ProBeam Facility & Interface Requirements
1 / 20Páginas

{{requestButtons}}

Excertos do catálogo

ProBeam Facility & Interface Requirements - 1

PROTON BEAM THERAPY FACILITY Facility and Interface Requirements VARIAN PARTICLE THERAPY I FACILITY REQUIREMENTS

Abrir o catálogo na página 1
ProBeam Facility & Interface Requirements - 4

Introduction The ProBeam™ proton therapy system from Varian Medical Systems must be installed in a building that meets Varian requirements and conforms to local regulations, especially regulations on radiation safety. Although the design and building of the facility are the customer’s responsibility, Varian will provide ongoing support to assist with the process. Varian’s proposal will be based on the following assumptions: • Varian will provide regular on-site support in order to assist with the building design and design approval, as well as to minimize delays. • The customer’s building...

Abrir o catálogo na página 4
ProBeam Facility & Interface Requirements - 5

Facility design and radiation protection The architectural design of the proton therapy facility is driven by clinical requirements such as the number and type of treatment rooms; the resulting radiation shielding layout necessary to protect personnel, patients and the public; and the space allocation for clinical and technical areas. The proposed building layout is based on a modular concept: Customer requirements can be incorporated without changing the base allocation for technical areas needed for the proton therapy equipment. Treatment rooms and other clinical areas can be added...

Abrir o catálogo na página 5
ProBeam Facility & Interface Requirements - 6

The standard Varian proton therapy system layout has been evaluated by several authorities and has been found to satisfy the respective local regulations. Therefore, the architect’s team can use the wall thicknesses in the standard layout as estimates until the customer’s radiation safety specialist provides final confirmation. Varian will support radiation shielding discussions by providing a standard set of documentation and input during meetings with the customer’s radiation safety specialist. Radiation protection equipment Typically, the BDCT is responsible for installing permanent...

Abrir o catálogo na página 6
ProBeam Facility & Interface Requirements - 7

Shielding wall openings The beam transport system area requires a wall opening for inserting magnets and related beam transport and energy selection system equipment. The opening must have restricted access and should be thermally insulated. In addition, the cyclotron maze must have a removable section as indicated in Figure 1. Non-radiation areas Equipment will be inserted via normal doors, facade doors in the upper level, and elevators installed in the building. Doors or wall openings up to the full ceiling height must be provided in many areas for transport into the building. Door sills...

Abrir o catálogo na página 7
ProBeam Facility & Interface Requirements - 8

Electrical rooms and control rooms All electrical rooms must be equipped with computer room floors. On the upper level, the load bearing capacity must be 50 kN/m2 (1,050 lbs/ft2). All cabinets require frames, some cabinets’ frames must support weights up to 5 metric tons. A detailed specification will be provided after contract approval. Building settlement Outside areas Ground reinforcements to install mobile cranes of up to 800 metric tons for rigging of heavy components will be provided by the customer, near the outside walls of the respective equipment room and near the entrance to the...

Abrir o catálogo na página 8
ProBeam Facility & Interface Requirements - 9

Required rooms and building occupation The rooms mentioned in Tables 2 and 3, and Figures 1 and 2 must be available for the installation of the equipment provided by Varian. Room dimensions are given in Figures 1 and 2. Any other equipment that is installed in these rooms must be coordinated with Varian. Otherwise, it will be assumed that no electrical cabinets or other devices from companies involved in the project will be installed in these rooms. Room description Cyclotron area Figure 1; please note pit and roof opening dimensions Energy selection system and beam transport system Main...

Abrir o catálogo na página 9
ProBeam Facility & Interface Requirements - 10

Room description 20 m2 (215 ft2) each; also container offices outside building Conference room Shared use with customer Changing rooms/restrooms­ Table 3. List of temporary rooms provided by the BDCT starting 6 months prior to RFE Room description For X-ray film development For storage of QA equipment near the clinical use area Immobilization couch storage Immobilization couches for active patients, including disinfection and raw material storage Table 4. Recommended rooms The project schedule defines one milestone for start of installation and commissioning of the proton therapy equipment:...

