anatomy. The two middle plate holes should be on either side of the fracture line to allow for nonlocking screw placement to stabilize a wedge fragment if needed (fig 3). Lag screws can be used to stabilize oblique fractures and to stabilize wedge fragments. Reduce the fracture according to standard practice. Caution: When using...

 Open the catalog to page 7
SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE-8

TABLE OF CONTENTS SURGICAL TECHNIQUE Non-Locking Screw Insertion (cont.) Place the 2.8mm end of the double-ended drill guide into the 2nd compression slot. The 2.8mm drill bit is inserted to the desired depth (fig 7). It is recommended that a non-locking cortical screw be used to lag the plate to the clavicle. fig. 7 The depth gauge is then inserted into the plate and hooked to the far cortex (fig 8). The selected screw is then inserted into the plate and tighten using the screwdriver in a clockwise direction (fig 9). Caution: Double-ended drill guide metal tips contain sharp features to provide...

 Open the catalog to page 8
SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE-9

TABLE OF CONTENTS SURGICAL TECHNIQUE Locking Screw Insertion (Optional) Locking screws are used in only the threaded holes. Using the locking drill guide and 2.8mm drill bit, drill to the desired depth (fig 10). Measure for screw length by locating the laser mark on the drill guide. Alternatively, detach the locking drill guide and use the depth gauge to measure for screw length (fig 11). Optional step The appropriate length locking screw is then inserted and tightened clockwise using the screwdriver (fig 12). CLAVICLE PLATES SURGICAL TECHNIQUE

 Open the catalog to page 9
SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE-10

TABLE OF CONTENTS SURGICAL TECHNIQUE Insertion of Remaining Screws Repeat steps 5-6 for insertion of remaining screws. Implant Removal An intraoperative radiograph is recommended to check the final reduction of the fracture and the position of the screws. The wound is closed according to surgeon preference. Unlock all locking screws from the plate, then remove the locking and non-locking screws completely from the bone. Remove the plate. Note: For unstable repair constructs, i.e., highly comminuted fractures, post-operative loading should be restricted to a level determined by the physician until...

 Open the catalog to page 10
SUPERIOR CLAVICLE PLATING SYSTEM SURGICAL TECHNIQUE-11

WishBoneMedical.com 100 Capital Drive Warsaw, IN 46582 +1-574-306-4006 All trademarks herein are the property of WishBone Medical, Inc. or its subsidiaries unless otherwise indicated. This material is intended for the sole use and benefit of Health Care Professionals and the WishBone Medical Sales Force. It is not to be redistributed, duplicated or disclosed without the express written consent of WishBone Medical. All WishBone Clavicle Plating System kits are only available as single-use, sterile packed kits. Always confirm product expiration date prior to use. For product information, including...

 Open the catalog to page 11
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.