Cleaning and Disinfecting Guide for Greentek Sintered Ag-AgCl Electrode Cap
Introduction: This document outlines the procedures for cleaning and disinfecting the Greentek Sintered Ag-AgCl Electrode Cap to ensure accurate EEG signals and prolong the life of the electrodes.
Cleaning Method:
  • Immediately after use, gently clean off any residual gel from the electrodes.
  • Soak the cap in warm water (up to 50°C) for 15 minutes to dissolve remaining gel. Use a cotton ball only if necessary.
  • If gel persists, soak the cap in diluted detergent for a few hours.
  • Rinse the cap thoroughly with tap water and dry it gently. Store in a dark, dry place.
  • Frequency: After every use.
Disinfecting Method:
  • Follow the initial cleaning steps.
  • Soak the cap in a 1% diluted bleach solution (100 ppm sodium hypochlorite) for up to 30 minutes.
  • Rinse thoroughly with tap water and dry gently. Store in a dark, dry place.
  • Frequency: Every two weeks.
Caution:
  • Handle the connection point of electrodes and lead wire gently.
  • Avoid contact with other metals.
  • Ensure no residual gel remains on electrodes or wires.
  • Keep electrodes and wires away from corrosive liquids.
  • Prevent water contact with the connector.
See more

Catalog excerpts

Cleaning and Disinfecting of Greentek Sintered Ag-AgCl Electrode Cap-1

WUHAN GREENTEK PTY LTD Add: Room 303, Dingye Hall B, International Enterprise Center, Guanggu Avenue, Hongshan District, Wuhan, China, 430074 Tel: +86-27-59262126; +86-18627764693 Web: www.gtsensor.com Email: [email protected] Cleaning and Disinfecting of Greentek Sintered Ag-AgCl Electrode Cap A routine schedule for the Cleaning and Disinfecting of Electrode Cap helps assure accurate EEG signals and the reuse of electrodes between different participants. Furthermore, you will preserve the excellent characteristics of your electrodes and will ensure a long product life. Cleaning Methode: 1. Clean the residual gel off the Sintered Ag-AgCl Electrodes in the cap softly immediately after use. 2. Soak the cap for about 15 minutes in warm water (up to 50 degrees Celsius), so that the remaining gel can dissolves off quickly into the water. Use a cotton ball for removing gel residues from the electrodes only if absolutely necessary. 3. In case the gel stays on the surface of the electrodes or the cap shell, then soak the cap into diluted detergent for a few hours. 4. Rinse the cap in tap water afterwards for several times. Dry the cap softly, Store the cap in a dark dry place. Cleaning frequency: Everytime after use. Disinfecting Methode: 1. Clean the residual gel off the Sintered Ag-AgCl Electrodes in the cap softly immediately after use. 2. Soak the cap for about 15 minutes in warm water (up to 50 degrees Celsius), so that the remaining gel can dissolves off quickly into the water. Use a cotton ball for removing gel residues from the electrodes only if absolutely necessary. 3. Then soak the cap for up to 30 minutes in 1% diluted bleach solution

 Open the catalog to page 1
Cleaning and Disinfecting of Greentek Sintered Ag-AgCl Electrode Cap-2

WUHAN GREENTEK PTY LTD Add: Room 303, Dingye Hall B, International Enterprise Center, Guanggu Avenue, Hongshan District, Wuhan, China, 430074 Tel: +86-27-59262126; +86-18627764693 Web: www.gtsensor.com Email: [email protected] (100 ppm sodium hypochlorite). Household bleach has a concentration of 52,500ppm of available chlorine (5.25% Sodium Hypochlorite). A 1:500 dilution of household bleach provides the 100ppm concentration required) 4. Rinse the cap in tap water afterwards for several times. Dry the cap softly, Store the cap in a dark dry place. Cleaning frequency: Every two weeks. Caution:...

 Open the catalog to page 2

All Wuhan Greentek catalogs and technical brochures

  1. Flex-Cap

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.