NIRS Analyzer Pro
1 / 8Páginas

Excertos do catálogo

NIRS Analyzer Pro - 1

NIRS Analyzer Pro is a process analysis system based on high-resolution diode array technology. It provides nondestructive analysis of granules, powders, slurries or opaque samples directly in the process line without bypass – a true in-line system.

Abrir o catálogo na página 1
NIRS Analyzer Pro - 2

• High resolution diode array technology for accurate and continuous analysis in reflectance or transmittance mode • Built-in instrument factory standardisation for quick and simple implementation of one or several analysers • Uptime protection systems for low maintenance • Dedicated sample interfaces provides accuracy and rapid implementation • Instant measurement of complete wavelength range for direct measurement of fast moving samples • Quantitative and qualitative data for better in-line process control • Dual lamp technology retaining accuracy when the backup lamp is activated...

Abrir o catálogo na página 2
NIRS Analyzer Pro - 3

NIRS Analyzer Pro is a process analysis system based on high re­­ lution diode array technology. It provides non so­­ destruc­­ analysis of products such as flour, feeds, tive meals, ground meat and dairy products directly in the process line without bypass. The analyser is housed in a robust cabinet mounted at the relevant location in the production area. Measurements are displayed in the control room and results can be fed into a regulation system for closed-loop automa­­ tic control. The solution helps to optimise the use of raw materials and to consistently run production closer to...

Abrir o catálogo na página 3
NIRS Analyzer Pro - 4

Dedicated sample interfaces 04 Window reflectance In-line analysis of paste, granulates, powdered products etc in pipes or transport systems without bypass can be performed. The products pass over the interface window. The window reflection interface can easily be installed into the production line using standard GEA Tuchenhagen flowcells or welding an interface flange into the wall of the pipe/transport system. Temperature 150 °C (302 °F) Pressure Vacuum > 1 torr, Pressure < 3’000 PSI Lens Sapphire; Diameter 45 mm, thickness 12 mm, with food grade EPDM O-ring seal Hygiene USDA, Dairy Pipe...

Abrir o catálogo na página 4
NIRS Analyzer Pro - 5

Powder probe In line analysis of fine powder like dairy powders etc. The powder probe interface can easily be installed into a hopper or pipe with free falling product. The probe has no moving parts. The probe is automatically cleaned with compressed air before each analysis. Material Diameter Length Fiber Air Temperature Installation Hygiene Cleaning Teflon (PCTFE) 1˝ 7˝ Steel armoured fiber bundle (1, 3, 5 or 10 meters) Clean compressed air 43-72 PSI Max 120 °C 1˝ Swagelok crimp fitting USDA, Dairy Clean with compressed air or wash in water, hot or cold depending on product

Abrir o catálogo na página 5
NIRS Analyzer Pro - 6

Lateral transmittance In line analysis of slurries and viscous products such as WPC, Cream Cheese, Mozarella etc. The Lateral transmittance probe does not restrict the flow rate of the product and can easily be installed in the production line using a standard GEA Tuchenhagen flowcell for installation in a pipe or by welding an interface flange into the wall of a tank. Materials Lens Temperature Pressure Hygiene Optical fiber Pipe flowcells Stainless steel Sapphire, 5 mm thick, with food grade EPDM O-ring seal Max 150 °C (302 °F) Max 200 PSI (13.79 bars) – Warning! Varinline access units...

Abrir o catálogo na página 6
NIRS Analyzer Pro - 7

Technical Specifications NIRS Analyzer Pro Reflection Analysis time 5 - 50 ms / integration time; Average time per result 3-15 seconds Measurement mode Reflectance (Window reflectance; Direct Light, Powder Probe) Wavelength range 1’100-1’650 nm Detector InGaAs Diode array Spectral dispersion 1.1 nm / pixel NIRS Analyzer Pro Transmittance Analysis time 5 ms-60 sec / integration time; Average time per result 3-15 seconds Measurement mode Transmittance (Lateral transmittance, transmittance probe pair) Wavelength range 850-1’050 nm Detector Si Diode array Spectral dispersion 1.0 nm / pixel...

Abrir o catálogo na página 7
NIRS Analyzer Pro - 8

NIRS Analyzer PRO – ContactReflection NIRS Analyzer PRO – FiberSystem NIRS Analyzer PRO – DirectLight/NonContact Software 6.6069.102 Vision Software Subject to modifications Layout by Ecknauer+Schoch ASW, printed in Switzerland by Metrohm AG, CH-9100 Herisau 8.928.5000EN – 2013-03 Standards 6.7450.070 NIRS Certified Process Reflectance and Wavelength Standard (1") 6.7450.080 NIRS Certified Process Reflectance and Wavelength Standard (0.5") 6.7450.090 NIRS Certified Process Reflectance and Wavelength Standard (1", 45 degrees) 6.7450.100 NIRS Certified System2 Reflectance Standards (set of 7...

Abrir o catálogo na página 8

Todos os catálogos e folhetos técnicos Metrohm

  1. 2029

    1 Páginas

  2. 2026 Titrolyzer

    8 Páginas

  3. Mira P Handheld

    8 Páginas

  4. µStat-i 400

    2 Páginas

  5. 899 Coulometer

    8 Páginas

  6. OMNIS

    16 Páginas

  7. Sample preparation

    12 Páginas

  8. Titrotherm

    12 Páginas

  9. 888/890 Titrando

    12 Páginas

  10. OMNIS

    8 Páginas

  11. PSTAT mini

    8 Páginas

  12. Dosing Package

    4 Páginas

  13. Dosing Interface

    3 Páginas

  14. Dosimat plus

    4 Páginas

  15. KF Evaporator

    4 Páginas

  16. 909 UV Digester

    8 Páginas