Certification Overview:
The Tianlong Biolum Portable ATP Hygiene Monitoring System, produced by Xi’an Tianlong Science and Technology Co., Ltd., has been certified by the AOAC Research Institute under certificate number 022401. This certification is valid from March 4, 2024, to December 31, 2024.

Method Description:
This system utilizes a bioluminescence assay to monitor hygiene by detecting Adenosine Triphosphate (ATP) on stainless steel surfaces, primarily in the food and beverage sectors. It employs an ATP Test Swab with recombinant luciferase reagent, which emits fluorescence upon reacting with ATP, measured in Relative Light Units (RLUs).

Performance Claims:
The system effectively monitors cleanliness with a detection limit of 0.9 fmol ATP, demonstrating good sensitivity and linearity in detecting ATP from various food and microbial sources.

Validation Study:
The validation study confirmed the system's sensitivity to ATP in both pure analyte and food/microbial matrices, capable of detecting ATP at low dilutions. Variations in results were noted due to environmental factors and sample handling.

Data Analysis:
Data tables and figures provide detailed performance data, including regression analyses and raw/calculated data for different matrices and microbial studies. Consistent results were observed across various conditions, with some variability due to sample preparation and environmental conditions.

Conclusion:
The Tianlong Biolum Portable ATP Hygiene Monitoring System is a reliable tool for hygiene monitoring, capable of detecting ATP from diverse sources and providing accurate cleanliness assessments. Regular cleaning is essential to achieve negative ATP results, as ATP remains stable on surfaces unless physically removed.

Overview:
The document includes detailed data and analysis on the selectivity and performance of the system, focusing on ATP detection in matrices like sweetened yogurt and doughnut residues on stainless steel surfaces.

1. Selectivity Data:
Table 7 shows raw and calculated data for selectivity, highlighting the system's ability to differentiate ATP from other nucleotides, with ATP showing significantly higher RLU values.

2. Stainless Steel Matrix Study:
Tables 11 and 12 present data from studies on stainless steel surfaces contaminated with sweetened yogurt and doughnut residues, comparing wet and dry sampling methods across different RLU target ranges. Key metrics include mean RLU, standard deviation (sr), and relative standard deviation (RSD).

3. Key Findings:
  • The system shows consistent performance across different matrices and sampling methods, with wet sampling generally yielding higher RLU values.
  • Higher RLU values correlate with higher contamination levels, indicating sensitivity to varying ATP concentrations.
  • The coefficient of variance (RSD) indicates measurement precision, with lower RSD values reflecting more reliable results.

4. References:
The document cites validation studies and user manuals that support the findings and provide additional context for using the Tianlong Biolum Portable ATP Hygiene Monitoring System.
See more

Catalog excerpts

AOAC Cert of Biolum Portable ATP Hygiene Monitoring System-1

CERTIFICATION AOAC Research Institute Performance Tested MethodsSM Certificate No. The AOAC Research Institute hereby certifies the method known as: Tianlong Biolum Portable ATP Hygiene Monitoring System manufactured by Xi’an Tianlong Science and Technology Co., Ltd. No.4266, Shanglin Road, Weiyang District Xi’an, 710021, Shaanxi P.R. China This method has been evaluated in the AOAC Research Institute Performance Tested MethodsSM Program and found to perform as stated in the applicability of the method. This certificate indicates an AOAC Research Institute Certification Mark License Agreement has been executed which authorizes the manufacturer to display the AOAC Research Institute Performance Tested Methods SM certification mark on the above-mentioned method for the period below. Renewal may be granted by the Expiration Date under the rules stated in the licensing agreement. Bradley A. Stawick, Senior Director Signature for AOAC Research Institute Issue Date Expiration Date 2275 Research Blvd., Ste. 300, Rockville, Maryland, USA Telephone: +1-301-924-7077 Fax: +1-301-924-7089 Internet e-mail: [email protected] * World Wide Web Site: http://www.aoac.org

 Open the catalog to page 1
AOAC Cert of Biolum Portable ATP Hygiene Monitoring System-2

SUBMITTING COMPANY Xi’an Tianlong Science and Technology Co., Ltd. No.4266, Shanglin Road, Weiyang District Xi’an, 710021, Shaanxi P.R. China METHOD NAME Tianlong Biolum Portable ATP Hygiene Monitoring System INDEPENDENT LABORATORY Q Laboratories 1930 Radcliff Drive Cincinnati, OH 45204 APPLICABILITY OF METHOD Target organism – Adenosine Triphosphate (ATP) Matrixes – Stainless steel (304) Performance claims – The Biolum Portable ATP Hygiene Monitoring System when used with the ATP Test Swab is effective at monitoring on-site cleanliness of stainless steel in food and beverage industries with...