Abrir o catálogo na página 10
ProBeam Facility & Interface Requirements - 11

The floors in the electrical rooms must be treated with a dust- and water-repellent coating. The coating must be sufficient to withstand installation of cable trays and double floors without degradation. An access floor (computer floor) must be installed in all electrical rooms. The floors, walls, and ceilings in all rooms in which the proton therapy equipment is installed must be dust- and water-repellent. The color white is recommended in order to allow contamination to be recognized more easily and to ensure better light conditions. At RFE HVAC, electrical supply, and cooling water,...

Abrir o catálogo na página 11
ProBeam Facility & Interface Requirements - 12

of 28°C ±2°C. The cyclotron requires a temperature regulation with ±1°C temperature tolerance and a freely selected set-point selection between 30°C and 37°C for the supply water temperature. The table below explains this in more detail. The total load in this circuit will not exceed ~ 650 kW. 2. “Power Supply”: The RF amplifier, test load, and magnet power supply units are cooled with deionized water. The activation of the water in this circuit by neutrons is irrelevant. Supply temperature: 24°C. The total load in this circuit will not exceed ~ 490 kW. 3. “Cryogenics”: The helium...

Abrir o catálogo na página 12
ProBeam Facility & Interface Requirements - 13

Humidity requirements: • All areas < 65%, non-condensing Compressed air (cooling room, workshops, electrical rooms) for operation of machine tools and cleaning • Additionally in electrical rooms noncondensing at 18ºC (64°F) Pressure: 5–6 bar, relative to the surrounding atmosphere A maximum of 200 kW of heat is estimated to be generated by the equipment (magnets and power supplies) which will be dissipated in the ambient air. During typical operation, Varian estimates that ≤100 kW is dissipated in the ambient air by heated equipment parts. Ventilation Varian recommends that all radiation...

Abrir o catálogo na página 13

Todos os catálogos e folhetos técnicos Varian Oncology

  1. Halcyon

    15 Páginas

  2. Eclipse

    7 Páginas

  3. TrueBeam Brochure

    11 Páginas

  4. ProBeam Brochure

    16 Páginas

Catálogos arquivados

  1. Trilogy Brochure

    7 Páginas

  2. VitalBeam

    8 Páginas

  3. HyperArc Overview

    2 Páginas

  4. Eclipse™

    2 Páginas

  5. ProBeam 360

    2 Páginas

  6. VELOCITY SOFTWARE

    3 Páginas

  7. 360 ONCOLOGY

    3 Páginas

  8. HALCYON

    12 Páginas

  9. PaxScan ® 3030X

    2 Páginas

  10. PaxScan ® 4336X

    4 Páginas

  11. PaxScan ® 4343R

    2 Páginas

  12. Nexus-DRF™

    2 Páginas

  13. 2520DX

    2 Páginas

  14. 1515DXT

    2 Páginas

  15. PaxScan 1508 DXT

    2 Páginas

  16. PaxScan 1313DXT

    2 Páginas

  17. 1308DXT

    2 Páginas

  18. Clinac Brochure

    12 Páginas

  19. Edge Brochure

    22 Páginas

  20. NDI-452

    4 Páginas

  21. M-101G

    6 Páginas

  22. M-105SP

    6 Páginas

  23. M-109

    6 Páginas

  24. M-113

    6 Páginas

  25. M-143

    6 Páginas

  26. M-147SP

    6 Páginas

  27. M-146

    6 Páginas

  28. M-147

    6 Páginas

  29. RAD-85S

    6 Páginas

  30. M-149

    6 Páginas

  31. M-151

    10 Páginas

  32. M-152

    12 Páginas

  33. M-153

    8 Páginas

  34. PaxScan 3030+

    2 Páginas

  35. M-171

    6 Páginas

  36. PaxScan 2520E+

    2 Páginas

  37. PaxScan 4030E

    2 Páginas

  38. PaxScan 4336R

    2 Páginas

  39. PaxScan® 2020X

    2 Páginas

  40. PaxScan® 2020+

    2 Páginas

  41. 1313

    2 Páginas

  42. PaxScan® 1308

    2 Páginas

  43. Vitesse 3.0

    3 Páginas

  44. BrachyVision

    7 Páginas

  45. RapidArc

    8 Páginas

  46. PaxScan ® 4030CB

    2 Páginas

  47. PaxScan ® 1313DX

    2 Páginas

  48. PaxScan ® 1308DX

    2 Páginas

  49. Vitesse? 2.5

    2 Páginas

  50. BrachyVision? 8.5

    7 Páginas

  51. Varian BrachyTherapy

    147 Páginas