EDUCATIONAL GUIDE QC Multi-Rules QUALITY CONTROL
Abrir o catálogo na página 1
QC Multi-Rules What are QC Multi-Rules? Based on statistical process control, QC Multi-rules use a combination of control rules to decide whether an analytical run is in-control or out-of-control. Control Data What is Measurement Uncertainty? Recognising out-of-control QC results using a single rule A laboratory favourite is the 1 result outside 2SD rule. This is a great rule for alerting lab staff to possible out-of-control events, however, you must be careful as it does have a high false rejection rate. Remember, it is normal for 1/20 results to be outside 2SD! It is important that you...
Abrir o catálogo na página 2
Types of Analytical Errors QC Multi-Rules can be used to detect systematic or random errors. Recognising if you have either a systematic or random error can help you identify the source of the problem. Random Errors The characteristics of Random Errors are as follows: • Error which varies in an unpredictable manner (in magnitude and sign) whenever a large number of measurements of the same quantity are made under effectively identical conditions. • Difficult to eliminate but repetition can reduce the influence of random errors.between ± one SD of the average and around 95% of all readings...
Abrir o catálogo na página 3
Explained QC Multi-Rules There are 6 basic QC Multi-Rules. Some of these detect random error and some detect systematic error (which may indicate a bias in the system). Rule 12S (1 result exceeds 2SD) This rule is designed as a warning rule only. If a single QC result is more than ±2 standard deviations from the mean, this run along with previous analytical runs should be evaluated before accepting the run and reporting results. The 12S rule is designed to warn that random error or systematic error may be present. Using the 12S rule alone may lead to frequent rejection of valid runs (false...
Abrir o catálogo na página 4
Rule 2 of 32S (2 of 3 results exceed 2SD) This is a variation of the 22S rule and is used to detect systematic error. The rule is violated when any two levels of control in a run exceed 2 standard deviations on the same side of the mean. Rule R4S (Range exceeds 4SD) This rule is violated if there is at least 4 standard deviations difference between control values within a single run. In this example the level 1 control is +2S above the mean and the level 2 control is -2S below the mean. The total difference is 4s. This rule identifies random error only. Rule 41S (4 results exceed 1SD) This...
Abrir o catálogo na página 5
Rule 10x (10 results same side of the mean) This rule is violated if ten consecutive control results regardless of level fall on the same side of the mean. The rule detects systematic error and can be applied both within and across runs. Violation of this rule may not require rejection of the run but rather indicate that instrument maintenance or calibration is needed. The 10x rule can be modified to 7,8,9 or 12 consecutive results on the same side of the mean, each giving varying levels of sensitivity to systematic bias. This rule is violated when a group of seven consecutive results for a...
Abrir o catálogo na página 6
Tools to Assist Labs Apply QC Multi-Rules To apply QC Multi-Rules you firstly collect your control measurements in the same way that you would for a Levey-Jennings chart. From this you establish a mean and standard deviation (SD). All you are changing is the control limits and how the data will be interpreted. Online QC software with real-time peer group statistics Acusera 24•7 will apply QC multi-rules automatically Compatible for use with the Acusera range of third party controls, the Acusera 24•7 software is designed to help laboratories monitor and interpret their QC data. Access to an...
Abrir o catálogo na página 7
Copyright © 2019 Randox Laboratories Ltd. All rights Reserved. VAT number: GB 151682708. Product availability may vary from country to country. Some products may be for Research Use Only. For more information on product application and availability, please contact your local Randox Representative.
Abrir o catálogo na página 8Todos os catálogos e folhetos técnicos Randox Laboratories
-
LT735 Vivalytic Overview
32 Páginas
-
Acusera New Controls
8 Páginas
-
Acusera 24.7
24 Páginas
-
Acusera Third Party Controls
108 Páginas
-
Acusera SMART Controls
12 Páginas
-
Company Overview
40 Páginas
-
EDUCATIONAL GUIDE Gastroenteritis
12 Páginas
-
Superior Performance & Unique Tests
32 Páginas
-
Reagents Brochure
64 Páginas
-
sdLDL-C
8 Páginas
-
RX SERIES ANALYSER OVERVIEW
24 Páginas
-
RX modena
20 Páginas
-
RX misano
20 Páginas
-
RX Daytona Plus
20 Páginas
-
Rx Imola
20 Páginas
-
RX Monaco
20 Páginas
-
2024 Product List
64 Páginas
-
LT107 Evidence Investigator
32 Páginas
-
VeraSTAT-V
16 Páginas
-
LT033 RIQAS Explained
64 Páginas
-
VERASTAT
16 Páginas
-
Endocrine Array
4 Páginas
-
Cerebral Arrays
4 Páginas
-
Thyroid Arrays
4 Páginas
-
Respiratory Multiplex Array
8 Páginas
-
LT252 Respiratory Multiplex Array
8 Páginas
-
LT253 Molecular Diagnostics
20 Páginas
-
LT367 FH Array Brochure
4 Páginas
-
The role of EQA in QC
8 Páginas
-
Basic QC Stastics
8 Páginas
-
Commutability Guide
4 Páginas
-
How to measure uncertainty
8 Páginas
-
ISO 15189 Educational Guide
8 Páginas
-
Meeting ISO 15189 Requirements
4 Páginas
-
Troubleshooting QC Errors
8 Páginas
-
Qnostics
52 Páginas
-
Adiponectin LT519
28 Páginas
-
Specific Proteins
40 Páginas
-
Linearity sets
12 Páginas
-
Antioxidants
16 Páginas
-
Cardiology & Lipid Testing
28 Páginas
-
Diabetes Portfolio
28 Páginas
-
Total Bile Acids
4 Páginas
-
VIVALYTIC
30 Páginas
-
RANDOX DISCOVERY
36 Páginas
-
HbA1c
2 Páginas
-
technology at your fingertips
2 Páginas
-
RIQAS Troubleshooting checklist
1 Páginas
-
Preparing QC
1 Páginas
-
Point of Care Testing
12 Páginas
-
LT630 Monitoring EQA Performance
1 Páginas
-
LT394 Using QC Multirules
1 Páginas
-
Guide to running QC
1 Páginas
-
Tumour Marker Arrays
4 Páginas
-
Which QC is the Right QC
8 Páginas
-
How often is right for QC
6 Páginas
-
Cardiac Risk Multiplex Array
4 Páginas
-
Biochip Array Technology explained
12 Páginas
-
LT250 STI Multiplex Array NOV14
8 Páginas
-
LT241 Metabolic Array MAY15
8 Páginas
-
KRAS / BRAF / PIK3CA Array*
4 Páginas
-
Custom Arrays for Biochip
12 Páginas
-
LT169 Cardiac Array
4 Páginas
Catálogos arquivados
-
ACUSERA
108 Páginas
-
Evidence Evolution
28 Páginas
-
Evidence
16 Páginas
-
Metabolic Syndrome Arrays
8 Páginas
-
STI Multiplex Array
8 Páginas
-
Molecular Testing
16 Páginas
-
Evidence Investigator
20 Páginas
-
Fertility Array
4 Páginas
-
LT107 Evidence Investigator - 2014
20 Páginas
-
Rx Daytona
16 Páginas
-
Evidence Investigator
16 Páginas
-
Evidence
16 Páginas