Product catalog summary
Technical Specifications of SHALYA
Electrosurgical Solutions: SHALYA offers a complete range of electrosurgical solutions designed to enhance surgical precision and safety. The system operates at a pure sine wave frequency of 390 kHz, providing various cutting and coagulation modes.
Cutting Modes:
  • Sp Cut: Programmable pulse or continuous output at 250 W with a crest factor (CF) of 1.5.
  • Pure: Continuous output at 250 W, CF 1.5.
  • Blend: Bursts at 29 kHz with 55% ON cycle, 150 W, CF 2.5.
Monopolar Coagulation:
  • Soft: 62 kHz, 100 W, CF 5.0.
  • Fulgurate: 52 kHz, 100 W, CF 6.5.
  • Spray: Randomized frequency between 34 kHz and 50 kHz, 100 W, CF 8.0.
PREM System: Monitors patient return electrode with a measuring frequency of 85 kHz ± 10 kHz and current < 10 µA. Acceptable resistance ranges are specified for dual and single area electrodes.
Design Standards: Classified as II b according to EC-Directive 93/42/EEC, with protective class I and type CF according to EN 60 601-1.
Input Power: Operates at 190-265 Volt AC with a maximum power input of 1000 VA and a 4 A fuse.
Dimensions: The device measures 38.0 x 30.5 x 11.5 cm and weighs less than 5 kg.
Advanced Features:
  • 6SENSE Technology: Maintains set power over a wide range of tissue types using tissue feedback technology.
  • PREM: Provides safety against patient return electrode site burns.
  • Randomized SPRAY Coag: Ensures homogeneous and faster coagulation over larger areas.
  • Self Diagnosis: Ensures safe functioning and displays error codes if any issues are detected.
  • Bipolar Cut: Macro mode available for bipolar cutting.
  • Argon Beam Coagulator Upgradable: The system can be upgraded for Argon Beam Coagulation.
  • User Programmable: Three surgical programs can be set after power ON.
  • BiCoag Alarm (Optional): Provides audio feedback at the completion of bipolar coagulation.
  • Sp Cut: Low voltage cut mode minimizes smoke and charring during endoscopic procedures.
  • Soft/Dissect Coag: Facilitates faster cutting with coagulation for thicker tissues.
Leakage Currents:
  • RF Leakage Current: Bipolar < 60 mA rms, Monopolar < 150 mA rms.
  • Low Frequency Leakage Currents: Normal polarity, intact ground < 10 µA; ground open < 50 µA.
Bipolar Modes:
  • Micro: Lower output voltage, 60 W at 100 Ù, CF 1.5.
  • Macro: Higher output voltage, 60 W at 100 Ù, CF 1.5.
Feedback System: Utilizes 3 microprocessor-controlled 6SENSE technology with a sampling rate above 2000 senses/second. Maximum output power deviation is < 15% or 5 watts, whichever is greater.
Contact Information: For further details, contact via email at [email protected] or visit www.xcellancetech.com. Located at 3057, Bhandup Indl. Estate, Pannalal Compound, LBS Marg, Bhandup (W), Mumbai - 400 078, India. Tel: +91-22-2596 4645 / 3299 5113.
See more

Catalog excerpts

Easy-1

SHALYA Complete range of electro surgical solutions... Helping you make the right choice.. Add Hi-Tech Precision to your Surgical Expertise... Technical Specifications of SHALYA Monopolar Cut: Pure Sine wave 390 kHz Sp Cut- Programmable Pulse or continues output 250 W at 300 Ù, CF 1.5 Pure - Continues output, 250 W at 300 Ù, CF 1.5 Blend - Bursts repeating at 29 kHz 55% ON cycle, 150 W at 300 Ù, CF 2.5 Monopolar Coag : Damped Sine wave 460 kHz Soft - Repetition Frequency 62 kHz, 100 W at 500 Ù, CF 5.0 Fulgurate - Repetition Frequency 52 kHz, 100 W at 500 Ù, CF 6.5 Spray - Randomized Repetition Frequency 34 kHz < f < 50 kHz, 100 W at 500 Ù, CF 8.0 Bipolar : Pure Sine wave 390 kHz Micro - Lower output voltage, 60 W at 100 Ù, CF 1.5 Macro - Higher output voltage, 60 W at 100 Ù, CF 1.5 DESIGN STANDARD Classification according to the EC-Directive 93/ 42/ EEC : II b Protective class according to EN 60 601-1 : I Type according to EN 60 601-1 : CF LEAKAGE CURRENTS • RF leakage current • Monopolar : < 150 mA rms • Bipolar : < 60 mA rms • Low frequency leakage currents • Normal polarity, intact ground : < 10 µ • Normal polarity, ground open : < 50 µA PREM SYSTEM Measuring Frequency : 85 kHz ± 10 kHz Measuring current : < 10 µA Acceptable Resistance Range Dual Area PREM Patient Return Electrode : 10-135 Ù Single Area Patient Return Electrode : < 10 Ù FEEDBACK SYSTEM • 3 Microprocessor Controlled 6SENSE Technology with sampling rate above 2000 sense/second • Maximum output power deviation < 15% or 5 watts, whichever is greater at any setting with PER-97 INPUT POWER Operating Voltage : 190-265 Volt AC, (other ratings available on request) Max. Power input : 1000 VA, Fuse : 4 A DIMENSIONS L x W x H in cm - 38.0 x 30.5 x 11.5, Extended Range Features of SHALYA 6SENSE Technology : Maintain set power over wider range of tissue by Tissue Feedback Technology. PREM : Patient Return Electrode Monitoring - Safety against the Patient Return Electrode site burns. Randomized SPRAY Coag : Homogeneous, superficial, faster Coagulation over larger area. Self Diagnosis : Ensure the ESU with accessories are safe for functioning & display Error Codes if any. Bipolar Cut : Macro mode can be used for bipolar cutting with hooks and other cutting devices. Argon Beam Coagulator up-gradable : ESU can be upgraded for Argon Beam Coagulation module. User Programmable : THREE Surgical Programs help in setting the system after power ON. BiCoag Alarm (Optional) : Audio feedback at completion of Bipolar coag, reduces charring and sticking. electro surgical Sp Cut : Low voltage cut mode for cutting with minimum smoke & charring for endoscopic procedure. Soft/Dissect Coag : Faster cutting with coagulation for thicker tissues & soft coagulation. 3057, Bhandup Indl. Estate, Pannalal Compound, LBS Marg, Bhandup (W), Mumbai - 400 078. India • Tel : +91-22-2596 4645 / 3299 5113 Email : [email protected]

 Open the catalog to page 1

Archived catalogs

  1. Shalya Lx

    2  Pages

  2. Shalya Easy

    2  Pages

  3. Shalya Mx

    2  Pages

  4. Shalya Dx

    2  Pages

  5. Shalya Easy +

    2  Pages

  6. APC

    1  Page

  7. Products

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.