Introduction
The ORTHOLOC™ Forefoot Fracture System is designed to enhance surgical flexibility and performance in forefoot fracture and osteotomy fixation. It features polyaxial locked plate technology, offering both traditional non-locking and locking screw/plate constructs for stability. Key attributes include strength with 1.5mm titanium locking plates, versatility with multiple screw/plate options, and specificity tailored by leading surgeons for forefoot fixation challenges.

Thread and Screw Design
The ORTHOLOC™ thread design provides multiple points of engagement for maximum holding power and stiffness, effective up to 15 degrees off-axis. The locking screw, coated with titanium nitride, ensures a secure fit by reforming internal plate threads upon full engagement.

Surgical Technique
Plate Selection: Choose a plate based on the fracture/osteotomy site and fixation goals. Options include Straight, T-Plate, L-Plate, Akin Plate, and 1st Metatarsal Plate, all with specific dimensions and lengths.
Temporary Fixation: Use temporary fixation pins in K-wire holes or plate screw holes to provisionally fixate the plate.
Screw Selection: Plates accept 2.0mm or 2.4mm locking or non-locking screws. Selection depends on anatomy and fixation goals, with specific lengths and drill sizes provided.
Screw Placement: Bicortical fixation is generally unnecessary but recommended for highly comminuted fractures. Use the Star 7 Driver for screw placement.
Compression Feature: Achieve compression using the oblong compression screw slot on specific plates, creating compression as the screw head moves.

Applications
The system is applicable for various forefoot conditions, including osteotomies and fractures of the metatarsal shaft, neck, base, and 5th metatarsal.

Ordering Information
Detailed part numbers and descriptions are provided for plates, screws, and instruments, ensuring precise ordering for surgical needs.
See more

Catalog excerpts

ORTHOLOC® Forefoot Fracture System ? FA255?409-1

ORTHOLOC™ Forefoot Fracture System SURGICAL TECHNIQUE

 Open the catalog to page 1
ORTHOLOC® Forefoot Fracture System ? FA255?409-2

ORTHOLOC™ Forefoot Fracture System Surgical Technique

 Open the catalog to page 2
ORTHOLOC® Forefoot Fracture System ? FA255?409-3

Introduction - Thread Design - Screw Design ORTHOLOC™ Forefoot Fracture System Surgical Technique - Plate Selection - Temporary Fixation - Screw Selection - Screw Placement - Compression Feature ORTHOLOC™ Forefoot Fracture System Applications Ordering Information Proper surgical procedures and techniques are the responsibility of the medical professional. The following guidelines are furnished for information purposes only. Each surgeon must evaluate the appropriateness of the procedures based on his or her personal medical training and experience. Prior to use of the system, the surgeon should...

 Open the catalog to page 3
ORTHOLOC® Forefoot Fracture System ? FA255?409-4

Introduction Chapter Title The ORTHOLOC™ Forefoot Fracture System marks a new level of surgical flexibility and performance for forefoot fracture and osteotomy fixation. Featuring the innovative ORTHOLOC™ polyaxial locked plate technology; this system gives the surgeon the options of a traditional non-locking plate with the stability of a locking screw/plate construct. »» STRENGTH: 1.5mm titanium locking plates for surgical strength and post-op holding power. »» VERSATILITY: Multiple screw/plate options and ORTHOLOC™ polyaxial locking technology provides a multitude of surgical options and fixation...

 Open the catalog to page 4
ORTHOLOC® Forefoot Fracture System ? FA255?409-5

ORTHOLOC™ Forefoot Fracture System Surgical Technique Plate Selection After appropriate dissection and reduction of the fracture/osteotomy site has been achieved, choose an ORTHOLOC™ Forefoot Fracture plate that best achieves the fixation goals specific to the anatomy and indication. | table 1 Straight Plate Akin Plate Plate Thickness Available Lengths Plate Width *Universal left/right Temporary Fixation Place the selected plate on the bone and provisionally fixate using the temporary fixation pins. The temporary fixation pins can be used in the K-wire holes if provided or in any of the plate...

 Open the catalog to page 5
ORTHOLOC® Forefoot Fracture System ? FA255?409-6

Screw Selection The ORTHOLOC™ 2.0/2.4mm plates will accept a 2.0mm or 2.4mm ORTHOLOC™ locking or non-locking screw. Screw selection is based on anatomy and fixation goals. Select the most appropriate screw to achieve the surgical goals specific to the indication and patient. | table 2 Screw Options SCREW Color Magenta (Gold Head) Bronze (Gold Head) Available Lengths The ORTHOLOC™ locking screw can be used at up to 15 degrees off axis to the plate threads. Once the appropriate screw angle is determined, place the polyaxial drill guide into the desired plate screw hole. | Figure 2 Screw length...

 Open the catalog to page 6
ORTHOLOC® Forefoot Fracture System ? FA255?409-7

Screw Placement In general, bicortical fixation is not required with ORTHOLOC™ locking screws; however, it should be strongly considered in cases of highly comminuted fractures or unstable osteotomies. Choose the locking screw that corresponds with the previously selected drill diameter and length determined. Use the Star 7 Driver to drive the screw until the proximal locking threads of the locking screw are fully engaged with the plate or until the head of the non-locking screws sits flush with the plate. Compression Feature Compression across a fracture/osteotomy site can be achieved using...

 Open the catalog to page 7
ORTHOLOC® Forefoot Fracture System ? FA255?409-8

ORTHOLOC™ Forefoot Fracture System Applications Forefoot Osteotomies Metatarsal Neck Fractures Metatarsal Shaft Fractures Metatarsal Base Fractures Appendix 1 ORTHOLOC™ Forefoot Fracture System Applications

 Open the catalog to page 8
ORTHOLOC® Forefoot Fracture System ? FA255?409-9

Ordering Information Plate Options PART NO. LESS MET STRAIGHT PLATE 3-HOLE LESS MET STRAIGHT PLATE 4-HOLE LESS MET STRAIGHT PLATE 6-HOLE 2.0mm Locking Screws PART NO. Appendix 2 Ordering Information

 Open the catalog to page 9
ORTHOLOC® Forefoot Fracture System ? FA255?409-10

Appendix 2 Ordering Information

 Open the catalog to page 10
ORTHOLOC® Forefoot Fracture System ? FA255?409-12

Wright Medical Technology, Inc. 5677 Airline Road Arlington, TN 38002 901.867.9971 800.238.7117 www.wmt.com Wright Medical EMEA Hoogoorddreef 5 1101 BA Amsterdam The Netherlands 011.31.20.545.0100 www.wmt-emea.com ™Trademarks and ®Registered marks of Wright Medical Technology, Inc. ©2010 Wright Medical Technology, Inc. All Rights Reserved.

 Open the catalog to page 12

All Wright Medical Technology catalogs and technical brochures

  1. FORCE FIBER

    4  Pages

  2. BIOFOAM®

    1  Page

  3. TENSIX

    2  Pages

  4. ALLOPURE

    6  Pages

  5. MEDIALMAX

    2  Pages

  6. DARCO™

    1  Page

  7. INVISION™

    2  Pages

  8. PROPHECY

    6  Pages

  9. css

    4  Pages

  10. CL AW ® II

    20  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.