BioHarness 3 Medical Data Sheet Zephyr BioHarness™ 3 PRODUCT DESCRIPTION The BioHarness™ 3 is a compact physiological monitoring module. It is attached to a lightweight Smart Fabric strap or garment which incorporates ECG and Breathing detection sensors. The BioHarness™ module can transmit physiological data or record it to internal memory. Bluetooth Connectivity to receiver or external sensors Heart Rate 25 – 240 BPM (±1 BPM) Breathing Rate 3 – 70 BPM (±1BPM) Device Temperature 10 – 60°C (±2°C) Position/posture ± 180° (Laying, standing) Activity in VMU (Stationary, walk, run) 3 axis Acceleration to 16g Red / Orange / Green subject status indication Transmit and/or Logging Modes EC38 Type 3 250Hz ECG Transmission & Logging 100Hz Accelerometer Logging USB connectivity for data download & charging Up to 500+ hours data storage Internal algorithms for o Estimated core temperature o Jump Test o Dash Test o Fall detection o Heart Rate Variability o Human Real Data Company Information Zephyr Technology 1 Annapolis St Suite 200 Annapolis MD 21401 Tel: +1 (443) 569-3603 Fax: +1 (443) 926-9402 Email: sales@zephyranywhere.com www.zephyr-technology.com Biomechanical and physiological research Remote Patient Monitoring Physical status monitoring in real world situations Data subject to change

Open the catalog to page 1
Zephyr BioHarness - 2

BioHarness 3 Medical Data Sheet Optional Live Long-Range Data Transmission Sensor strap/shirt User Interface Bluetooth Radio Link Zephyr Training System Optional Live Short-Range Data Transmission Sensor strap/shirt User Interface Bluetooth Radio Link Bluetooth Receiver Optional Transmit to Mobile Device Application Sensor strap/shirt Mobile Application Bluetooth Radio Link Bluetooth Mobile Device Optional Data Logging Wear & record* Download data using USB charging cradle User Interface *Device must be configured to log. Software needed for data import.

Open the catalog to page 2
Zephyr BioHarness - 3

BioHarness 3 Medical Data Sheet ° Specifications (Unless otherwise stated: Temperature = 25 C, Pressure = 1 ATM, Fresh battery, g = gravity units) Power Supply: Internal Lithium cell, rechargeable via USB charging cradle or USB wall charger. Parameter General Logging capacity Power supply voltage Battery Life – Radio transmitting Battery Life - Logging Log Download Time Charging Time Storage resolu tion Charging Cycles 300 Cycles ECG Digital resolution 10 12 bits Heart Rate ECG sensor sampling frequency 1000 Hz Range 7 25 240 ±1 BPM R-R 250 1500 ms Time to first lock At 60 bpm 15 25 S No...

Open the catalog to page 3
Zephyr BioHarness - 4

BioHarness 3 Medical Data Sheet Parameter Max. Acc’y Unit Breathing Rate Sampling frequency 25 Hz Range 9 3 70 ±1 BPM No signal response time 15 s Step change response time 15 s Device Temperature Sampling frequency 10 1 s Range 10 60 ±2 °C Stabilization Time 20 minutes Acceleration Sampling Frequency 100 Hz Range (any axis) -16 +16 g Bandwidth 50 Hz Sensitivity 12 mg Activity VMU (vector magnitude units) 11 16 g Epoch 1 s Bandwidth 0.06 9 Hz Dynamic Range (any axis) 16 g −16 Sensitivity 10 mg Noise 7.2 mg Posture Reporting frequency 1 Hz Dynamic Range 12 -180 +180 Degrees Epoch 1 s...

Open the catalog to page 4
Zephyr BioHarness - 5

BioHarness 3 Medical Data Sheet Data Output – Transmitted Data Data output is in the form of a number of messages, each of which can be enabled or disabled. Parameter Reporting Range Units Description Frequency (Hz) General Data Packet Heart Rate 1 25 – 240 BPM Beats per Minute Breathing Rate 1 3 – 70 BPM Breaths per Minute Skin Temperature 1 °C Invalid (-3276.8) always returned. Not supported. Posture 1 ±180 Degrees Vertical = 0°, Inverted = 180° Activity Level 1 16 VMU (g) Peak Acceleration 1 ±16 g Battery Voltage 1 3.5 – 4.2 V Breathing Wave Amplitude 1 0 - 65534 bits Indicative only ECG...

Open the catalog to page 5
Zephyr BioHarness - 6

BioHarness 3 Medical Data Sheet Parameter Invalid Value Posture Activity Level Peak Acceleration Battery Voltage Battery Level Breathing Wave Amplitude Breathing Wave Noise Breathing Rate Confidence ECG Amplitude Breaths per Minute Invalid always returned – not supported. Vertical = 0°, Inverted = 180° Heart Rate Confidence HR Variability System Confidence Galvanic Skin Response ROG Status Vertical Acceleration Min Vertical Acceleration Peak Lateral Acceleration Min Lateral Acceleration Peak Sagittal Acceleration Min Sagittal Acceleration Peak Device Internal Temperature Status Info Link...

Open the catalog to page 6
Zephyr BioHarness - 7

Tx Power Estimated Subject Core Temperature Aux ADC Channel 1 Aux ADC Channel 2 Aux ADC Channel 3 Reserved BioHarness 3 Medical Data Sheet 10 = +10dBm -128 386=38.6 6553.5 Implementation specific Implementation specific Implementation specific 1. The Summary data packet has a configurable reporting frequency, in the range 1 sec ~ 18 hrs. Data reported is the latest value for each parameter at time of transmission, other than Activity, Peak, Max & Min accelerations, which use the reporting interval as the epoch over which they are calculated. 2. Invalid Values may be reported if the data is...

Open the catalog to page 7
Zephyr BioHarness - 8

BioHarness 3 Medical Data Sheet Data Output – Logged Data Logging Modes • General (default) • General + ECG • General + Acceleration Summary Summary + Waveform Where used, “Bits’ refers to raw ADC output throughout. Note that some parameters are logged with greater resolution than is possible for transmitted data. General Log Parameter Heart Rate Breathing Rate Skin Temperature Posture Activity Level Peak Acceleration Battery Voltage Breathing Wave Amp GSR Level ECG Amplitude ECG Noise X Acceleration Min X Acceleration Peak Y Acceleration Min Y Acceleration Peak Z Acceleration Min Z...

Open the catalog to page 8
Zephyr BioHarness - 9

BioHarness 3 Medical Data Sheet Summary Log Parameter Heart Rate Breathing Rate Skin Temperature Posture Activity Level Peak Acceleration Battery Voltage Breathing Wave Amp Breathing Wave Noise Breathing Rate Confidence ECG Amplitude ECG Noise Heart Rate Confidence Heart Rate Variability System Confidence GSR ROG Vertical Acceleration Min Vertical Acceleration Peak Lateral Acceleration Min Lateral Acceleration Peak Sagittal Acceleration Min Sagittal Acceleration Peak Internal Device Temperature Status Info Link Quality RSSI BPM BPM °C Degrees VMU(g) g V bits Tx Power Estimated Core...