EDUCATIONAL GUIDE Troubleshooting QC Errors QUALITY CONTROL
Abrir o catálogo na página 1
Troubleshooting QC Errors So you ran QC this morning and realised that one of your analytes has been flagged as “out-of-control”, what do you do next? Do you ignore the warning and continue patient testing, repeat the control until it’s within range or do you halt patient testing and investigate the source of the error? Your QC Has Failed, What Do You Do Next? When it comes to troubleshooting QC errors, unfortunately there is no easy path to take. However, it’s important that you have standard operating procedures in place, outlining what to do in the event of an out-of-control error....
Abrir o catálogo na página 2
2. Review Your Levey-Jennings Charts To Understand The Type Of Error That Has Occurred There are two types of analytical errors: Systematic Errors Random Errors Both of these types of errors can be recognised on a standard Levey-Jennings chart and by using QC multi-rules. Identifying the type of error will help you relate the error to a possible cause. Systematic Error Systematic errors create a characteristic bias and can be identified by a change in the mean of control values. The change in the mean may be either a gradual trend or an abrupt shift in control results. Standard Deviation...
Abrir o catálogo na página 3
Using QC Multi-Rules To Identify Systematic & Random Error You can use QC multi-rules to differentiate between systematic and random error. +3s Standard Deviation Standard Deviation This is a warning rule that is broken when a single control observation is outside the +/-2s limits. This rule warns that either a random or systematic error is present in the test system. If no other unacceptable results are apparent in the test system, it must be assumed that this is simply a random error and no further troubleshooting action is necessary. Any QC result outside +/-3s breaks this rule. This...
Abrir o catálogo na página 4
3. Relate The Type Of Error To Possible Causes Ask yourself questions in order of likely relevance and review the most common solutions to the problem. Dependent on whether you have identified a systematic or random error in your system, this can help you determine the possible root cause of the error. See below for some common causes of systematic and random errors. Systematic Error: Shift 1. Change or failure in light source 2. Major instrument maintenance 3. Failure in sampling system 4. Change in temperature 5. Failure in reagent dispense system 6. Change in reagent formulation 7. New...
Abrir o catálogo na página 5
ACUSERA True third party quality controls As a world leading manufacturer of multi-analyte true third party controls, thousands of laboratories rely on Randox to accurately assess test system performance and ultimately empower them with the confidence required to release patient test results. With more than 390 analytes available, the number of individual controls required to cover your test menu is significantly reduced while simultaneously reducing costs, time and storage space. A choice of formats is available, including liquid or lyophilised, which ensures flexibility and suitability...
Abrir o catálogo na página 6
ACUSERA 24•7 Interlaboratory Data Management Designed for use with the Acusera range of third party controls, the Acusera 24•7 software helps laboratories monitor and interpret their QC data. Access to an impressive range of features, including interactive charts, the automatic calculation of Measurement Uncertainty & Sigma Metrics and live peer group data generated from our extensive database of laboratory participants, ensures Acusera 24•7 is the most comprehensive package available. • Advanced statistical analysis with automatic calculation of performance metrics including; Sigma, UM, TE...
Abrir o catálogo na página 7
Copyright © 2019 Randox Laboratories Ltd. All rights Reserved. VAT number: GB 151682708. Product availability may vary from country to country. Some products may be for Research Use Only. For more information on product application and availability, please contact your local Randox Representative.
Abrir o catálogo na página 8Todos os catálogos e folhetos técnicos Randox Laboratories
-
LT735 Vivalytic Overview
32 Páginas
-
Acusera New Controls
8 Páginas
-
Acusera 24.7
24 Páginas
-
Acusera Third Party Controls
108 Páginas
-
Acusera SMART Controls
12 Páginas
-
Company Overview
40 Páginas
-
EDUCATIONAL GUIDE Gastroenteritis
12 Páginas
-
Superior Performance & Unique Tests
32 Páginas
-
Reagents Brochure
64 Páginas
-
sdLDL-C
8 Páginas
-
RX SERIES ANALYSER OVERVIEW
24 Páginas
-
RX modena
20 Páginas
-
RX misano
20 Páginas
-
RX Daytona Plus
20 Páginas
-
Rx Imola
20 Páginas
-
RX Monaco
20 Páginas
-
2024 Product List
64 Páginas
-
LT107 Evidence Investigator
32 Páginas
-
VeraSTAT-V
16 Páginas
-
LT033 RIQAS Explained
64 Páginas
-
VERASTAT
16 Páginas
-
Endocrine Array
4 Páginas
-
Cerebral Arrays
4 Páginas
-
Thyroid Arrays
4 Páginas
-
Respiratory Multiplex Array
8 Páginas
-
LT252 Respiratory Multiplex Array
8 Páginas
-
LT253 Molecular Diagnostics
20 Páginas
-
LT367 FH Array Brochure
4 Páginas
-
The role of EQA in QC
8 Páginas
-
Basic QC Stastics
8 Páginas
-
Commutability Guide
4 Páginas
-
How to measure uncertainty
8 Páginas
-
ISO 15189 Educational Guide
8 Páginas
-
Meeting ISO 15189 Requirements
4 Páginas
-
QC Multirules Educational Guide
8 Páginas
-
Qnostics
52 Páginas
-
Adiponectin LT519
28 Páginas
-
Specific Proteins
40 Páginas
-
Linearity sets
12 Páginas
-
Antioxidants
16 Páginas
-
Cardiology & Lipid Testing
28 Páginas
-
Diabetes Portfolio
28 Páginas
-
Total Bile Acids
4 Páginas
-
VIVALYTIC
30 Páginas
-
RANDOX DISCOVERY
36 Páginas
-
HbA1c
2 Páginas
-
technology at your fingertips
2 Páginas
-
RIQAS Troubleshooting checklist
1 Páginas
-
Preparing QC
1 Páginas
-
Point of Care Testing
12 Páginas
-
LT630 Monitoring EQA Performance
1 Páginas
-
LT394 Using QC Multirules
1 Páginas
-
Guide to running QC
1 Páginas
-
Tumour Marker Arrays
4 Páginas
-
Which QC is the Right QC
8 Páginas
-
How often is right for QC
6 Páginas
-
Cardiac Risk Multiplex Array
4 Páginas
-
Biochip Array Technology explained
12 Páginas
-
LT250 STI Multiplex Array NOV14
8 Páginas
-
LT241 Metabolic Array MAY15
8 Páginas
-
KRAS / BRAF / PIK3CA Array*
4 Páginas
-
Custom Arrays for Biochip
12 Páginas
-
LT169 Cardiac Array
4 Páginas
Catálogos arquivados
-
ACUSERA
108 Páginas
-
Evidence Evolution
28 Páginas
-
Evidence
16 Páginas
-
Metabolic Syndrome Arrays
8 Páginas
-
STI Multiplex Array
8 Páginas
-
Molecular Testing
16 Páginas
-
Evidence Investigator
20 Páginas
-
Fertility Array
4 Páginas
-
LT107 Evidence Investigator - 2014
20 Páginas
-
Rx Daytona
16 Páginas
-
Evidence Investigator
16 Páginas
-
Evidence
16 Páginas