 Open the catalog to page 2
AOAC Cert of Biolum Portable ATP Hygiene Monitoring System-3

Xi’an Tianlong Science and Technology Co., Ltd. Biolum ATP, AOAC Research Institute Performance Tested MethodsSM certification number 022401 Mean RLUa srb c RSD r Method Mean fmold developer sre RSDrf Mean RLU sr Independent RSDr Mean fmol laboratory sr RSDr a Average RLU from 10 replicates per ATP level. bs calculated from 10 replicates per ATP level. r c RSDr calculated from 10 replicates per RLU level. d Average ATP (femtomoles) from 10 theoretical replicates per ATP level. e s calculated from 10 predicted replicates per RLU level. r f s calculated from 10 predicted replicates per RLU level....

 Open the catalog to page 3
AOAC Cert of Biolum Portable ATP Hygiene Monitoring System-4

Xi’an Tianlong Science and Technology Co., Ltd. Biolum ATP, AOAC Research Institute Performance Tested MethodsSM certification number 022401 Figure 2. Regression Analysis used to Calculate LOD in Method Developer Study (1) Figure 3. Regression Analysis of RLUs to Determine ATP Concentration in Independent Laboratory Study (1)

 Open the catalog to page 4
AOAC Cert of Biolum Portable ATP Hygiene Monitoring System-5

Xi’an Tianlong Science and Technology Co., Ltd. Biolum ATP, AOAC Research Institute Performance Tested MethodsSM certification number 022401 Figure 4. Regression Analysis used to Calculate LOD in Independent Laboratory Study (1) Table 2. Raw and Calculated Data for the Stainless Steel Matrix Study – Wet Sampling (1) Dilutiona Mean RLUb

 Open the catalog to page 5
AOAC Cert of Biolum Portable ATP Hygiene Monitoring System-6

Xi’an Tianlong Science and Technology Co., Ltd. Biolum ATP, AOAC Research Institute Performance Tested MethodsSM certification number 022401 Mean RLUb

 Open the catalog to page 6
AOAC Cert of Biolum Portable ATP Hygiene Monitoring System-7

Xi’an Tianlong Science and Technology Co., Ltd. Biolum ATP, AOAC Research Institute Performance Tested MethodsSM certification number 022401 Table 7. Raw and Calculated Data for the Selectivity (1) RLU at 2500 fmol compound,0 fmol ATP NAb analyte-free water 1 ATP Adenosine 5'-triphosphate sodium salt hydrate 6720 dATP 2'-deoxyadenosine 5'-triphosphate sodium salt 431 UTP Uridine 5'-triphosphate trisodium salt 2 GTP Guanosine 5'-triphosphate sodium salt 1 TTP Thymidine 5'-triphosphate sodium salt 2 dUTP 2'-Deoxyuridine 5'-triphosphate sodium salt 1 CTP Cytidine 5'-triphosphate 2 dGTP 2'-deoxyguanosine...

 Open the catalog to page 7
AOAC Cert of Biolum Portable ATP Hygiene Monitoring System-8

Xi’an Tianlong Science and Technology Co., Ltd. Biolum ATP, AOAC Research Institute Performance Tested MethodsSM certification number 022401 Table 12. Raw and Calculated Data for the Stainless Steel Matrix Study in Independent Laboratory Study – Doughnut Residue (1) Target Wet Sampling Dry Sampling RLU a b c Sample No. RLU Mean sr RSDr Sample No. RLU Mean

 Open the catalog to page 8
AOAC Cert of Biolum Portable ATP Hygiene Monitoring System-9

Xi’an Tianlong Science and Technology Co., Ltd. Biolum ATP, AOAC Research Institute Performance Tested MethodsSM certification number 022401 REFERENCES CITED 1. Han, W. and Li, H., Validation Study for the Tianlong Biolum Portable ATP Hygiene Monitoring System for Detection of ATP with ATP Test Swab for Stainless Steel Surface, AOAC Performance TestedSM certification number 022401. Approved February 29, 2024. 2. User Manual – Biolum Portable ATP Hygiene Monitoring System User Manual. Document No. M3001, Version Biolum_Manual_TL_V1.0.002 3. Instruction Manual – Instruction Manual of ATP Test Swab....

 Open the catalog to page 9